Clinical Validation Study Protocol for SmartEyes Smart Optometry Measurement Device (Medical Device Clinical Validation Study) (CVSP-SOMD)

2026년 9월 2일 업데이트: The Hong Kong Polytechnic University

Clinical Validation Study of SmartEyes Smart Optometry Measurement Instrument for Objective Refraction Agreement, Repeatability, Optical Eye Relaxation Function, and Refractive Fine-Scale Prediction of Axial Length Growth in Pediatric and Adolescent Outpatient Refraction Settings

This study aims to evaluate the SmartEyes smart refraction measurement device in two phases under real outpatient optometry room visits at Wuhan Children's Hospital. The first phase used a prospective, single-center, stratified paired cross-sectional clinical validation design, evaluating the consistency between SmartEyes, Topcon KR-800, and subjective refraction/post-dilation refraction results among outpatient children and adolescents without and those requiring dilation, and evaluating the repeatability of the three devices under corresponding conditions. The second phase added prospective cohort follow-up to explore the relationship between SmartEyes fine scale refractive data and axial length growth, evaluating whether these results better match or predict short-term axial growth trends than traditional 0.25D stepped objective refraction results.

연구 개요

상세 설명

This prospective, single-center, non-interventional clinical validation study is designed to systematically evaluate the clinical performance, accuracy, repeatability, and functional utility of the SmartEyes optometric measurement device in pediatric and adolescent outpatient settings.

The study enrolls pediatric and adolescent subjects presenting for routine optometric examinations, covering a diverse spectrum of refractive statuses, including myopia, hyperopia, and astigmatism. The research design is structured into distinct phases and cohorts to thoroughly assess device performance under varying physiological conditions:

Phase I (Cross-Sectional Evaluation): Subjects are allocated into two parallel subgroups based on clinical routine and requirements:

Non-Cycloplegic Subgroup: Subjects complete two independent measurements with each of the three devices (Topcon KR-800, and the SmartEyes device) within a single visit, followed by subjective refraction. Between consecutive measurements, subjects are required to leave the chin/forehead rest and reposition to verify device repeatability. SmartEyes outputs its refractive results after completing its built-in optical eye relaxation protocol.

Cycloplegic Subgroup: Subjects undergo one measurement with each of the three devices prior to cycloplegia. Following routine clinical cycloplegia/ciliary muscle paralysis and subsequent subjective refraction, subjects complete two additional independent measurements with each of the three devices to evaluate repeatability and agreement under cycloplegic conditions. (Note: No extra or study-mandated cycloplegic drugs are administered solely for research purposes).

Phase II (Follow-up Evaluation): Conducts longitudinal tracking and follow-up assessments to evaluate the ongoing stability and consistency of the measurement devices over time.

To thoroughly validate the device's diagnostic capabilities, the study focuses on primary evaluation dimensions including accuracy validation (assessing agreement of sphere, cylinder, axis, spherical equivalent, and corneal curvature via Bland-Altman analysis, ICC, and Pearson correlation), repeatability validation through consecutive independent scans, and the assessment of adaptive optical relaxation functions.

This study does not involve any investigational drugs, invasive procedures, or alterations to standard clinical care pathways. The generated data will establish robust clinical evidence supporting the device's accuracy and stability, while providing standardized datasets for the continuous optimization of artificial intelligence-driven optometric algorithms and smart assistance platforms.

연구 유형

관찰

등록 (추정된)

200

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 어린이

건강한 자원 봉사자를 받아들입니다

아니

샘플링 방법

비확률 샘플

연구 인구

Pediatric and adolescent outpatients presenting for routine optometric examinations and refractive status evaluation at the clinical site.

설명

Inclusion Criteria:

  1. Aged 6-17 years old, no gender limit;
  2. Visiting the outpatient optometry room of Wuhan Children's Hospital, clinically required refraction examination;
  3. Ability to cooperate in completing SmartEyes, Topcon KR-800, Topcon KR-1W examinations;
  4. Able to complete reliable subjective refraction;
  5. Subjects in the mydriasis subgroup should be those who need mydriasis/cycloplegia refraction for clinical diagnosis and treatment;
  6. Those who participate in the second phase cohort must be able to complete the axial length measurement and be willing to return to complete the 3-month (and 6-month) follow-up as planned;
  7. Guardians are fully informed and sign written informed consent;
  8. Subject children accept research notifications based on their age and understanding ability, and express agreement or no objection to participating in the research.

Exclusion Criteria:

  1. Those unable to maintain stable fixation, unable to cooperate with desktop equipment for examination, or unable to perform reliable subjective refraction;
  2. Obvious strabismus, nystagmus, severe amblyopia, or other conditions affecting the reliability of objective/subjective refraction;
  3. Keratoconus, obviously irregular astigmatism, corneal scarring, corneal edema, or other corneal diseases that significantly affect refractive measurement;
  4. Moderate to severe cataracts, obvious lens opacity, vitreous opacity, fundus diseases, and other conditions that may affect the refraction endpoint;
  5. History of previous corneal refractive surgery or intraocular surgery;
  6. Still wearing orthokeratology lenses, RGP, or other contact lenses on the day and determined by the investigator to affect measurement results;
  7. Participants in the second phase cohort who are unable to complete follow-up, plan to undergo eye surgery in the short term, or have investigators who judge axial measurement results to be clearly unreliable will not be included in the cohort follow-up analysis;
  8. Other circumstances where the investigator determines the participant is unsuitable for this study.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

코호트 및 개입

그룹/코호트
개입 / 치료
No pupil dilation subgroup(Phase I)
No pupil dilation subgroup: Subjects completed two independent tests each of Topcon KR-800 and SmartEyes smart optometry devices during the same outpatient procedure, followed by subjective refraction. Between two independent tests, subjects were asked to leave the mandibular rest/forehead rest and reposition to verify the repeatability of each device. Each SmartEyes test outputs the refraction results after completing the device's built-in optical eye relaxation process.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Routine refractive examination
Required pupil dilation subgroup(Phase I)
Before pupil dilation, subjects completed one test each of Topcon KR-800, and SmartEyes devices; Then, following the routine outpatient procedures of Wuhan Children's Hospital, the clinically required pupil dilation/ciliary muscle palsy was completed, followed by subjective refraction after pupil dilation. After meeting clinical examination requirements, each of the three devices underwent two independent tests to evaluate repeatability and consistency under dilated pupil conditions. No additional mydriatic drugs were used for research purposes.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Routine refractive examination
Follow-up visit group (Phase II)
Subjects may be selected from those willing to continue follow-up in the first phase, or from newly screened eligible subjects in the same outpatient setting. All cohort subjects completed baseline refraction and axial measurement, followed up at 3 months (and 6 months). To reduce the impact of inspection sequence, the first phase uses a pre-established random sequence or balanced rotation sequence for the three devices; The dilation subgroup should try to maintain the same device sequence for the same subject before and after dilation. Final subjective optometry is performed by an experienced optometrist, who must not view the results of SmartEyes, or KR-800 before recording the subjective refraction endpoint.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Used to acquire objective optometric parameters, including refractive errors and corneal measurements, during the clinical evaluation and comparison protocol.
Routine refractive examination
FDA-certified equipment used for measuring the axial length

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Refractive Measurement Agreement and Repeatability across Subgroups
기간: Phase I clinical evaluation (completed during the baseline outpatient examination sessions)

Evaluate the agreement between SmartEyes, Topcon KR-800, and subjective/cycloplegic reference refractions, as well as the independent measurement repeatability of each device, across both non-cycloplegic and cycloplegic subgroups.

Assess whether the optical eye relaxation function of SmartEyes yields refractive outcomes closer to subjective and cycloplegic references than conventional objective methods.

Phase I clinical evaluation (completed during the baseline outpatient examination sessions)

2차 결과 측정

결과 측정
측정값 설명
기간
Absolute axial length change measured by optical biometry
기간: From baseline through the Phase II longitudinal follow-up period (6 months)
Absolute change in axial length (unit: mm) measured by optical biometry device.
From baseline through the Phase II longitudinal follow-up period (6 months)
Axial length growth rate measured by optical biometry
기간: From baseline through the Phase II longitudinal follow-up period (6 months)
Axial length monthly growth rate (unit: mm/month), derived from serial axial-length measurements obtained by optical biometry device.
From baseline through the Phase II longitudinal follow-up period (6 months)
Correlation between fine-graded refractive change and absolute axial length change
기간: From baseline through the Phase II longitudinal follow-up period (6 months)
Pearson correlation coefficient between fine-graded spherical equivalent change (SmartEyes autorefractor, diopter) and absolute axial-length change (optical biometry, mm).
From baseline through the Phase II longitudinal follow-up period (6 months)

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 수석 연구원: Hong Jie, MD, Wuhan Children's Hospital affiliated with Tongji Medical College, Huazhong University of Science and Technology
  • 연구 의자: Mingguang He, The Hong Kong Polytechnic University

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (추정된)

2026년 8월 28일

기본 완료 (추정된)

2027년 2월 2일

연구 완료 (추정된)

2027년 6월 2일

연구 등록 날짜

최초 제출

2026년 8월 21일

QC 기준을 충족하는 최초 제출

2026년 9월 2일

처음 게시됨 (실제)

2026년 9월 3일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 9월 3일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 9월 2일

마지막으로 확인됨

2026년 8월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

예

미국에서 제조되어 미국에서 수출되는 제품

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다