Evaluation of 4D Intracardiac Echocardiography (ICE) During Structural Heart and Electrophysiology Procedures (4D-ICE)
Assessment of Utility of Siemens Healthineers 4D ICE and Premium CV System Workflow in Evaluation of Cardiac Anatomy and Physiology During Electrophysiology and Structural Heart Procedures
The purpose of this study is to evaluate the imaging performance and procedural safety of a prototype 4D ICE catheter during real-time intracardiac imaging and to optimize device imaging parameters, compared with current standard imaging modalities.
TEE is the current standard of care and will serve as the primary imaging modality for clinical guidance during the EP and Structural Heart Disease (SHD) procedures. In this research protocol, TEE is not being investigated. ICE will be used as an adjunct imaging modality for research data collection and workflow evaluation only. All procedures will be performed using commercially available, FDA-approved Watchman devices, and no investigational implantable devices will be used. The use of ICE in this study does not modify the approved therapeutic device indication, implantation technique, or standard clinical decision-making.
연구 개요
상세 설명
This study is a prospective clinical feasibility study designed to evaluate the imaging performance, workflow efficiency, and safety of a prototype 4D intracardiac echocardiography (ICE) catheter (ACUNAV Luxera) used in conjunction with the Siemens Premium CV ultrasound system during cardiac interventional procedures.
The ACUNAV Luxera 4D ICE system will be used adjunctively while standard-of-care transesophageal echocardiography (TEE) serves as the primary imaging modality. The study aims to determine whether 4D ICE provides comparable image quality and measurement accuracy to TEE, while maintaining procedural safety and enabling optimization of intracardiac imaging workflows.
연구 유형
등록 (추정된)
단계
- 해당 없음
연락처 및 위치
연구 연락처
- 이름: Walid Saliba, MD
- 전화번호: 216-444-6810
- 이메일: salibaw@ccf.org
연구 장소
-
-
Ohio
-
Cleveland, Ohio, 미국, 44195
- 모병
- Cleveland Clinical Foundation
-
부수사관:
- Oussama Wazni, MD
-
연락하다:
- Walid Saliba, MD
- 전화번호: 216-444-6810
- 이메일: salibaw@ccf.org
-
-
참여기준
자격 기준
공부할 수 있는 나이
- 성인
- 고령자
건강한 자원 봉사자를 받아들입니다
설명
Inclusion Criteria:
- Be aged above 18 for adult participants
- Be scheduled for a LAAO transcatheter procedure with or without concurrent procedures.
- Be able and willing to give informed consent
- (For concomitant cases only) Ablation will occur prior to LAAO implantation
Exclusion Criteria:
Potential subjects must not include:
- Subjects who are currently enrolled in any other investigational clinical studies
- Subjects with any bleeding diathesis
- Subjects with known vascular abnormality rendering right heart catheterization from right or left femoral vein impossible.
- Subjects with known intracardiac thrombus
- Subjects with myocardial infarction within the prior two months
- Subjects with class IV angina or heart failure
- Subjects with sepsis or active infection
- Subjects with immune-deficiency or immune-compromised status
- Subjects with cerebral ischemic event in the prior six months
- Subjects with deep vein thrombosis
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 특수 증상
- 할당: 해당 없음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: 4D ICE Adjunct Imaging
Participants undergoing left atrial appendage occlusion (LAAO) procedures (with or without concomitant electrophysiology procedures) will receive adjunctive imaging using the 4D ICE catheter while TEE remains the primary imaging modality for procedural guidance.
|
Use of investigational 4D ICE catheter for real-time intracardiac imaging during cardiac interventional procedures, performed adjunctively to standard-of-care TEE guidance.
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Feasibility of 4D ICE Imaging Performance
기간: During the Procedure
|
A composite score evaluating the overall technical and clinical adequacy of the 4D Intracardiac Echocardiography (ICE) catheter compared to standard-of-care imaging. The composite score is derived from three individual component assessments: Image Quality: Clarity, resolution, and artifact presence. Anatomical Visualization: Ability to clearly identify target cardiac structures. Procedural Guidance: Effectiveness of real-time images in guiding the procedure. Each component is graded by independent reviewers on a 5-point Likert scale (1=Inadequate, 2=Suboptimal, 3=Adequate, 4=Good, 5=Excellent). The final composite score is calculated as the sum of the three component scores, ranging from a minimum of 3 (worst performance) to a maximum of 15 (best performance). |
During the Procedure
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Image Quality Assessment
기간: During the Procedure
|
Evaluation of 2D, 4D, and Doppler imaging quality using Likert scale (1=Inadequate, 2=Suboptimal, 3=Adequate, 4=Good, 5=Excellent).
|
During the Procedure
|
|
Workflow Efficiency
기간: During the Procedure
|
Procedural workflow metrics including imaging setup and acquisition time
|
During the Procedure
|
|
Measurement Acccuacy
기간: During the Procedure
|
Quantitative comparison of ICE-derived measurements (e.g., LAA sizing) with reference imaging (TEE or CT)
|
During the Procedure
|
|
Safety Outcomes
기간: Up to 45 ± 15 days follow-up
|
Incidence of device- or procedure-related adverse events
|
Up to 45 ± 15 days follow-up
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Walid Saliba, MD, The Cleveland Clinic
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
기타 연구 ID 번호
- 26-552
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
미국에서 제조되어 미국에서 수출되는 제품
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .