Evaluation of an Karajishi Contour Investigational Blood Glucose Monitoring System (ISO/FDA Study)

29. januar 2016 oppdatert av: Ascensia Diabetes Care

User Performance of the Karajishi Contour Blood Glucose Monitoring System (ISO/FDA Study)

The purpose of this study was to determine if untrained persons with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid blood glucose results.

Studieoversikt

Studietype

Intervensjonell

Registrering (Faktiske)

372

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

    • California
      • Escondido, California, Forente stater, 92026
        • AMCR Institute
    • Washington
      • Renton, Washington, Forente stater, 98057
        • Rainier Clinical Research Center

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

18 år og eldre (Voksen, Eldre voksen)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  • Males and females, 18 years of age and older
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion Criteria:

  • Hemophilia or any other bleeding disorder
  • Pregnancy
  • Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM
  • Previously participated in a BG monitor study using the Karajishi Contour BGMS (or used a Bayer Contour meter)
  • Work for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors
  • Work for a competitive medical device company, or having an immediate family member who works for such a company
  • A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk

General enrollment guidelines:

  • Most subjects will be persons with type 1 or type 2 diabetes
  • At least 10% of subjects with diabetes will have type 1 diabetes
  • Approximately 10% of subjects will be persons without diabetes
  • At least 60% of subjects will be younger than age 65

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Støttende omsorg
  • Tildeling: Ikke-randomisert
  • Intervensjonsmodell: Enkeltgruppeoppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: Persons With Diabetes
Untrained subjects WITH Diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System).
Untrained subjects WITH diabetes use the Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
Eksperimentell: Persons With and Without Diabetes
Untrained subjects WITH/WITHOUT Diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System).
Untrained subjects WITH and WITHOUT diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Tidsramme: 1 hour
Untrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma).
1 hour

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Tidsramme: 1 hour
Study staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI venous plasma) and +/-15% (>=100 mg/dL YSI venous plasma).
1 hour
Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff
Tidsramme: 1 hour
Study staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma).
1 hour
Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Tidsramme: 1 hour
Staff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
1 hour
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range
Tidsramme: 1 hour
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range.
1 hour
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range
Tidsramme: 1 hour
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range.
1 hour
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Tidsramme: 1 hour
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range.
1 hour
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Tidsramme: 1 hour
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range.
1 hour
Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Tidsramme: 1 hour
Staff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
1 hour

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Etterforskere

  • Hovedetterforsker: Leslie J Klaff, MD, Rainier Clinical Research Center

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart

1. juni 2014

Primær fullføring (Faktiske)

1. juni 2014

Studiet fullført (Faktiske)

1. juni 2014

Datoer for studieregistrering

Først innsendt

27. mai 2014

Først innsendt som oppfylte QC-kriteriene

27. mai 2014

Først lagt ut (Anslag)

29. mai 2014

Oppdateringer av studieposter

Sist oppdatering lagt ut (Anslag)

29. februar 2016

Siste oppdatering sendt inn som oppfylte QC-kriteriene

29. januar 2016

Sist bekreftet

1. juni 2015

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • GCA-2014-002-01

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .