Non-invasive Perioperative Hb Monitoring in Spinal Surgery

18 april 2018 uppdaterad av: Tehran University of Medical Sciences

Evaluating and Comparing the Accuracy of Non-invasive Hemoglobin Monitoring by Spectrophotometry With and Without Digital Nerve Block in Patients Undergoing Spinal Surgery in Sina Hospital During 2016-2017

The decision to transfuse or not to transfuse blood products is one of the main issues in patients undergoing surgery.

The standard laboratory method of hemoglobin assessment is time-consuming, gives intermittent data, and requires venipuncture which is invasive and painful. In the past decade, the use of non-invasive and faster methods, which allow physicians to measure hemoglobin levels at the patient's bedside, have become widespread. One of the tools that make this possible is Spectrophotometric Hemoglobin.

However, one of the main concerns regarding this method is its accuracy. Since the accuracy of this assessment depends on the extent of perfusion of the organ on which the probe is placed, use of digital nerve blocks proposed to increase its accuracy.

This study aims to evaluate the effects of digital nerve block (with bupivacaine) on the accuracy of non-invasive hemoglobin monitoring by spectrophotometry in patients undergoing spinal surgery.

Studieöversikt

Status

Avslutad

Detaljerad beskrivning

The decision to transfuse or not to transfuse blood products is one of the main issues in healthcare systems. Timely transfusion of blood to a patient bleeding, either due to the surgery or because of the trauma, can prevent tissue hypoxia and ischemia in vital organs. However, transfusion of blood products should be done carefully and cautiously due to the potential complications of transfusions, and also because of the cost of providing blood products 1, 2, 3. Hence, during surgeries, hemoglobin level has a vital role in the diagnosis and management of patients with anemia.

The standard laboratory method of hemoglobin assessment is time-consuming, gives intermittent data, and requires venipuncture which is invasive and painful. In the past decade, the use of non-invasive and faster methods, which allow physicians to measure hemoglobin levels at the patient's bedside, have become widespread. One of the tools that make this possible is Spectrophotometric Hemoglobin (SpHb Masimo Rainbow Co-oximeter). This procedure is applied by a digital probe and allows physicians to monitor hemoglobin levels continuously. 4, 5, 6 However, one of the main concerns regarding this method is its accuracy. Actually, the difference between the measured hemoglobin level and the actual hemoglobin level of the patient, might make it impossible for physicians to decide with sufficient confidence3.

From the time that SpHb has been available in operating rooms, various studies have evaluated its accuracy. While its accuracy has been reported relatively acceptable (mean difference is about 1-1.5 g/dl comparing to the lab data),4 there are still concerns regarding this issue. Furthermore, it is not clear whether this difference would remain the same during the surgery or not. Because blood transfusion is rarely needed when the hemoglobin level of patient is higher than 10g/dl during the surgery. Also, blood transfusion becomes obviously necessary if hemoglobin levels drop to 6g/dl or lower during the surgery. Therefore, the accuracy of the tool in assessing hemoglobin levels at this critical interval (6-10g/dl) is of greater importance.6,7 Experimental studies have confirmed that accuracy could change in different hemoglobin levels5,8, and with the passage of time.9 In addition, it has been reported that after significant blood loss or even following injection of blood, the average difference between SpHb and laboratory might change, as well. 5,8 Since the accuracy of this assessment depends on the extent of perfusion of the organ on which the probe is placed, methods that enhance blood circulation in the organ (including local heating of the organ and use of digital nerve blocks (DNB), proposed to increase its accuracy.10, 11 In a study conducted in 2012 by Miller et al. on 20 patients undergoing spinal surgery, DNB was performed using lidocaine. The differences between SpHb and lab hemoglobin were recorded first after injection, and then at almost hourly intervals. These differences were compared to data from another study without DNB. Thirty seven percent of data in the intervention group and 12% of data in the control group was considered as "very accurate". Results of this study showed that DNB increased perfusion indices and SpHb accuracy.10 Following the previous study, Miller et al. carried out a study in 2014 to compare the effects of DNB with lidocaine and bupivacaine on 12 healthy volunteers. DNB was performed on two identical fingers on both hands of the volunteers. Lidocaine was used on one hand and bupivacaine on the other one. Results of this study showed that both lidocaine and bupivacaine raised the perfusion indices and temperatures of the fingers. The duration and extent of the rise in temperature were more for bupivacaine, and there was a direct correlation between changes in perfusion index and those in SpHb.11 Since, the previous study data had been gathered from healthy individuals, this study aims at investigate the effects of DNB on the accuracy SpHb in patients undergoing surgery.

Studietyp

Interventionell

Inskrivning (Faktisk)

23

Fas

  • Fas 4

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

18 år och äldre (Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Beskrivning

Inclusion Criteria: All patients 18 years of age or older, admitted to Sina hospital to have spinal surgery, whose surgery is not an emergency, are included in the study. Moreover, the patients must be in grade 2 or lower according to the ASA Physical Status Classification, and volunteer to participate in the study.

-

Exclusion Criteria: Patients who are going to have an emergency surgery are excluded from the study. Also patients with grade 3 ASA or higher, or the ones suffering from anatomic or vascular disorders that cause problems in attaching the SpHb sensor, are excluded from the study.

-

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Diagnostisk
  • Tilldelning: Randomiserad
  • Interventionsmodell: Parallellt uppdrag
  • Maskning: Dubbel

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Experimentell: Digital nerve block in left hand
Patients in this group will receive DNB in left hand. DNB will be performed by injecting 2 ml of 0.25% bupivacaine at the base of medial and lateral sides of the finger attached to the sensor (1 ml on each side of the base).
The aim of the study, the way it will be conducted, and the complications and possible risks are explained to them. written informed consent is obtained from them. Then, after induction of anaesthesia, an arterial line will be inserted in order to monitor blood pressure and to take blood samples. During the surgery, two rainbow adult ReSposable sensors will be attached to the third or fourth fingers of both hands, and each of sensors will be attached to a monitor. Patients will be randomly assigned to two groups: group A will receive DNB in left hand and group B will receive DNB in right hand. DNB will be performed by injecting 2 ml of 0.25% bupivacaine at the base of medial and lateral sides of the finger attached to the sensor (1 ml on each side of the base).
Experimentell: Digital nerve block in right hand
Patients in this group will receive DNB in right hand. DNB will be performed by injecting 2 ml of 0.25% bupivacaine at the base of medial and lateral sides of the finger attached to the sensor (1 ml on each side of the base).
The aim of the study, the way it will be conducted, and the complications and possible risks are explained to them. written informed consent is obtained from them. Then, after induction of anaesthesia, an arterial line will be inserted in order to monitor blood pressure and to take blood samples. During the surgery, two rainbow adult ReSposable sensors will be attached to the third or fourth fingers of both hands, and each of sensors will be attached to a monitor. Patients will be randomly assigned to two groups: group A will receive DNB in left hand and group B will receive DNB in right hand. DNB will be performed by injecting 2 ml of 0.25% bupivacaine at the base of medial and lateral sides of the finger attached to the sensor (1 ml on each side of the base).

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
accuracy of SpHb
Tidsram: At the start of the operation, once every 1.5 hours during the operation till its end
in each patient, Sphb values will be documented for both hands. Also, the arterial blood sample (lab Hb) will be sent to the laboratory. The difference between two measure in each hand (with or without DNB) shows the accuracy of Sphb for each hand.
At the start of the operation, once every 1.5 hours during the operation till its end

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Perfusion index (PI)
Tidsram: At the start of the operation, once every 1.5 hours during the operation till its end
In each patient, PI value of hand with DNB will be compared to PI measure in the other hand.
At the start of the operation, once every 1.5 hours during the operation till its end
Pleth Variability Index (PVI)
Tidsram: At the start of the operation, once every 1.5 hours during the operation till its end
In each patient, PVI value of hand with DNB will be compared to PI measure in the other hand.
At the start of the operation, once every 1.5 hours during the operation till its end

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Publikationer och användbara länkar

Den som ansvarar för att lägga in information om studien tillhandahåller frivilligt dessa publikationer. Dessa kan handla om allt som har med studien att göra.

Allmänna publikationer

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

4 januari 2017

Primärt slutförande (Faktisk)

1 oktober 2017

Avslutad studie (Faktisk)

1 januari 2018

Studieregistreringsdatum

Först inskickad

15 september 2016

Först inskickad som uppfyllde QC-kriterierna

15 september 2016

Första postat (Uppskatta)

21 september 2016

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

20 april 2018

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

18 april 2018

Senast verifierad

1 december 2017

Mer information

Termer relaterade till denna studie

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

produkt tillverkad i och exporterad från U.S.A.

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .

Prenumerera