The SATURİSTA Device (SATURİSTA)
Development and Pilot Evaluation of an Artificial Intelligence-Assisted System for Automated Oxygen Flow Regulation: The Saturista Device
This study aims to evaluate the functional performance and feasibility of the Saturista system, a novel device designed for automated oxygen flow regulation based on continuous monitoring of oxygen saturation (SpO₂). The system integrates real-time physiological monitoring, wireless data transmission, and centralized visualization to support timely detection of oxygen desaturation and clinical decision-making.
A pilot feasibility study will be conducted in healthy volunteers using a scenario-based protocol, including baseline monitoring, controlled desaturation, probe disconnection, and signal recovery. The primary objective is to assess the system's ability to generate alarms in response to changes in SpO₂ and to display data in real time. Secondary objectives include evaluation of response time, signal management, and overall system stability.
研究概览
地位
条件
详细说明
The Saturista system is a novel, integrated platform designed to support automated oxygen therapy management through continuous monitoring of oxygen saturation (SpO₂) and real-time decision support. The system combines pulse oximetry-based sensing, wireless data transmission, a cloud-based data infrastructure, and a web-based user interface within a distributed closed-loop control architecture.
In conventional clinical practice, oxygen therapy is typically adjusted manually based on intermittent assessment of oxygen saturation. This approach may delay recognition of physiological deterioration and timely intervention. The Saturista system addresses this limitation by enabling continuous monitoring, automated alarm generation, and centralized visualization of patient data at the nursing station.
The system consists of three main components: (i) a pulse oximetry-based monitoring module, (ii) an embedded device with wireless communication capabilities that regulates oxygen flow via a pneumatic valve mechanism, and (iii) a cloud-based software platform that processes physiological data and provides real-time visualization and alarm management.
This study is designed as a pilot feasibility study conducted in healthy volunteers to evaluate the technical and functional performance of the system. A scenario-based testing protocol will be applied, including baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. These scenarios simulate common clinical monitoring conditions and allow assessment of system responsiveness and reliability.
The primary outcome measures include the system's ability to detect decreases in SpO₂ and generate appropriate alarms, as well as its capacity to display physiological data in real time. Secondary outcomes include system response time, detection of signal loss, recovery performance, and overall operational stability.
The findings of this study are expected to provide preliminary evidence regarding the feasibility and functional performance of the Saturista system and to inform the design of future clinical studies in patient populations.
研究类型
注册 (估计的)
联系人和位置
学习地点
-
-
-
Istanbul、土耳其(türkiye)
- Istinye University
-
-
参与标准
资格标准
适合学习的年龄
- 成人
- 年长者
接受健康志愿者
取样方法
研究人群
The study population consists of healthy adult volunteers recruited for a pilot feasibility evaluation of the Saturista system. Participants are individuals without known respiratory or cardiovascular disease and with normal baseline oxygen saturation levels.
The purpose of selecting a healthy population is to assess the technical performance and functional reliability of the system under controlled conditions before clinical testing in patient populations. All participants will undergo a structured, scenario-based testing protocol, including baseline monitoring, controlled desaturation maneuvers within safe physiological limits, probe disconnection, and signal recovery.
This population allows for standardized assessment of the system's ability to monitor SpO₂, generate alarms, display real-time data, and manage signal-related events without the confounding effects of underlying clinical conditions.
描述
Inclusion Criteria:
- Healthy adult volunteers aged 18 years or older
- Resting oxygen saturation (SpO₂) within the normal range at baseline
- Ability to understand the study procedures and provide written informed consent
- Willingness to participate in all planned testing scenarios
Exclusion Criteria:
- History of chronic respiratory disease, including asthma, chronic obstructive pulmonary disease, or interstitial lung disease
- Known cardiovascular disease or any condition that may affect oxygen saturation or hemodynamic stability
- Current respiratory symptoms, acute infection, or fever at the time of participation
- Anemia, peripheral circulatory disorders, or any condition that may interfere with pulse oximetry measurements
- Skin lesions, deformity, or injury at the intended probe placement site
- Pregnancy
- History of syncope, intolerance to short-duration breath-holding, or any condition that may increase risk during testing
- Inability or unwillingness to comply with study procedures
学习计划
研究是如何设计的?
设计细节
队列和干预
团体/队列 |
|---|
|
Healthy Adult Volunteers (Pilot Feasibility Cohort)
This cohort includes healthy adult volunteers participating in a pilot feasibility study to evaluate the functional performance of the Saturista system. Participants will undergo a structured, scenario-based testing protocol under controlled conditions. The protocol includes baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. The purpose of this cohort is to assess the system's ability to monitor oxygen saturation (SpO₂), generate alarms in response to predefined threshold changes, display data in real time, and manage signal interruption and recovery. No therapeutic intervention is applied, and all procedures are conducted within safe physiological limits under supervision. |
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Real-Time Data Display Performance
大体时间:Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Proportion of measurements successfully displayed on the user interface in real time during all testing scenarios.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Alarm Response Time
大体时间:Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Time (in seconds) between a decrease in SpO₂ below the predefined threshold and activation of the system alarm.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
其他结果措施
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Display Delay
大体时间:Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Time (in seconds) between SpO₂ measurement and visualization on the user interface.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
|
Signal Recovery Performance
大体时间:Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Proportion of scenarios in which the system successfully resumes valid data acquisition and terminates alarms after probe reconnection.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
|
Technical Reliability
大体时间:Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Frequency of technical issues or system failures observed during testing sessions.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
合作者和调查者
出版物和有用的链接
一般刊物
- Hansen EF, Hove JD, Bech CS, Jensen JS, Kallemose T, Vestbo J. Automated oxygen control with O2matic(R) during admission with exacerbation of COPD. Int J Chron Obstruct Pulmon Dis. 2018 Dec 14;13:3997-4003. doi: 10.2147/COPD.S183762. eCollection 2018.
- McMullan J, Hart KW, Barczak C, Lindsell CJ, Branson R. Supplemental Oxygen Requirements of Critically Injured Adults: An Observational Trial. Mil Med. 2016 Aug;181(8):767-72. doi: 10.7205/MILMED-D-15-00356.
- Douin DJ, Rice JD, Xiao M, Beaty L, Guo C, Withers C, Sullivan A, Anderson EL, Cheng AC, Banasiewicz MK, Semler MW, Lloyd BD, Maiga A, Gibbs KW, Stettler GR, Khan A, Sally MB, Wright FL, Aggarwal N, Bebarta VS, Ginde AA; SAVE-O2 AI Investigators. Statistical analysis plan for the Strategy to Avoid Excessive Oxygen using Autonomous Oxygen Titration Intervention (SAVE-O2 AI) trial: protocol. BMJ Open. 2026 Jan 28;16(1):e110739. doi: 10.1136/bmjopen-2025-110739.
研究记录日期
研究主要日期
学习开始 (实际的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
IPD 计划说明
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.