- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT07576946
The SATURİSTA Device (SATURİSTA)
Development and Pilot Evaluation of an Artificial Intelligence-Assisted System for Automated Oxygen Flow Regulation: The Saturista Device
This study aims to evaluate the functional performance and feasibility of the Saturista system, a novel device designed for automated oxygen flow regulation based on continuous monitoring of oxygen saturation (SpO₂). The system integrates real-time physiological monitoring, wireless data transmission, and centralized visualization to support timely detection of oxygen desaturation and clinical decision-making.
A pilot feasibility study will be conducted in healthy volunteers using a scenario-based protocol, including baseline monitoring, controlled desaturation, probe disconnection, and signal recovery. The primary objective is to assess the system's ability to generate alarms in response to changes in SpO₂ and to display data in real time. Secondary objectives include evaluation of response time, signal management, and overall system stability.
Descripción general del estudio
Estado
Condiciones
Descripción detallada
The Saturista system is a novel, integrated platform designed to support automated oxygen therapy management through continuous monitoring of oxygen saturation (SpO₂) and real-time decision support. The system combines pulse oximetry-based sensing, wireless data transmission, a cloud-based data infrastructure, and a web-based user interface within a distributed closed-loop control architecture.
In conventional clinical practice, oxygen therapy is typically adjusted manually based on intermittent assessment of oxygen saturation. This approach may delay recognition of physiological deterioration and timely intervention. The Saturista system addresses this limitation by enabling continuous monitoring, automated alarm generation, and centralized visualization of patient data at the nursing station.
The system consists of three main components: (i) a pulse oximetry-based monitoring module, (ii) an embedded device with wireless communication capabilities that regulates oxygen flow via a pneumatic valve mechanism, and (iii) a cloud-based software platform that processes physiological data and provides real-time visualization and alarm management.
This study is designed as a pilot feasibility study conducted in healthy volunteers to evaluate the technical and functional performance of the system. A scenario-based testing protocol will be applied, including baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. These scenarios simulate common clinical monitoring conditions and allow assessment of system responsiveness and reliability.
The primary outcome measures include the system's ability to detect decreases in SpO₂ and generate appropriate alarms, as well as its capacity to display physiological data in real time. Secondary outcomes include system response time, detection of signal loss, recovery performance, and overall operational stability.
The findings of this study are expected to provide preliminary evidence regarding the feasibility and functional performance of the Saturista system and to inform the design of future clinical studies in patient populations.
Tipo de estudio
Inscripción (Estimado)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
-
Istanbul, Turquía (Türkiye)
- Istinye University
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
- Adulto
- Adulto Mayor
Acepta Voluntarios Saludables
Método de muestreo
Población de estudio
The study population consists of healthy adult volunteers recruited for a pilot feasibility evaluation of the Saturista system. Participants are individuals without known respiratory or cardiovascular disease and with normal baseline oxygen saturation levels.
The purpose of selecting a healthy population is to assess the technical performance and functional reliability of the system under controlled conditions before clinical testing in patient populations. All participants will undergo a structured, scenario-based testing protocol, including baseline monitoring, controlled desaturation maneuvers within safe physiological limits, probe disconnection, and signal recovery.
This population allows for standardized assessment of the system's ability to monitor SpO₂, generate alarms, display real-time data, and manage signal-related events without the confounding effects of underlying clinical conditions.
Descripción
Inclusion Criteria:
- Healthy adult volunteers aged 18 years or older
- Resting oxygen saturation (SpO₂) within the normal range at baseline
- Ability to understand the study procedures and provide written informed consent
- Willingness to participate in all planned testing scenarios
Exclusion Criteria:
- History of chronic respiratory disease, including asthma, chronic obstructive pulmonary disease, or interstitial lung disease
- Known cardiovascular disease or any condition that may affect oxygen saturation or hemodynamic stability
- Current respiratory symptoms, acute infection, or fever at the time of participation
- Anemia, peripheral circulatory disorders, or any condition that may interfere with pulse oximetry measurements
- Skin lesions, deformity, or injury at the intended probe placement site
- Pregnancy
- History of syncope, intolerance to short-duration breath-holding, or any condition that may increase risk during testing
- Inability or unwillingness to comply with study procedures
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
|---|
|
Healthy Adult Volunteers (Pilot Feasibility Cohort)
This cohort includes healthy adult volunteers participating in a pilot feasibility study to evaluate the functional performance of the Saturista system. Participants will undergo a structured, scenario-based testing protocol under controlled conditions. The protocol includes baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. The purpose of this cohort is to assess the system's ability to monitor oxygen saturation (SpO₂), generate alarms in response to predefined threshold changes, display data in real time, and manage signal interruption and recovery. No therapeutic intervention is applied, and all procedures are conducted within safe physiological limits under supervision. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Real-Time Data Display Performance
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Proportion of measurements successfully displayed on the user interface in real time during all testing scenarios.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Alarm Response Time
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Time (in seconds) between a decrease in SpO₂ below the predefined threshold and activation of the system alarm.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Otras medidas de resultado
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Display Delay
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Time (in seconds) between SpO₂ measurement and visualization on the user interface.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
|
Signal Recovery Performance
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Proportion of scenarios in which the system successfully resumes valid data acquisition and terminates alarms after probe reconnection.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
|
Technical Reliability
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Frequency of technical issues or system failures observed during testing sessions.
|
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
|
Colaboradores e Investigadores
Patrocinador
Publicaciones y enlaces útiles
Publicaciones Generales
- Hansen EF, Hove JD, Bech CS, Jensen JS, Kallemose T, Vestbo J. Automated oxygen control with O2matic(R) during admission with exacerbation of COPD. Int J Chron Obstruct Pulmon Dis. 2018 Dec 14;13:3997-4003. doi: 10.2147/COPD.S183762. eCollection 2018.
- McMullan J, Hart KW, Barczak C, Lindsell CJ, Branson R. Supplemental Oxygen Requirements of Critically Injured Adults: An Observational Trial. Mil Med. 2016 Aug;181(8):767-72. doi: 10.7205/MILMED-D-15-00356.
- Douin DJ, Rice JD, Xiao M, Beaty L, Guo C, Withers C, Sullivan A, Anderson EL, Cheng AC, Banasiewicz MK, Semler MW, Lloyd BD, Maiga A, Gibbs KW, Stettler GR, Khan A, Sally MB, Wright FL, Aggarwal N, Bebarta VS, Ginde AA; SAVE-O2 AI Investigators. Statistical analysis plan for the Strategy to Avoid Excessive Oxygen using Autonomous Oxygen Titration Intervention (SAVE-O2 AI) trial: protocol. BMJ Open. 2026 Jan 28;16(1):e110739. doi: 10.1136/bmjopen-2025-110739.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Estimado)
Finalización del estudio (Estimado)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Otros números de identificación del estudio
- 25-134
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Descripción del plan IPD
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .