Esta página se tradujo automáticamente y no se garantiza la precisión de la traducción. por favor refiérase a versión inglesa para un texto fuente.

The SATURİSTA Device (SATURİSTA)

4 de mayo de 2026 actualizado por: Koç University

Development and Pilot Evaluation of an Artificial Intelligence-Assisted System for Automated Oxygen Flow Regulation: The Saturista Device

This study aims to evaluate the functional performance and feasibility of the Saturista system, a novel device designed for automated oxygen flow regulation based on continuous monitoring of oxygen saturation (SpO₂). The system integrates real-time physiological monitoring, wireless data transmission, and centralized visualization to support timely detection of oxygen desaturation and clinical decision-making.

A pilot feasibility study will be conducted in healthy volunteers using a scenario-based protocol, including baseline monitoring, controlled desaturation, probe disconnection, and signal recovery. The primary objective is to assess the system's ability to generate alarms in response to changes in SpO₂ and to display data in real time. Secondary objectives include evaluation of response time, signal management, and overall system stability.

Descripción general del estudio

Estado

Activo, no reclutando

Condiciones

Descripción detallada

The Saturista system is a novel, integrated platform designed to support automated oxygen therapy management through continuous monitoring of oxygen saturation (SpO₂) and real-time decision support. The system combines pulse oximetry-based sensing, wireless data transmission, a cloud-based data infrastructure, and a web-based user interface within a distributed closed-loop control architecture.

In conventional clinical practice, oxygen therapy is typically adjusted manually based on intermittent assessment of oxygen saturation. This approach may delay recognition of physiological deterioration and timely intervention. The Saturista system addresses this limitation by enabling continuous monitoring, automated alarm generation, and centralized visualization of patient data at the nursing station.

The system consists of three main components: (i) a pulse oximetry-based monitoring module, (ii) an embedded device with wireless communication capabilities that regulates oxygen flow via a pneumatic valve mechanism, and (iii) a cloud-based software platform that processes physiological data and provides real-time visualization and alarm management.

This study is designed as a pilot feasibility study conducted in healthy volunteers to evaluate the technical and functional performance of the system. A scenario-based testing protocol will be applied, including baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. These scenarios simulate common clinical monitoring conditions and allow assessment of system responsiveness and reliability.

The primary outcome measures include the system's ability to detect decreases in SpO₂ and generate appropriate alarms, as well as its capacity to display physiological data in real time. Secondary outcomes include system response time, detection of signal loss, recovery performance, and overall operational stability.

The findings of this study are expected to provide preliminary evidence regarding the feasibility and functional performance of the Saturista system and to inform the design of future clinical studies in patient populations.

Tipo de estudio

De observación

Inscripción (Estimado)

7

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

Sí

Método de muestreo

Muestra de probabilidad

Población de estudio

The study population consists of healthy adult volunteers recruited for a pilot feasibility evaluation of the Saturista system. Participants are individuals without known respiratory or cardiovascular disease and with normal baseline oxygen saturation levels.

The purpose of selecting a healthy population is to assess the technical performance and functional reliability of the system under controlled conditions before clinical testing in patient populations. All participants will undergo a structured, scenario-based testing protocol, including baseline monitoring, controlled desaturation maneuvers within safe physiological limits, probe disconnection, and signal recovery.

This population allows for standardized assessment of the system's ability to monitor SpO₂, generate alarms, display real-time data, and manage signal-related events without the confounding effects of underlying clinical conditions.

Descripción

Inclusion Criteria:

  • Healthy adult volunteers aged 18 years or older
  • Resting oxygen saturation (SpO₂) within the normal range at baseline
  • Ability to understand the study procedures and provide written informed consent
  • Willingness to participate in all planned testing scenarios

Exclusion Criteria:

  • History of chronic respiratory disease, including asthma, chronic obstructive pulmonary disease, or interstitial lung disease
  • Known cardiovascular disease or any condition that may affect oxygen saturation or hemodynamic stability
  • Current respiratory symptoms, acute infection, or fever at the time of participation
  • Anemia, peripheral circulatory disorders, or any condition that may interfere with pulse oximetry measurements
  • Skin lesions, deformity, or injury at the intended probe placement site
  • Pregnancy
  • History of syncope, intolerance to short-duration breath-holding, or any condition that may increase risk during testing
  • Inability or unwillingness to comply with study procedures

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

Cohortes e Intervenciones

Grupo / Cohorte
Healthy Adult Volunteers (Pilot Feasibility Cohort)

This cohort includes healthy adult volunteers participating in a pilot feasibility study to evaluate the functional performance of the Saturista system. Participants will undergo a structured, scenario-based testing protocol under controlled conditions. The protocol includes baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery.

The purpose of this cohort is to assess the system's ability to monitor oxygen saturation (SpO₂), generate alarms in response to predefined threshold changes, display data in real time, and manage signal interruption and recovery. No therapeutic intervention is applied, and all procedures are conducted within safe physiological limits under supervision.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Real-Time Data Display Performance
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of measurements successfully displayed on the user interface in real time during all testing scenarios.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Alarm Response Time
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between a decrease in SpO₂ below the predefined threshold and activation of the system alarm.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Otras medidas de resultado

Medida de resultado
Medida Descripción
Periodo de tiempo
Display Delay
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between SpO₂ measurement and visualization on the user interface.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Signal Recovery Performance
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of scenarios in which the system successfully resumes valid data acquisition and terminates alarms after probe reconnection.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Technical Reliability
Periodo de tiempo: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Frequency of technical issues or system failures observed during testing sessions.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Publicaciones y enlaces útiles

La persona responsable de ingresar información sobre el estudio proporciona voluntariamente estas publicaciones. Estos pueden ser sobre cualquier cosa relacionada con el estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

15 de marzo de 2026

Finalización primaria (Estimado)

1 de junio de 2026

Finalización del estudio (Estimado)

1 de diciembre de 2026

Fechas de registro del estudio

Enviado por primera vez

24 de abril de 2026

Primero enviado que cumplió con los criterios de control de calidad

4 de mayo de 2026

Publicado por primera vez (Actual)

8 de mayo de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

8 de mayo de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

4 de mayo de 2026

Última verificación

1 de abril de 2026

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

INDECISO

Descripción del plan IPD

Individual participant data (IPD) sharing has not yet been determined. As this study is a pilot feasibility study involving a small sample of healthy volunteers, no formal data sharing plan has been established at this stage. Data sharing may be considered in future studies following evaluation of study outcomes, ethical considerations, and institutional policies.

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir