The SATURİSTA Device (SATURİSTA)

May 4, 2026 updated by: Koç University

Development and Pilot Evaluation of an Artificial Intelligence-Assisted System for Automated Oxygen Flow Regulation: The Saturista Device

This study aims to evaluate the functional performance and feasibility of the Saturista system, a novel device designed for automated oxygen flow regulation based on continuous monitoring of oxygen saturation (SpO₂). The system integrates real-time physiological monitoring, wireless data transmission, and centralized visualization to support timely detection of oxygen desaturation and clinical decision-making.

A pilot feasibility study will be conducted in healthy volunteers using a scenario-based protocol, including baseline monitoring, controlled desaturation, probe disconnection, and signal recovery. The primary objective is to assess the system's ability to generate alarms in response to changes in SpO₂ and to display data in real time. Secondary objectives include evaluation of response time, signal management, and overall system stability.

Study Overview

Status

Active, not recruiting

Conditions

Detailed Description

The Saturista system is a novel, integrated platform designed to support automated oxygen therapy management through continuous monitoring of oxygen saturation (SpO₂) and real-time decision support. The system combines pulse oximetry-based sensing, wireless data transmission, a cloud-based data infrastructure, and a web-based user interface within a distributed closed-loop control architecture.

In conventional clinical practice, oxygen therapy is typically adjusted manually based on intermittent assessment of oxygen saturation. This approach may delay recognition of physiological deterioration and timely intervention. The Saturista system addresses this limitation by enabling continuous monitoring, automated alarm generation, and centralized visualization of patient data at the nursing station.

The system consists of three main components: (i) a pulse oximetry-based monitoring module, (ii) an embedded device with wireless communication capabilities that regulates oxygen flow via a pneumatic valve mechanism, and (iii) a cloud-based software platform that processes physiological data and provides real-time visualization and alarm management.

This study is designed as a pilot feasibility study conducted in healthy volunteers to evaluate the technical and functional performance of the system. A scenario-based testing protocol will be applied, including baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. These scenarios simulate common clinical monitoring conditions and allow assessment of system responsiveness and reliability.

The primary outcome measures include the system's ability to detect decreases in SpO₂ and generate appropriate alarms, as well as its capacity to display physiological data in real time. Secondary outcomes include system response time, detection of signal loss, recovery performance, and overall operational stability.

The findings of this study are expected to provide preliminary evidence regarding the feasibility and functional performance of the Saturista system and to inform the design of future clinical studies in patient populations.

Study Type

Observational

Enrollment (Estimated)

7

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Sampling Method

Probability Sample

Study Population

The study population consists of healthy adult volunteers recruited for a pilot feasibility evaluation of the Saturista system. Participants are individuals without known respiratory or cardiovascular disease and with normal baseline oxygen saturation levels.

The purpose of selecting a healthy population is to assess the technical performance and functional reliability of the system under controlled conditions before clinical testing in patient populations. All participants will undergo a structured, scenario-based testing protocol, including baseline monitoring, controlled desaturation maneuvers within safe physiological limits, probe disconnection, and signal recovery.

This population allows for standardized assessment of the system's ability to monitor SpO₂, generate alarms, display real-time data, and manage signal-related events without the confounding effects of underlying clinical conditions.

Description

Inclusion Criteria:

  • Healthy adult volunteers aged 18 years or older
  • Resting oxygen saturation (SpO₂) within the normal range at baseline
  • Ability to understand the study procedures and provide written informed consent
  • Willingness to participate in all planned testing scenarios

Exclusion Criteria:

  • History of chronic respiratory disease, including asthma, chronic obstructive pulmonary disease, or interstitial lung disease
  • Known cardiovascular disease or any condition that may affect oxygen saturation or hemodynamic stability
  • Current respiratory symptoms, acute infection, or fever at the time of participation
  • Anemia, peripheral circulatory disorders, or any condition that may interfere with pulse oximetry measurements
  • Skin lesions, deformity, or injury at the intended probe placement site
  • Pregnancy
  • History of syncope, intolerance to short-duration breath-holding, or any condition that may increase risk during testing
  • Inability or unwillingness to comply with study procedures

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Healthy Adult Volunteers (Pilot Feasibility Cohort)

This cohort includes healthy adult volunteers participating in a pilot feasibility study to evaluate the functional performance of the Saturista system. Participants will undergo a structured, scenario-based testing protocol under controlled conditions. The protocol includes baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery.

The purpose of this cohort is to assess the system's ability to monitor oxygen saturation (SpO₂), generate alarms in response to predefined threshold changes, display data in real time, and manage signal interruption and recovery. No therapeutic intervention is applied, and all procedures are conducted within safe physiological limits under supervision.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Real-Time Data Display Performance
Time Frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of measurements successfully displayed on the user interface in real time during all testing scenarios.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Alarm Response Time
Time Frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between a decrease in SpO₂ below the predefined threshold and activation of the system alarm.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Display Delay
Time Frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between SpO₂ measurement and visualization on the user interface.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Signal Recovery Performance
Time Frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of scenarios in which the system successfully resumes valid data acquisition and terminates alarms after probe reconnection.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Technical Reliability
Time Frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Frequency of technical issues or system failures observed during testing sessions.
Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 15, 2026

Primary Completion (Estimated)

June 1, 2026

Study Completion (Estimated)

December 1, 2026

Study Registration Dates

First Submitted

April 24, 2026

First Submitted That Met QC Criteria

May 4, 2026

First Posted (Actual)

May 8, 2026

Study Record Updates

Last Update Posted (Actual)

May 8, 2026

Last Update Submitted That Met QC Criteria

May 4, 2026

Last Verified

April 1, 2026

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

UNDECIDED

IPD Plan Description

Individual participant data (IPD) sharing has not yet been determined. As this study is a pilot feasibility study involving a small sample of healthy volunteers, no formal data sharing plan has been established at this stage. Data sharing may be considered in future studies following evaluation of study outcomes, ethical considerations, and institutional policies.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Oxygen Therapy

Subscribe