Validation of Non-Invasive Hemoglobin Measurement Using Optical Method for Anemia Screening Among Adults in Jakarta (SPECTRO-Hb)

2026年7月13日 更新者:Yohanes Firmansyah, dr, MH, MM、Tarumanagara University

Validation of Non-Invasive Hemoglobin Measurement (Using Optical Method) Compared With Point-of-Care Testing and Hematology Analyzer Among Adult in Jakarta

Anemia remains a major public health concern and is commonly diagnosed through hemoglobin measurement using capillary or venous blood samples. Although laboratory-based and point-of-care testing methods are widely used, they require blood collection, trained personnel, and may cause discomfort to participants. Non-invasive spectrophotometric technology has emerged as a potential alternative for rapid hemoglobin assessment without blood sampling.

This study aims to evaluate the validity and diagnostic performance of a non-invasive hemoglobin using optical method compared with capillary point-of-care testing (POCT) and venous blood hemoglobin measured using an automated hematology analyzer as the reference standard. A total of 150 adults from Kelurahan Kota Bambu, Jakarta, will undergo non-invasive hemoglobin measurement, capillary blood testing, and venous blood testing during a single study visit.

The study will assess agreement between methods, diagnostic accuracy for anemia detection, and factors that may influence measurement performance. Findings from this study are expected to support the development of convenient, community-based anemia screening strategies and provide evidence for future implementation of non-invasive hemoglobin assessment technologies.

研究概览

地位

尚未招聘

条件

详细说明

Background

Anemia affects billions of people worldwide and remains an important public health challenge, particularly in low- and middle-income countries. Undiagnosed anemia is associated with impaired physical performance, reduced productivity, cognitive dysfunction, poor maternal outcomes, and increased morbidity. Early detection of anemia is therefore essential for timely intervention and prevention of complications.

Conventional hemoglobin measurement relies on venous blood analysis using automated hematology analyzers or capillary blood testing using point-of-care devices. While these methods provide reliable results, they require blood collection, trained healthcare personnel, consumable materials, and appropriate laboratory infrastructure. In addition, invasive sampling may cause pain, anxiety, and reluctance to participate in large-scale screening programs.

Recent advances in optical and spectrophotometric technologies have enabled the development of non-invasive methods for estimating hemoglobin concentration without blood collection. These systems utilize light absorption and reflection characteristics of peripheral tissues to estimate hemoglobin levels rapidly and painlessly. Non-invasive approaches may improve participant acceptance, facilitate repeated measurements, and support large-scale community screening initiatives. However, their accuracy and agreement with established laboratory methods remain variable across populations and settings. Therefore, local validation studies are necessary before widespread implementation.

Objectives

Primary Objective

To evaluate the validity and agreement of a non-invasive hemoglobin measurement method compared with capillary POCT hemoglobin measurement and venous blood hemoglobin measured using an automated hematology analyzer.

Secondary Objectives

To determine the correlation between non-invasive hemoglobin measurements and hematology analyzer results.

To assess agreement between non-invasive measurements, POCT results, and hematology analyzer measurements using Bland-Altman analysis.

To evaluate the sensitivity, specificity, positive predictive value, negative predictive value, likelihood ratios, and diagnostic accuracy for anemia detection.

To explore the feasibility of implementing non-invasive hemoglobin measurement in community-based screening programs.

Study Design

This study is a community-based cross-sectional diagnostic validation study.

Study Population

The study population consists of adults aged 18 years and older participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.

Sample Size

A total of 150 participants will be recruited using consecutive sampling. Sample size calculation was based on an expected diagnostic accuracy of 90%, a confidence level of 95%, and a margin of error of 5%, with additional participants included to compensate for incomplete data.

Study Procedures

Each participant will undergo three hemoglobin measurements during a single study visit:

Non-Invasive Hemoglobin Measurement Performed on the thumb using a using optical method device. No blood collection is required. Capillary Hemoglobin Measurement Obtained through finger-prick blood sampling. Measured using a point-of-care testing (POCT) device. Venous Hemoglobin Measurement Venous blood collected by trained healthcare personnel. Analyzed using an automated hematology analyzer as the reference standard.

Demographic information, medical history, smoking status, anthropometric measurements, and factors potentially affecting measurement accuracy will also be collected.

Outcome Measures

Primary Outcome

Agreement between non-invasive hemoglobin measurements and venous hemoglobin measured using a hematology analyzer.

Secondary Outcomes

Correlation coefficients between methods. Mean measurement bias and limits of agreement. Sensitivity and specificity for anemia detection. Positive and negative predictive values. Likelihood ratios. Area under the receiver operating characteristic (ROC) curve. Cohen's kappa coefficient for anemia classification agreement. Statistical Analysis

Descriptive statistics will summarize participant characteristics and hemoglobin measurements. Agreement between methods will be assessed using Bland-Altman analysis and Intraclass Correlation Coefficients (ICC). Correlation analyses will be conducted using Pearson or Spearman tests as appropriate. Diagnostic performance will be evaluated through sensitivity, specificity, predictive values, likelihood ratios, ROC analysis, and area under the curve (AUC). Cohen's kappa coefficient will assess agreement in anemia classification. Subgroup analyses may be performed according to age, sex, and measurement conditions.

Expected Impact

The study is expected to provide evidence regarding the accuracy and screening capability of non-invasive hemoglobin measurement in community settings. Results may contribute to the development of accessible, rapid, and patient-friendly anemia screening programs and support future research involving pediatric and other vulnerable populations.

研究类型

观察性的

注册 (估计的)

200

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

研究联系人备份

学习地点

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 成人
  • 年长者

接受健康志愿者

是的

取样方法

非概率样本

研究人群

Adults aged 18 years and older participating in community-based health screening activities in Jakarta. Participants include both healthy individuals and individuals with potential anemia or other underlying medical conditions. All participants undergo non-invasive hemoglobin measurement, capillary hemoglobin testing using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer during a single study visit.

描述

Inclusion Criteria:

  • Adults aged 18 years and older.
  • Residents or community members participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.
  • Willing to undergo non-invasive hemoglobin measurement.
  • Willing to undergo capillary blood hemoglobin testing using point-of-care testing (POCT).
  • Willing to undergo venous blood collection for hematology analyzer examination.
  • Able to provide written informed consent.

Exclusion Criteria:

  • Active bleeding at the time of examination.
  • Severe acute illness requiring urgent medical attention.
  • Refusal to undergo finger-prick or venous blood sampling.
  • History of severe anxiety, syncope, or adverse reactions related to blood collection procedures.
  • Visible infection, wound, or skin lesion at the measurement site.
  • Severe hand deformity preventing proper non-invasive measurement.
  • Nail polish or artificial nails that cannot be removed before examination.
  • Markedly cold hands or poor peripheral perfusion that may interfere with optical measurements.
  • Incomplete hemoglobin measurement data from any of the study methods.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

队列和干预

团体/队列
Diagnostic Validation Cohort
A cohort of 200 adults recruited from community screening activities. Each participant undergoes three hemoglobin assessments: non-invasive hemoglobin measurement, capillary blood hemoglobin measurement using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer as the reference standard. Data are used to evaluate agreement, diagnostic accuracy, and anemia classification performance of the non-invasive measurement method.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Agreement Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
大体时间:Day 1
Assessment of agreement between non-invasive hemoglobin values and venous hemoglobin values measured using an automated hematology analyzer using Bland-Altman analysis, mean bias, 95% limits of agreement, and intraclass correlation coefficient (ICC).
Day 1
Diagnostic Accuracy for Anemia Detection
大体时间:Day 1
Diagnostic performance of non-invasive hemoglobin measurement for identifying anemia compared with hematology analyzer results, including sensitivity, specificity, positive predictive value, negative predictive value, and overall diagnostic accuracy.
Day 1

次要结果测量

结果测量
措施说明
大体时间
Correlation Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
大体时间:Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with venous hemoglobin concentration measured using a laboratory hematology analyzer. The strength and direction of the relationship will be quantified using Pearson's correlation coefficient for normally distributed data or Spearman's rank correlation coefficient for non-normally distributed data. Correlation coefficients (r), 95% confidence intervals, and p-values will be reported.
Day 1
Agreement Between Non-Invasive Hemoglobin Measurement and Point-of-Care Testing
大体时间:Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with capillary hemoglobin concentration measured using a point-of-care testing (POCT) device. Agreement will be evaluated using Bland-Altman analysis, including mean bias and 95% limits of agreement. Intraclass correlation coefficients (ICC) with 95% confidence intervals will also be calculated to assess measurement reliability between methods.
Day 1
Area Under the ROC Curve (AUC) for Anemia Detection
大体时间:Day 1
The diagnostic performance of the non-invasive hemoglobin measurement device for identifying anemia will be evaluated using receiver operating characteristic (ROC) curve analysis, with venous hemoglobin concentration measured by a hematology analyzer serving as the reference standard. The area under the ROC curve (AUC), optimal cutoff value, sensitivity, specificity, positive predictive value, and negative predictive value will be reported.
Day 1
Agreement in Anemia Classification
大体时间:Day 1
Participants will be classified as anemic or non-anemic based on established hemoglobin thresholds. Agreement between anemia classifications derived from non-invasive hemoglobin measurements and venous hemoglobin measurements obtained using a hematology analyzer will be assessed using Cohen's kappa coefficient. Kappa values and corresponding 95% confidence intervals will be reported.
Day 1

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 研究主任:Alexander Halim Santoso、Tarumanagara University

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年7月11日

初级完成 (估计的)

2027年7月30日

研究完成 (估计的)

2028年7月30日

研究注册日期

首次提交

2026年7月8日

首先提交符合 QC 标准的

2026年7月8日

首次发布 (实际的)

2026年7月14日

研究记录更新

最后更新发布 (实际的)

2026年7月15日

上次提交的符合 QC 标准的更新

2026年7月13日

最后验证

2026年7月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

未定

IPD 计划说明

IPD Sharing Description De-identified individual participant data underlying the results reported in publications may be shared upon reasonable request to the principal investigator, following approval by the institutional ethics committee and execution of an appropriate data-sharing agreement.

Supporting Information Study protocol, statistical analysis plan, and data dictionary.

Time Frame Beginning 6 months after publication and ending 5 years after publication.

Access Criteria Access will be granted to qualified researchers for scientifically sound research proposals, subject to ethical approval and institutional review.