Validation of Non-Invasive Hemoglobin Measurement Using Optical Method for Anemia Screening Among Adults in Jakarta (SPECTRO-Hb)

13 de julio de 2026 actualizado por: Yohanes Firmansyah, dr, MH, MM, Tarumanagara University

Validation of Non-Invasive Hemoglobin Measurement (Using Optical Method) Compared With Point-of-Care Testing and Hematology Analyzer Among Adult in Jakarta

Anemia remains a major public health concern and is commonly diagnosed through hemoglobin measurement using capillary or venous blood samples. Although laboratory-based and point-of-care testing methods are widely used, they require blood collection, trained personnel, and may cause discomfort to participants. Non-invasive spectrophotometric technology has emerged as a potential alternative for rapid hemoglobin assessment without blood sampling.

This study aims to evaluate the validity and diagnostic performance of a non-invasive hemoglobin using optical method compared with capillary point-of-care testing (POCT) and venous blood hemoglobin measured using an automated hematology analyzer as the reference standard. A total of 150 adults from Kelurahan Kota Bambu, Jakarta, will undergo non-invasive hemoglobin measurement, capillary blood testing, and venous blood testing during a single study visit.

The study will assess agreement between methods, diagnostic accuracy for anemia detection, and factors that may influence measurement performance. Findings from this study are expected to support the development of convenient, community-based anemia screening strategies and provide evidence for future implementation of non-invasive hemoglobin assessment technologies.

Descripción general del estudio

Estado

Aún no reclutando

Condiciones

Descripción detallada

Background

Anemia affects billions of people worldwide and remains an important public health challenge, particularly in low- and middle-income countries. Undiagnosed anemia is associated with impaired physical performance, reduced productivity, cognitive dysfunction, poor maternal outcomes, and increased morbidity. Early detection of anemia is therefore essential for timely intervention and prevention of complications.

Conventional hemoglobin measurement relies on venous blood analysis using automated hematology analyzers or capillary blood testing using point-of-care devices. While these methods provide reliable results, they require blood collection, trained healthcare personnel, consumable materials, and appropriate laboratory infrastructure. In addition, invasive sampling may cause pain, anxiety, and reluctance to participate in large-scale screening programs.

Recent advances in optical and spectrophotometric technologies have enabled the development of non-invasive methods for estimating hemoglobin concentration without blood collection. These systems utilize light absorption and reflection characteristics of peripheral tissues to estimate hemoglobin levels rapidly and painlessly. Non-invasive approaches may improve participant acceptance, facilitate repeated measurements, and support large-scale community screening initiatives. However, their accuracy and agreement with established laboratory methods remain variable across populations and settings. Therefore, local validation studies are necessary before widespread implementation.

Objectives

Primary Objective

To evaluate the validity and agreement of a non-invasive hemoglobin measurement method compared with capillary POCT hemoglobin measurement and venous blood hemoglobin measured using an automated hematology analyzer.

Secondary Objectives

To determine the correlation between non-invasive hemoglobin measurements and hematology analyzer results.

To assess agreement between non-invasive measurements, POCT results, and hematology analyzer measurements using Bland-Altman analysis.

To evaluate the sensitivity, specificity, positive predictive value, negative predictive value, likelihood ratios, and diagnostic accuracy for anemia detection.

To explore the feasibility of implementing non-invasive hemoglobin measurement in community-based screening programs.

Study Design

This study is a community-based cross-sectional diagnostic validation study.

Study Population

The study population consists of adults aged 18 years and older participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.

Sample Size

A total of 150 participants will be recruited using consecutive sampling. Sample size calculation was based on an expected diagnostic accuracy of 90%, a confidence level of 95%, and a margin of error of 5%, with additional participants included to compensate for incomplete data.

Study Procedures

Each participant will undergo three hemoglobin measurements during a single study visit:

Non-Invasive Hemoglobin Measurement Performed on the thumb using a using optical method device. No blood collection is required. Capillary Hemoglobin Measurement Obtained through finger-prick blood sampling. Measured using a point-of-care testing (POCT) device. Venous Hemoglobin Measurement Venous blood collected by trained healthcare personnel. Analyzed using an automated hematology analyzer as the reference standard.

Demographic information, medical history, smoking status, anthropometric measurements, and factors potentially affecting measurement accuracy will also be collected.

Outcome Measures

Primary Outcome

Agreement between non-invasive hemoglobin measurements and venous hemoglobin measured using a hematology analyzer.

Secondary Outcomes

Correlation coefficients between methods. Mean measurement bias and limits of agreement. Sensitivity and specificity for anemia detection. Positive and negative predictive values. Likelihood ratios. Area under the receiver operating characteristic (ROC) curve. Cohen's kappa coefficient for anemia classification agreement. Statistical Analysis

Descriptive statistics will summarize participant characteristics and hemoglobin measurements. Agreement between methods will be assessed using Bland-Altman analysis and Intraclass Correlation Coefficients (ICC). Correlation analyses will be conducted using Pearson or Spearman tests as appropriate. Diagnostic performance will be evaluated through sensitivity, specificity, predictive values, likelihood ratios, ROC analysis, and area under the curve (AUC). Cohen's kappa coefficient will assess agreement in anemia classification. Subgroup analyses may be performed according to age, sex, and measurement conditions.

Expected Impact

The study is expected to provide evidence regarding the accuracy and screening capability of non-invasive hemoglobin measurement in community settings. Results may contribute to the development of accessible, rapid, and patient-friendly anemia screening programs and support future research involving pediatric and other vulnerable populations.

Tipo de estudio

De observación

Inscripción (Estimado)

200

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Copia de seguridad de contactos de estudio

Ubicaciones de estudio

    • DKI Jakarta
      • Jakarta, DKI Jakarta, Indonesia, 11420

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

Sí

Método de muestreo

Muestra no probabilística

Población de estudio

Adults aged 18 years and older participating in community-based health screening activities in Jakarta. Participants include both healthy individuals and individuals with potential anemia or other underlying medical conditions. All participants undergo non-invasive hemoglobin measurement, capillary hemoglobin testing using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer during a single study visit.

Descripción

Inclusion Criteria:

  • Adults aged 18 years and older.
  • Residents or community members participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.
  • Willing to undergo non-invasive hemoglobin measurement.
  • Willing to undergo capillary blood hemoglobin testing using point-of-care testing (POCT).
  • Willing to undergo venous blood collection for hematology analyzer examination.
  • Able to provide written informed consent.

Exclusion Criteria:

  • Active bleeding at the time of examination.
  • Severe acute illness requiring urgent medical attention.
  • Refusal to undergo finger-prick or venous blood sampling.
  • History of severe anxiety, syncope, or adverse reactions related to blood collection procedures.
  • Visible infection, wound, or skin lesion at the measurement site.
  • Severe hand deformity preventing proper non-invasive measurement.
  • Nail polish or artificial nails that cannot be removed before examination.
  • Markedly cold hands or poor peripheral perfusion that may interfere with optical measurements.
  • Incomplete hemoglobin measurement data from any of the study methods.

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

Cohortes e Intervenciones

Grupo / Cohorte
Diagnostic Validation Cohort
A cohort of 200 adults recruited from community screening activities. Each participant undergoes three hemoglobin assessments: non-invasive hemoglobin measurement, capillary blood hemoglobin measurement using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer as the reference standard. Data are used to evaluate agreement, diagnostic accuracy, and anemia classification performance of the non-invasive measurement method.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Agreement Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
Periodo de tiempo: Day 1
Assessment of agreement between non-invasive hemoglobin values and venous hemoglobin values measured using an automated hematology analyzer using Bland-Altman analysis, mean bias, 95% limits of agreement, and intraclass correlation coefficient (ICC).
Day 1
Diagnostic Accuracy for Anemia Detection
Periodo de tiempo: Day 1
Diagnostic performance of non-invasive hemoglobin measurement for identifying anemia compared with hematology analyzer results, including sensitivity, specificity, positive predictive value, negative predictive value, and overall diagnostic accuracy.
Day 1

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Correlation Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
Periodo de tiempo: Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with venous hemoglobin concentration measured using a laboratory hematology analyzer. The strength and direction of the relationship will be quantified using Pearson's correlation coefficient for normally distributed data or Spearman's rank correlation coefficient for non-normally distributed data. Correlation coefficients (r), 95% confidence intervals, and p-values will be reported.
Day 1
Agreement Between Non-Invasive Hemoglobin Measurement and Point-of-Care Testing
Periodo de tiempo: Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with capillary hemoglobin concentration measured using a point-of-care testing (POCT) device. Agreement will be evaluated using Bland-Altman analysis, including mean bias and 95% limits of agreement. Intraclass correlation coefficients (ICC) with 95% confidence intervals will also be calculated to assess measurement reliability between methods.
Day 1
Area Under the ROC Curve (AUC) for Anemia Detection
Periodo de tiempo: Day 1
The diagnostic performance of the non-invasive hemoglobin measurement device for identifying anemia will be evaluated using receiver operating characteristic (ROC) curve analysis, with venous hemoglobin concentration measured by a hematology analyzer serving as the reference standard. The area under the ROC curve (AUC), optimal cutoff value, sensitivity, specificity, positive predictive value, and negative predictive value will be reported.
Day 1
Agreement in Anemia Classification
Periodo de tiempo: Day 1
Participants will be classified as anemic or non-anemic based on established hemoglobin thresholds. Agreement between anemia classifications derived from non-invasive hemoglobin measurements and venous hemoglobin measurements obtained using a hematology analyzer will be assessed using Cohen's kappa coefficient. Kappa values and corresponding 95% confidence intervals will be reported.
Day 1

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Investigadores

  • Director de estudio: Alexander Halim Santoso, Tarumanagara University

Publicaciones y enlaces útiles

La persona responsable de ingresar información sobre el estudio proporciona voluntariamente estas publicaciones. Estos pueden ser sobre cualquier cosa relacionada con el estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Estimado)

11 de julio de 2026

Finalización primaria (Estimado)

30 de julio de 2027

Finalización del estudio (Estimado)

30 de julio de 2028

Fechas de registro del estudio

Enviado por primera vez

8 de julio de 2026

Primero enviado que cumplió con los criterios de control de calidad

8 de julio de 2026

Publicado por primera vez (Actual)

14 de julio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

15 de julio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

13 de julio de 2026

Última verificación

1 de julio de 2026

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

INDECISO

Descripción del plan IPD

IPD Sharing Description De-identified individual participant data underlying the results reported in publications may be shared upon reasonable request to the principal investigator, following approval by the institutional ethics committee and execution of an appropriate data-sharing agreement.

Supporting Information Study protocol, statistical analysis plan, and data dictionary.

Time Frame Beginning 6 months after publication and ending 5 years after publication.

Access Criteria Access will be granted to qualified researchers for scientifically sound research proposals, subject to ethical approval and institutional review.

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .