Validation of Non-Invasive Hemoglobin Measurement Using Optical Method for Anemia Screening Among Adults in Jakarta (SPECTRO-Hb)

13 lipca 2026 zaktualizowane przez: Yohanes Firmansyah, dr, MH, MM, Tarumanagara University

Validation of Non-Invasive Hemoglobin Measurement (Using Optical Method) Compared With Point-of-Care Testing and Hematology Analyzer Among Adult in Jakarta

Anemia remains a major public health concern and is commonly diagnosed through hemoglobin measurement using capillary or venous blood samples. Although laboratory-based and point-of-care testing methods are widely used, they require blood collection, trained personnel, and may cause discomfort to participants. Non-invasive spectrophotometric technology has emerged as a potential alternative for rapid hemoglobin assessment without blood sampling.

This study aims to evaluate the validity and diagnostic performance of a non-invasive hemoglobin using optical method compared with capillary point-of-care testing (POCT) and venous blood hemoglobin measured using an automated hematology analyzer as the reference standard. A total of 150 adults from Kelurahan Kota Bambu, Jakarta, will undergo non-invasive hemoglobin measurement, capillary blood testing, and venous blood testing during a single study visit.

The study will assess agreement between methods, diagnostic accuracy for anemia detection, and factors that may influence measurement performance. Findings from this study are expected to support the development of convenient, community-based anemia screening strategies and provide evidence for future implementation of non-invasive hemoglobin assessment technologies.

Przegląd badań

Status

Jeszcze nie rekrutacja

Szczegółowy opis

Background

Anemia affects billions of people worldwide and remains an important public health challenge, particularly in low- and middle-income countries. Undiagnosed anemia is associated with impaired physical performance, reduced productivity, cognitive dysfunction, poor maternal outcomes, and increased morbidity. Early detection of anemia is therefore essential for timely intervention and prevention of complications.

Conventional hemoglobin measurement relies on venous blood analysis using automated hematology analyzers or capillary blood testing using point-of-care devices. While these methods provide reliable results, they require blood collection, trained healthcare personnel, consumable materials, and appropriate laboratory infrastructure. In addition, invasive sampling may cause pain, anxiety, and reluctance to participate in large-scale screening programs.

Recent advances in optical and spectrophotometric technologies have enabled the development of non-invasive methods for estimating hemoglobin concentration without blood collection. These systems utilize light absorption and reflection characteristics of peripheral tissues to estimate hemoglobin levels rapidly and painlessly. Non-invasive approaches may improve participant acceptance, facilitate repeated measurements, and support large-scale community screening initiatives. However, their accuracy and agreement with established laboratory methods remain variable across populations and settings. Therefore, local validation studies are necessary before widespread implementation.

Objectives

Primary Objective

To evaluate the validity and agreement of a non-invasive hemoglobin measurement method compared with capillary POCT hemoglobin measurement and venous blood hemoglobin measured using an automated hematology analyzer.

Secondary Objectives

To determine the correlation between non-invasive hemoglobin measurements and hematology analyzer results.

To assess agreement between non-invasive measurements, POCT results, and hematology analyzer measurements using Bland-Altman analysis.

To evaluate the sensitivity, specificity, positive predictive value, negative predictive value, likelihood ratios, and diagnostic accuracy for anemia detection.

To explore the feasibility of implementing non-invasive hemoglobin measurement in community-based screening programs.

Study Design

This study is a community-based cross-sectional diagnostic validation study.

Study Population

The study population consists of adults aged 18 years and older participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.

Sample Size

A total of 150 participants will be recruited using consecutive sampling. Sample size calculation was based on an expected diagnostic accuracy of 90%, a confidence level of 95%, and a margin of error of 5%, with additional participants included to compensate for incomplete data.

Study Procedures

Each participant will undergo three hemoglobin measurements during a single study visit:

Non-Invasive Hemoglobin Measurement Performed on the thumb using a using optical method device. No blood collection is required. Capillary Hemoglobin Measurement Obtained through finger-prick blood sampling. Measured using a point-of-care testing (POCT) device. Venous Hemoglobin Measurement Venous blood collected by trained healthcare personnel. Analyzed using an automated hematology analyzer as the reference standard.

Demographic information, medical history, smoking status, anthropometric measurements, and factors potentially affecting measurement accuracy will also be collected.

Outcome Measures

Primary Outcome

Agreement between non-invasive hemoglobin measurements and venous hemoglobin measured using a hematology analyzer.

Secondary Outcomes

Correlation coefficients between methods. Mean measurement bias and limits of agreement. Sensitivity and specificity for anemia detection. Positive and negative predictive values. Likelihood ratios. Area under the receiver operating characteristic (ROC) curve. Cohen's kappa coefficient for anemia classification agreement. Statistical Analysis

Descriptive statistics will summarize participant characteristics and hemoglobin measurements. Agreement between methods will be assessed using Bland-Altman analysis and Intraclass Correlation Coefficients (ICC). Correlation analyses will be conducted using Pearson or Spearman tests as appropriate. Diagnostic performance will be evaluated through sensitivity, specificity, predictive values, likelihood ratios, ROC analysis, and area under the curve (AUC). Cohen's kappa coefficient will assess agreement in anemia classification. Subgroup analyses may be performed according to age, sex, and measurement conditions.

Expected Impact

The study is expected to provide evidence regarding the accuracy and screening capability of non-invasive hemoglobin measurement in community settings. Results may contribute to the development of accessible, rapid, and patient-friendly anemia screening programs and support future research involving pediatric and other vulnerable populations.

Typ studiów

Obserwacyjny

Zapisy (Szacowany)

200

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Kontakt w sprawie studiów

Kopia zapasowa kontaktu do badania

Lokalizacje studiów

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

  • Dorosły
  • Starszy dorosły

Akceptuje zdrowych ochotników

Tak

Metoda próbkowania

Próbka bez prawdopodobieństwa

Badana populacja

Adults aged 18 years and older participating in community-based health screening activities in Jakarta. Participants include both healthy individuals and individuals with potential anemia or other underlying medical conditions. All participants undergo non-invasive hemoglobin measurement, capillary hemoglobin testing using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer during a single study visit.

Opis

Inclusion Criteria:

  • Adults aged 18 years and older.
  • Residents or community members participating in community health screening activities in Kelurahan Kota Bambu, Jakarta.
  • Willing to undergo non-invasive hemoglobin measurement.
  • Willing to undergo capillary blood hemoglobin testing using point-of-care testing (POCT).
  • Willing to undergo venous blood collection for hematology analyzer examination.
  • Able to provide written informed consent.

Exclusion Criteria:

  • Active bleeding at the time of examination.
  • Severe acute illness requiring urgent medical attention.
  • Refusal to undergo finger-prick or venous blood sampling.
  • History of severe anxiety, syncope, or adverse reactions related to blood collection procedures.
  • Visible infection, wound, or skin lesion at the measurement site.
  • Severe hand deformity preventing proper non-invasive measurement.
  • Nail polish or artificial nails that cannot be removed before examination.
  • Markedly cold hands or poor peripheral perfusion that may interfere with optical measurements.
  • Incomplete hemoglobin measurement data from any of the study methods.

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

Kohorty i interwencje

Grupa / Kohorta
Diagnostic Validation Cohort
A cohort of 200 adults recruited from community screening activities. Each participant undergoes three hemoglobin assessments: non-invasive hemoglobin measurement, capillary blood hemoglobin measurement using POCT, and venous blood hemoglobin measurement using an automated hematology analyzer as the reference standard. Data are used to evaluate agreement, diagnostic accuracy, and anemia classification performance of the non-invasive measurement method.

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Agreement Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
Ramy czasowe: Day 1
Assessment of agreement between non-invasive hemoglobin values and venous hemoglobin values measured using an automated hematology analyzer using Bland-Altman analysis, mean bias, 95% limits of agreement, and intraclass correlation coefficient (ICC).
Day 1
Diagnostic Accuracy for Anemia Detection
Ramy czasowe: Day 1
Diagnostic performance of non-invasive hemoglobin measurement for identifying anemia compared with hematology analyzer results, including sensitivity, specificity, positive predictive value, negative predictive value, and overall diagnostic accuracy.
Day 1

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Correlation Between Non-Invasive Hemoglobin Measurement and Hematology Analyzer
Ramy czasowe: Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with venous hemoglobin concentration measured using a laboratory hematology analyzer. The strength and direction of the relationship will be quantified using Pearson's correlation coefficient for normally distributed data or Spearman's rank correlation coefficient for non-normally distributed data. Correlation coefficients (r), 95% confidence intervals, and p-values will be reported.
Day 1
Agreement Between Non-Invasive Hemoglobin Measurement and Point-of-Care Testing
Ramy czasowe: Day 1
Hemoglobin concentration obtained from the non-invasive device will be compared with capillary hemoglobin concentration measured using a point-of-care testing (POCT) device. Agreement will be evaluated using Bland-Altman analysis, including mean bias and 95% limits of agreement. Intraclass correlation coefficients (ICC) with 95% confidence intervals will also be calculated to assess measurement reliability between methods.
Day 1
Area Under the ROC Curve (AUC) for Anemia Detection
Ramy czasowe: Day 1
The diagnostic performance of the non-invasive hemoglobin measurement device for identifying anemia will be evaluated using receiver operating characteristic (ROC) curve analysis, with venous hemoglobin concentration measured by a hematology analyzer serving as the reference standard. The area under the ROC curve (AUC), optimal cutoff value, sensitivity, specificity, positive predictive value, and negative predictive value will be reported.
Day 1
Agreement in Anemia Classification
Ramy czasowe: Day 1
Participants will be classified as anemic or non-anemic based on established hemoglobin thresholds. Agreement between anemia classifications derived from non-invasive hemoglobin measurements and venous hemoglobin measurements obtained using a hematology analyzer will be assessed using Cohen's kappa coefficient. Kappa values and corresponding 95% confidence intervals will be reported.
Day 1

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Śledczy

  • Dyrektor Studium: Alexander Halim Santoso, Tarumanagara University

Publikacje i pomocne linki

Osoba odpowiedzialna za wprowadzenie informacji o badaniu dobrowolnie udostępnia te publikacje. Mogą one dotyczyć wszystkiego, co jest związane z badaniem.

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Szacowany)

11 lipca 2026

Zakończenie podstawowe (Szacowany)

30 lipca 2027

Ukończenie studiów (Szacowany)

30 lipca 2028

Daty rejestracji na studia

Pierwszy przesłany

8 lipca 2026

Pierwszy przesłany, który spełnia kryteria kontroli jakości

8 lipca 2026

Pierwszy wysłany (Rzeczywisty)

14 lipca 2026

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

15 lipca 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

13 lipca 2026

Ostatnia weryfikacja

1 lipca 2026

Więcej informacji

Terminy związane z tym badaniem

Plan dla danych uczestnika indywidualnego (IPD)

Planujesz udostępniać dane poszczególnych uczestników (IPD)?

NIEZDECYDOWANY

Opis planu IPD

IPD Sharing Description De-identified individual participant data underlying the results reported in publications may be shared upon reasonable request to the principal investigator, following approval by the institutional ethics committee and execution of an appropriate data-sharing agreement.

Supporting Information Study protocol, statistical analysis plan, and data dictionary.

Time Frame Beginning 6 months after publication and ending 5 years after publication.

Access Criteria Access will be granted to qualified researchers for scientifically sound research proposals, subject to ethical approval and institutional review.

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Nie

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .