Blood'up Device Versus Manual Palpation for Arterial Puncture in Healthy Volunteers: A Pilot Randomized Study Under Normal and Simulated Hypotension

2026年8月10日 更新者:Centre Hospitalier Régional d'Orléans

Arterial puncture is a common procedure in the management of hospitalized patients. It is most often performed at the level of the radial artery and allows arterial blood gas analysis.

The success rate of arterial puncture varies widely in the literature. This procedure is moderately painful and may cause significant discomfort, particularly in cases of repeated attempts. First-attempt success is therefore important to reduce both pain and the risk of local complications associated with multiple punctures.

In routine practice, arterial puncture is performed using manual palpation of the radial artery, typically by a trained nurse. In case of failure, a physician may attempt the procedure, often using ultrasound guidance. However, ultrasound is not routinely used and requires specific training and organizational protocols for nurse use.

In hypotensive conditions, such as shock, the radial pulse may become difficult to palpate, making arterial puncture more challenging when using conventional techniques.

The Blood'up device (Arterya, Hérouville, France) is a non-invasive device currently under development that assists in arterial localization. It is applied to the wrist and projects a light signal indicating the position of the artery. This localization is based on ultrasound technology using piezoelectric sensors integrated into the device.

The aim of this study is to evaluate the effectiveness of the Blood'up device for radial artery localization in healthy volunteers, both under physiological conditions (normal blood pressure) and under simulated hypotensive conditions.

研究概览

研究类型

介入性

注册 (估计的)

32

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

学习地点

      • Orléans、法国、45067
        • Chu Orleans
        • 接触:
          • Mai-Anh NAY, MD
        • 首席研究员:
          • May Anh NAY, MD

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 成人
  • 年长者

接受健康志愿者

是的

描述

Inclusion Criteria:

  • Participant who has agreed to take part in the study
  • No history of cardiovascular disease and a positive Allen test (the Allen test confirms the presence of collateral circulation between the radial and ulnar arteries)
  • Age ≥ 18 years
  • Body mass index (BMI) < 30 kg/m²

Exclusion Criteria:

  • Raynaud's syndrome/disease
  • Damaged or injured skin at the puncture site or at the Blood'Up device contact area
  • Upper limb amputation
  • Contraindication to arterial puncture in one or both upper limbs (e.g., skin lesion, lymph node dissection, etc.)
  • Contraindication to extrinsic arterial compression
  • Therapeutic anticoagulation
  • Known allergy to thermoplastic (ABS)
  • Known allergy to EMLA patch (lidocaine or prilocaine)
  • Pregnant or breastfeeding women

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:治疗
  • 分配:随机化
  • 介入模型:并行分配
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
有源比较器:Arterial puncture using manual palpation

The traditional method for performing radial arterial puncture by manual palpation is as follows:

  • Identification of the radial artery pulsation by palpation by the nurse
  • Puncture performed at the location where the pulse amplitude is perceived as strongest.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

实验性的:Arterial puncture using Blood'Up device
Blood'Up is a portable, reusable, non-sterile Class I medical device that uses ultrasound to detect the radial artery beneath the skin. It projects two puncture points connected by a line onto the skin surface directly above the artery. Unlike ultrasound-guided or palpation-based techniques, the device allows the healthcare professional to keep both hands free, enabling the puncture to be performed under optimal conditions.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

Blood'Up will be used to detect the radial artery beneath the skin.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Success rate of radial arterial puncture performed with and without the Blood'Up device in healthy volunteers with no cardiovascular history
大体时间:5 minutes

Success is defined as successful puncture of the radial artery with confirmation of the arterial nature of the sample by blood gas analysis. Only one puncture (defined as a single skin puncture) will be performed on each wrist.

If the needle is withdrawn without obtaining blood return, this will be considered a failed arterial puncture. Similarly, the absence of blood return within one minute after skin puncture will also be considered a failure.

5 minutes

次要结果测量

结果测量
措施说明
大体时间
Procedure time: from device placement to blood return in the syringe when using the Blood'Up device; and from the start of palpation to blood return in the syringe for manual palpation.
大体时间:30 minutes

Procedure time is defined as the time from the start of device placement to the onset of blood return in the arterial syringe (expected volume: 1 mL).

For the Blood'Up device, the start is defined as the moment the device is placed on the volunteer's skin.

For manual palpation, the start of palpation defines the start of the procedure.

In both cases (Blood'Up and manual palpation), the maximum allowed procedure time is 1 minute. Similarly, the absence of blood return within one minute after skin puncture will be considered a failure.

30 minutes
Pain with and without the Blood'Up device
大体时间:30 minutes
Pain will be assessed immediately after each puncture by asking the participant to rate the maximum pain experienced during the procedure using a visual analog scale (VAS) ranging from 0 (no pain) to 10 (worst imaginable pain).
30 minutes
Number of needle redirections with and without the Blood'Up device.
大体时间:30 minutes
Number of needle redirections during each puncture, defined as any change in needle angulation (self-reported by the nurse performing the procedure). A puncture is defined as a single skin breach by the needle; as long as the needle is not completely withdrawn from the skin, all manipulations are considered part of the same puncture, regardless of the number of redirections.
30 minutes
Immediate complications with and without the Blood'Up device.
大体时间:30 minutes
Complications related to arterial puncture occurring within 5 minutes after the procedure: hematoma at the puncture site, suspected nerve injury (paresthesia), or vasovagal response (pallor, sweating, bradycardia, syncope).
30 minutes

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年9月1日

初级完成 (估计的)

2027年9月1日

研究完成 (估计的)

2027年9月1日

研究注册日期

首次提交

2026年6月26日

首先提交符合 QC 标准的

2026年8月10日

首次发布 (实际的)

2026年8月14日

研究记录更新

最后更新发布 (实际的)

2026年8月14日

上次提交的符合 QC 标准的更新

2026年8月10日

最后验证

2026年8月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅