Blood'up Device Versus Manual Palpation for Arterial Puncture in Healthy Volunteers: A Pilot Randomized Study Under Normal and Simulated Hypotension

10 août 2026 mis à jour par: Centre Hospitalier Régional d'Orléans

Arterial puncture is a common procedure in the management of hospitalized patients. It is most often performed at the level of the radial artery and allows arterial blood gas analysis.

The success rate of arterial puncture varies widely in the literature. This procedure is moderately painful and may cause significant discomfort, particularly in cases of repeated attempts. First-attempt success is therefore important to reduce both pain and the risk of local complications associated with multiple punctures.

In routine practice, arterial puncture is performed using manual palpation of the radial artery, typically by a trained nurse. In case of failure, a physician may attempt the procedure, often using ultrasound guidance. However, ultrasound is not routinely used and requires specific training and organizational protocols for nurse use.

In hypotensive conditions, such as shock, the radial pulse may become difficult to palpate, making arterial puncture more challenging when using conventional techniques.

The Blood'up device (Arterya, Hérouville, France) is a non-invasive device currently under development that assists in arterial localization. It is applied to the wrist and projects a light signal indicating the position of the artery. This localization is based on ultrasound technology using piezoelectric sensors integrated into the device.

The aim of this study is to evaluate the effectiveness of the Blood'up device for radial artery localization in healthy volunteers, both under physiological conditions (normal blood pressure) and under simulated hypotensive conditions.

Aperçu de l'étude

Type d'étude

Interventionnel

Inscription (Estimé)

32

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Lieux d'étude

      • Orléans, France, 45067
        • Chu Orleans
        • Contact:
          • Mai-Anh NAY, MD
        • Chercheur principal:
          • May Anh NAY, MD

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte
  • Adulte plus âgé

Accepte les volontaires sains

Oui

La description

Inclusion Criteria:

  • Participant who has agreed to take part in the study
  • No history of cardiovascular disease and a positive Allen test (the Allen test confirms the presence of collateral circulation between the radial and ulnar arteries)
  • Age ≥ 18 years
  • Body mass index (BMI) < 30 kg/m²

Exclusion Criteria:

  • Raynaud's syndrome/disease
  • Damaged or injured skin at the puncture site or at the Blood'Up device contact area
  • Upper limb amputation
  • Contraindication to arterial puncture in one or both upper limbs (e.g., skin lesion, lymph node dissection, etc.)
  • Contraindication to extrinsic arterial compression
  • Therapeutic anticoagulation
  • Known allergy to thermoplastic (ABS)
  • Known allergy to EMLA patch (lidocaine or prilocaine)
  • Pregnant or breastfeeding women

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation parallèle
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Comparateur actif: Arterial puncture using manual palpation

The traditional method for performing radial arterial puncture by manual palpation is as follows:

  • Identification of the radial artery pulsation by palpation by the nurse
  • Puncture performed at the location where the pulse amplitude is perceived as strongest.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

Expérimental: Arterial puncture using Blood'Up device
Blood'Up is a portable, reusable, non-sterile Class I medical device that uses ultrasound to detect the radial artery beneath the skin. It projects two puncture points connected by a line onto the skin surface directly above the artery. Unlike ultrasound-guided or palpation-based techniques, the device allows the healthcare professional to keep both hands free, enabling the puncture to be performed under optimal conditions.
Each healthy volunteer will undergo arterial blood gas sampling on both wrists using each randomized technique (manual palpation versus the Blood'Up device).

Only for the second cohort with simulated hypotension, which will include 16 healthy volunteers:

Hypotension will be simulated by extrinsic compression using a blood pressure cuff adapted to the volunteer's arm circumference. The (manual) blood pressure cuff will be inflated and maintained to achieve a mean arterial pressure of 60 mmHg (with hypotension defined as a mean arterial pressure below 65 mmHg). Blood pressure will be measured using a non-invasive hemodynamic monitoring device (ClearSight®, Edwards Lifesciences, Mississauga, Canada).

Blood'Up will be used to detect the radial artery beneath the skin.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Success rate of radial arterial puncture performed with and without the Blood'Up device in healthy volunteers with no cardiovascular history
Délai: 5 minutes

Success is defined as successful puncture of the radial artery with confirmation of the arterial nature of the sample by blood gas analysis. Only one puncture (defined as a single skin puncture) will be performed on each wrist.

If the needle is withdrawn without obtaining blood return, this will be considered a failed arterial puncture. Similarly, the absence of blood return within one minute after skin puncture will also be considered a failure.

5 minutes

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Procedure time: from device placement to blood return in the syringe when using the Blood'Up device; and from the start of palpation to blood return in the syringe for manual palpation.
Délai: 30 minutes

Procedure time is defined as the time from the start of device placement to the onset of blood return in the arterial syringe (expected volume: 1 mL).

For the Blood'Up device, the start is defined as the moment the device is placed on the volunteer's skin.

For manual palpation, the start of palpation defines the start of the procedure.

In both cases (Blood'Up and manual palpation), the maximum allowed procedure time is 1 minute. Similarly, the absence of blood return within one minute after skin puncture will be considered a failure.

30 minutes
Pain with and without the Blood'Up device
Délai: 30 minutes
Pain will be assessed immediately after each puncture by asking the participant to rate the maximum pain experienced during the procedure using a visual analog scale (VAS) ranging from 0 (no pain) to 10 (worst imaginable pain).
30 minutes
Number of needle redirections with and without the Blood'Up device.
Délai: 30 minutes
Number of needle redirections during each puncture, defined as any change in needle angulation (self-reported by the nurse performing the procedure). A puncture is defined as a single skin breach by the needle; as long as the needle is not completely withdrawn from the skin, all manipulations are considered part of the same puncture, regardless of the number of redirections.
30 minutes
Immediate complications with and without the Blood'Up device.
Délai: 30 minutes
Complications related to arterial puncture occurring within 5 minutes after the procedure: hematoma at the puncture site, suspected nerve injury (paresthesia), or vasovagal response (pallor, sweating, bradycardia, syncope).
30 minutes

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Publications et liens utiles

La personne responsable de la saisie des informations sur l'étude fournit volontairement ces publications. Il peut s'agir de tout ce qui concerne l'étude.

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Estimé)

1 septembre 2026

Achèvement primaire (Estimé)

1 septembre 2027

Achèvement de l'étude (Estimé)

1 septembre 2027

Dates d'inscription aux études

Première soumission

26 juin 2026

Première soumission répondant aux critères de contrôle qualité

10 août 2026

Première publication (Réel)

14 août 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

14 août 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

10 août 2026

Dernière vérification

1 août 2026

Plus d'information

Termes liés à cette étude

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .