Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)
SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia
SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.
Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.
The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.
This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.
研究概览
研究类型
注册 (估计的)
阶段
- 不适用
联系人和位置
学习联系方式
- 姓名:Eva Tamburro, MD
- 电话号码:33 06 95 16 61 97
- 邮箱:e-tamburro@chu-montpellier.fr
学习地点
-
-
-
Montpellier、法国
- University hospital Montpellier
-
接触:
- Eva Tamburro, MD
- 电话号码:33 06 95 16 61 97
- 邮箱:e-tamburro@chu-montpellier.fr
-
首席研究员:
- Eva Tamburro, MD
-
-
参与标准
资格标准
适合学习的年龄
接受健康志愿者
有资格学习的性别
描述
Inclusion Criteria:
- Adult subject
- Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
- Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.
Exclusion Criteria:
- Subject lacking the cognitive capacity to perform and understand the experimental protocol
- Subject unable to read and/or write
- Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
- Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
- Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
- Pregnant or breastfeeding woman.
- Absence of free and informed consent
- Subject not affiliated with a social security scheme
- Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
- Subject deprived of liberty by judicial or administrative decision
- Persons participating in another study with an exclusion period still ongoing.
学习计划
研究是如何设计的?
设计细节
- 主要用途:其他
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:Instrumented spasticity assessment
Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility. |
Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility. |
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
大体时间:Day 1 and Day 2
|
Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2. |
Day 1 and Day 2
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Difference in total elbow range of motion between the affected and unaffected upper limbs
大体时间:Day 1
|
Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the mean stretch velocity between the affected and unaffected upper limbs
大体时间:Day 1
|
Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the maximum resistance force between the affected and unaffected upper limbs
大体时间:Day 1
|
Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the mean resistance force between the affected and unaffected upper limbs
大体时间:Day 1
|
Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
大体时间:Day 1
|
Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Comparison of Instrumental and Clinical Measures of total elbow range of motion
大体时间:Day 1
|
Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Instrumental and Clinical Measures of stretch velocity
大体时间:Day 1
|
Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
大体时间:Day 1
|
Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
大体时间:Day 1
|
Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
大体时间:Day 1
|
Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds. |
Day 1
|
|
Feasibility of post hoc biomechanical parameter extraction from video recordings
大体时间:Day 1
|
Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed. |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of motion
大体时间:Day 1
|
Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental mean stretch velocity
大体时间:Day 1
|
Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental maximum resistance force
大体时间:Day 1
|
Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of mean resistance force
大体时间:Day 1
|
Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
大体时间:Day 1
|
Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of total range of motion
大体时间:Day 1
|
Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of stretch velocity
大体时间:Day 1
|
Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
大体时间:Day 1
|
Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Comparison of Video-Derived and Clinical Resistance Force Measurements
大体时间:Day 1
|
Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
合作者和调查者
调查人员
- 首席研究员:Eva Tamburro, MD、University Hospital, Montpellier
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.