Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)

15 novembre 2026 mis à jour par: University Hospital, Montpellier

SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia

SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.

Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.

The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.

This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.

Aperçu de l'étude

Statut

Pas encore de recrutement

Type d'étude

Interventionnel

Inscription (Estimé)

40

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Lieux d'étude

      • Montpellier, France
        • University hospital Montpellier
        • Contact:
        • Chercheur principal:
          • Eva Tamburro, MD

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

18 ans et plus (Adulte, Adulte plus âgé)

Accepte les volontaires sains

Non

Sexes éligibles pour l'étude

Tout

La description

Inclusion Criteria:

  • Adult subject
  • Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
  • Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.

Exclusion Criteria:

  • Subject lacking the cognitive capacity to perform and understand the experimental protocol
  • Subject unable to read and/or write
  • Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
  • Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
  • Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
  • Pregnant or breastfeeding woman.
  • Absence of free and informed consent
  • Subject not affiliated with a social security scheme
  • Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
  • Subject deprived of liberty by judicial or administrative decision
  • Persons participating in another study with an exclusion period still ongoing.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Autre
  • Répartition: N / A
  • Modèle interventionnel: Affectation à un seul groupe
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: Instrumented spasticity assessment

Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility.

Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
Délai: Day 1 and Day 2

Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2.

Day 1 and Day 2

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Difference in total elbow range of motion between the affected and unaffected upper limbs
Délai: Day 1

Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean stretch velocity between the affected and unaffected upper limbs
Délai: Day 1

Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the maximum resistance force between the affected and unaffected upper limbs
Délai: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean resistance force between the affected and unaffected upper limbs
Délai: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
Délai: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Comparison of Instrumental and Clinical Measures of total elbow range of motion
Délai: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental and Clinical Measures of stretch velocity
Délai: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
Délai: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
Délai: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
Délai: Day 1

Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds.

Day 1
Feasibility of post hoc biomechanical parameter extraction from video recordings
Délai: Day 1

Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed.

Day 1
Comparison of Video-Derived and Instrumental total range of motion
Délai: Day 1

Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental mean stretch velocity
Délai: Day 1

Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental maximum resistance force
Délai: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of mean resistance force
Délai: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
Délai: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Clinical Measures of total range of motion
Délai: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of stretch velocity
Délai: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
Délai: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Video-Derived and Clinical Resistance Force Measurements
Délai: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Les enquêteurs

  • Chercheur principal: Eva Tamburro, MD, University Hospital, Montpellier

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Estimé)

1 février 2027

Achèvement primaire (Estimé)

1 février 2029

Achèvement de l'étude (Estimé)

1 février 2029

Dates d'inscription aux études

Première soumission

25 août 2026

Première soumission répondant aux critères de contrôle qualité

15 novembre 2026

Première publication (Réel)

20 novembre 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

20 novembre 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

15 novembre 2026

Dernière vérification

1 novembre 2026

Plus d'information

Termes liés à cette étude

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

INDÉCIS

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner