Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)

15 ноября 2026 г. обновлено: University Hospital, Montpellier

SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia

SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.

Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.

The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.

This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.

Обзор исследования

Тип исследования

Интервенционный

Регистрация (Оцененный)

40

Фаза

  • Непригодный

Контакты и местонахождение

В этом разделе приведены контактные данные лиц, проводящих исследование, и информация о том, где проводится это исследование.

Контакты исследования

Места учебы

      • Montpellier, Франция
        • University hospital Montpellier
        • Контакт:
        • Главный следователь:
          • Eva Tamburro, MD

Критерии участия

Исследователи ищут людей, которые соответствуют определенному описанию, называемому критериям приемлемости. Некоторыми примерами этих критериев являются общее состояние здоровья человека или предшествующее лечение.

Критерии приемлемости

Возраст, подходящий для обучения

18 лет и старше (Взрослый, Пожилой взрослый)

Принимает здоровых добровольцев

Нет

Полы, имеющие право на обучение

Все

Описание

Inclusion Criteria:

  • Adult subject
  • Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
  • Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.

Exclusion Criteria:

  • Subject lacking the cognitive capacity to perform and understand the experimental protocol
  • Subject unable to read and/or write
  • Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
  • Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
  • Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
  • Pregnant or breastfeeding woman.
  • Absence of free and informed consent
  • Subject not affiliated with a social security scheme
  • Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
  • Subject deprived of liberty by judicial or administrative decision
  • Persons participating in another study with an exclusion period still ongoing.

Учебный план

В этом разделе представлена ​​подробная информация о плане исследования, в том числе о том, как планируется исследование и что оно измеряет.

Как устроено исследование?

Детали дизайна

  • Основная цель: Другой
  • Распределение: Н/Д
  • Интервенционная модель: Одногрупповое задание
  • Маскировка: Нет (открытая этикетка)

Оружие и интервенции

Группа участников / Армия
Вмешательство/лечение
Экспериментальный: Instrumented spasticity assessment

Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility.

Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility.

Что измеряет исследование?

Первичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
Временное ограничение: Day 1 and Day 2

Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2.

Day 1 and Day 2

Вторичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Difference in total elbow range of motion between the affected and unaffected upper limbs
Временное ограничение: Day 1

Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean stretch velocity between the affected and unaffected upper limbs
Временное ограничение: Day 1

Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the maximum resistance force between the affected and unaffected upper limbs
Временное ограничение: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean resistance force between the affected and unaffected upper limbs
Временное ограничение: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
Временное ограничение: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Comparison of Instrumental and Clinical Measures of total elbow range of motion
Временное ограничение: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental and Clinical Measures of stretch velocity
Временное ограничение: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
Временное ограничение: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
Временное ограничение: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
Временное ограничение: Day 1

Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds.

Day 1
Feasibility of post hoc biomechanical parameter extraction from video recordings
Временное ограничение: Day 1

Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed.

Day 1
Comparison of Video-Derived and Instrumental total range of motion
Временное ограничение: Day 1

Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental mean stretch velocity
Временное ограничение: Day 1

Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental maximum resistance force
Временное ограничение: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of mean resistance force
Временное ограничение: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
Временное ограничение: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Clinical Measures of total range of motion
Временное ограничение: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of stretch velocity
Временное ограничение: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
Временное ограничение: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Video-Derived and Clinical Resistance Force Measurements
Временное ограничение: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1

Соавторы и исследователи

Здесь вы найдете людей и организации, участвующие в этом исследовании.

Следователи

  • Главный следователь: Eva Tamburro, MD, University Hospital, Montpellier

Даты записи исследования

Эти даты отслеживают ход отправки отчетов об исследованиях и сводных результатов на сайт ClinicalTrials.gov. Записи исследований и сообщаемые результаты проверяются Национальной медицинской библиотекой (NLM), чтобы убедиться, что они соответствуют определенным стандартам контроля качества, прежде чем публиковать их на общедоступном веб-сайте.

Изучение основных дат

Начало исследования (Оцененный)

1 февраля 2027 г.

Первичное завершение (Оцененный)

1 февраля 2029 г.

Завершение исследования (Оцененный)

1 февраля 2029 г.

Даты регистрации исследования

Первый отправленный

25 августа 2026 г.

Впервые представлено, что соответствует критериям контроля качества

15 ноября 2026 г.

Первый опубликованный (Действительный)

20 ноября 2026 г.

Обновления учебных записей

Последнее опубликованное обновление (Действительный)

20 ноября 2026 г.

Последнее отправленное обновление, отвечающее критериям контроля качества

15 ноября 2026 г.

Последняя проверка

1 ноября 2026 г.

Дополнительная информация

Термины, связанные с этим исследованием

Планирование данных отдельных участников (IPD)

Планируете делиться данными об отдельных участниках (IPD)?

НЕ РЕШЕНО

Информация о лекарствах и устройствах, исследовательские документы

Изучает лекарственный продукт, регулируемый FDA США.

Нет

Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.

Нет

Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .

Подписаться