- ICH GCP
- Registr klinických studií v USA
- Klinická studie NCT07954232
Safety and Efficacy of TSF Function in HFT750 Device for Infants (TSF-NEST) (TSF-NEST)
A Prospective, Randomized, Crossover, Single-Blind, Multicenter, Sponsor-Initiated Exploratory Clinical Trial for the Evaluation of the Safety and Efficacy of the Target SpO2 Feedback Control by FiO2 (TSF) Function in the HFT750 Device in Infants (TSF-NEST)
The purpose of this study is to evaluate the safety and effectiveness of the automated oxygen control function, called Target SpO2 Feedback control by FiO2 (TSF), in the HFT750 high-flow nasal cannula (HFNC) device for infants admitted to the neonatal intensive care unit (NICU). Maintaining optimal blood oxygen saturation (SpO2) within the target range of 90% to 95% is critical for infants receiving respiratory support, as both too little oxygen (hypoxia) and too much oxygen (hyperoxia) can cause serious complications. Currently, healthcare providers manually adjust the fraction of inspired oxygen (FiO2), which can be challenging due to rapid changes in an infant's breathing. In this crossover study, eligible infants aged 28 days to 1 year requiring HFNC therapy will be randomly assigned to receive both automated FiO2 control (TSF function ON) and standard manual FiO2 control (TSF function OFF) in one of two sequences. Each control mode will be applied for 6 hours, separated by a 30-minute wash-out period, for a total device application time of 12 hours and 30 minutes. Researchers will compare the occurrence of composite safety events (severe hypoxia, severe hyperoxia, or need for urgent medical intervention) and the percentage of time spent within the target SpO2 range between the two modes.
Přehled studie
Postavení
Detailní popis
This is a prospective, randomized, crossover, single-blind, multicenter, sponsor-initiated exploratory clinical trial comparing automated FiO2 adjustment (TSF ON) versus manual FiO2 adjustment (TSF OFF) using the HFT750 universal ventilator (MEKICS Co., Ltd.) in infants requiring high-flow nasal cannula (HFNC) therapy.
Intervention Mechanism and Settings:
The TSF algorithm in the HFT750 continuously monitors the patient's SpO2 via an integrated pulse oximeter module every second and automatically adjusts the delivered FiO2 to maintain the target SpO2 range of 90-95%:
- When SpO2 is below 90%, FiO2 increases by 1% at intervals of 10 to 20 seconds.
- When SpO2 is within 90-95%, FiO2 is maintained.
- When SpO2 is above 95%, FiO2 decreases by 1% at intervals of 30 to 120 seconds.
During both automated and manual periods, the upper FiO2 limit is restricted to baseline FiO2 + 0.2 (maximum 0.5). During the TSF ON period, healthcare providers do not manually intervene in FiO2 adjustments except in emergency situations such as bradycardia or apnea.
Study Procedure:
A total of 20 infants will be stratified by gestational age (<32 weeks vs. ≥32 weeks) and randomized in a 1:1 ratio into Group A (TSF ON followed by TSF OFF) or Group B (TSF OFF followed by TSF ON). Data from the first 30 minutes after switching modes (wash-out period) will be excluded from analysis to prevent carryover effects. Participants will be followed up until the completion of their first feeding immediately following the 12.5-hour device application period.
Typ studie
Zápis (Odhadovaný)
Fáze
- Nelze použít
Kontakty a umístění
Studijní místa
-
-
-
Seoul, Jižní Korea
- Samsung Medical Center
-
Seoul, Jižní Korea
- Seoul National University Hospital
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Popis
Inclusion Criteria:
- Infants aged 28 days to 1 year admitted to the neonatal intensive care unit (NICU)
- Infants requiring high-flow nasal cannula (HFNC) oxygen therapy for respiratory support
- Infants requiring supplemental oxygen (FiO2 > 0.21) to maintain target peripheral oxygen saturation (SpO2) within 90% to 95%
- Infants whose parents or legal guardians have provided written informed consent
Exclusion Criteria:
- Infants with congenital cyanotic heart disease or other major congenital anomalies affecting cardiopulmonary function
- Infants with severe upper airway obstruction or craniofacial anomalies preventing proper nasal cannula application
- Infants requiring immediate invasive mechanical ventilation or non-invasive positive pressure ventilation (NIPPV/CPAP)
- Infants with hemodynamic instability requiring high-dose vasopressors
- Infants participating in another interventional clinical trial
- Infants deemed unsuitable for participation by the principal investigator
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Léčba
- Přidělení: Randomizované
- Intervenční model: Crossover Assignment
- Maskování: Dvojnásobek
Zbraně a zásahy
Skupina účastníků / Arm |
Intervence / Léčba |
|---|---|
|
Experimentální: Group A (TSF ON followed by TSF OFF)
Participants receive automated FiO2 adjustment (TSF function ON) via the HFT750 device for 6 hours, followed by a 30-minute wash-out period, and then receive manual FiO2 adjustment (TSF function OFF) by healthcare providers for 6 hours. |
Automated FiO2 adjustment using the Target SpO2 Feedback (TSF) algorithm on the HFT750 universal ventilator to maintain target SpO2 between 90% and 95% (FiO2 increases by 1% every 10-20 sec if SpO2 <90%; decreases by 1% every 30-120 sec if SpO2 >95%; FiO2 limit = baseline + 0.2, max 0.5) for 6 hours. Standard manual FiO2 adjustment by healthcare providers using the HFT750 universal ventilator with the TSF function turned OFF to maintain target SpO2 between 90% and 95% (FiO2 limit = baseline + 0.2, max 0.5) for 6 hours. |
|
Experimentální: Group B (TSF OFF followed by TSF ON)
Participants receive standard manual FiO2 adjustment (TSF function OFF) by healthcare providers via the HFT750 device for 6 hours, followed by a 30-minute wash-out period, and then receive automated FiO2 adjustment (TSF function ON) for 6 hours. |
Automated FiO2 adjustment using the Target SpO2 Feedback (TSF) algorithm on the HFT750 universal ventilator to maintain target SpO2 between 90% and 95% (FiO2 increases by 1% every 10-20 sec if SpO2 <90%; decreases by 1% every 30-120 sec if SpO2 >95%; FiO2 limit = baseline + 0.2, max 0.5) for 6 hours. Standard manual FiO2 adjustment by healthcare providers using the HFT750 universal ventilator with the TSF function turned OFF to maintain target SpO2 between 90% and 95% (FiO2 limit = baseline + 0.2, max 0.5) for 6 hours. |
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Incidence of Composite Safety Events
Časové okno: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
Number of participants experiencing at least one composite safety event, defined as: (1) severe hypoxemia (SpO2 <85% lasting >60 seconds), (2) severe hyperoxemia (SpO2 >98% lasting >60 seconds while FiO2 >0.21), or (3) requirement for urgent medical intervention (manual FiO2 override due to bradycardia/apnea, or escalation to noninvasive/invasive mechanical ventilation) |
During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Percentage of Time Spent Within the Target SpO2 Range (90% to 95%)
Časové okno: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
Proportion of total valid recording time during which peripheral oxygen saturation (SpO2) is maintained within the target range of 90% to 95%. |
During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
|
Percentage of Time Spent in Hypoxemia (SpO2 <90%) and Hyperoxemia (SpO2 >95%)
Časové okno: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
Proportion of total valid recording time spent below the target range (SpO2 <90%, and severe hypoxemia <85%) and above the target range (SpO2 >95% while FiO2 >0.21, and severe hyperoxemia >98%). |
During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
|
|
Frequency of Manual FiO2 Adjustments and Signs of Respiratory Distress
Časové okno: During each 6-hour intervention period and up to the completion of the first feeding immediately following the 12.5-hour device application period
|
Number of manual FiO2 adjustments performed by healthcare providers, mean delivered FiO2, and clinical assessment of respiratory distress signs (such as tachypnea, chest wall retractions, or nasal flaring) and feeding tolerance. |
During each 6-hour intervention period and up to the completion of the first feeding immediately following the 12.5-hour device application period
|
|
Incidence of Adverse Events (AEs) and Adverse Device Effects (ADEs)
Časové okno: From the start of device application until the completion of the first feeding immediately following the 12.5-hour device application period
|
Number and severity of all adverse events, serious adverse events (SAEs), and device-related adverse effects (e.g., nasal mucosal irritation, device malfunction) observed during the study. |
From the start of device application until the completion of the first feeding immediately following the 12.5-hour device application period
|
Spolupracovníci a vyšetřovatelé
Sponzor
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Primární dokončení (Odhadovaný)
Dokončení studie (Odhadovaný)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
Další relevantní podmínky MeSH
Další identifikační čísla studie
- HFT750-TSF 001
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .