Deze pagina is automatisch vertaald en de nauwkeurigheid van de vertaling kan niet worden gegarandeerd. Raadpleeg de Engelse versie voor een brontekst.

Safety and Efficacy of TSF Function in HFT750 Device for Infants (TSF-NEST) (TSF-NEST)

17 november 2026 bijgewerkt door: MEKICS Co., Ltd.

A Prospective, Randomized, Crossover, Single-Blind, Multicenter, Sponsor-Initiated Exploratory Clinical Trial for the Evaluation of the Safety and Efficacy of the Target SpO2 Feedback Control by FiO2 (TSF) Function in the HFT750 Device in Infants (TSF-NEST)

The purpose of this study is to evaluate the safety and effectiveness of the automated oxygen control function, called Target SpO2 Feedback control by FiO2 (TSF), in the HFT750 high-flow nasal cannula (HFNC) device for infants admitted to the neonatal intensive care unit (NICU). Maintaining optimal blood oxygen saturation (SpO2) within the target range of 90% to 95% is critical for infants receiving respiratory support, as both too little oxygen (hypoxia) and too much oxygen (hyperoxia) can cause serious complications. Currently, healthcare providers manually adjust the fraction of inspired oxygen (FiO2), which can be challenging due to rapid changes in an infant's breathing. In this crossover study, eligible infants aged 28 days to 1 year requiring HFNC therapy will be randomly assigned to receive both automated FiO2 control (TSF function ON) and standard manual FiO2 control (TSF function OFF) in one of two sequences. Each control mode will be applied for 6 hours, separated by a 30-minute wash-out period, for a total device application time of 12 hours and 30 minutes. Researchers will compare the occurrence of composite safety events (severe hypoxia, severe hyperoxia, or need for urgent medical intervention) and the percentage of time spent within the target SpO2 range between the two modes.

Studie Overzicht

Gedetailleerde beschrijving

This is a prospective, randomized, crossover, single-blind, multicenter, sponsor-initiated exploratory clinical trial comparing automated FiO2 adjustment (TSF ON) versus manual FiO2 adjustment (TSF OFF) using the HFT750 universal ventilator (MEKICS Co., Ltd.) in infants requiring high-flow nasal cannula (HFNC) therapy.

Intervention Mechanism and Settings:

The TSF algorithm in the HFT750 continuously monitors the patient's SpO2 via an integrated pulse oximeter module every second and automatically adjusts the delivered FiO2 to maintain the target SpO2 range of 90-95%:

  1. When SpO2 is below 90%, FiO2 increases by 1% at intervals of 10 to 20 seconds.
  2. When SpO2 is within 90-95%, FiO2 is maintained.
  3. When SpO2 is above 95%, FiO2 decreases by 1% at intervals of 30 to 120 seconds.

During both automated and manual periods, the upper FiO2 limit is restricted to baseline FiO2 + 0.2 (maximum 0.5). During the TSF ON period, healthcare providers do not manually intervene in FiO2 adjustments except in emergency situations such as bradycardia or apnea.

Study Procedure:

A total of 20 infants will be stratified by gestational age (<32 weeks vs. ≥32 weeks) and randomized in a 1:1 ratio into Group A (TSF ON followed by TSF OFF) or Group B (TSF OFF followed by TSF ON). Data from the first 30 minutes after switching modes (wash-out period) will be excluded from analysis to prevent carryover effects. Participants will be followed up until the completion of their first feeding immediately following the 12.5-hour device application period.

Studietype

Ingrijpend

Inschrijving (Geschat)

20

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

      • Seoul, Zuid -Korea
        • Samsung Medical Center
      • Seoul, Zuid -Korea
        • Seoul National University Hospital

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

4 weken tot 1 jaar (Kind)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  • Infants aged 28 days to 1 year admitted to the neonatal intensive care unit (NICU)
  • Infants requiring high-flow nasal cannula (HFNC) oxygen therapy for respiratory support
  • Infants requiring supplemental oxygen (FiO2 > 0.21) to maintain target peripheral oxygen saturation (SpO2) within 90% to 95%
  • Infants whose parents or legal guardians have provided written informed consent

Exclusion Criteria:

  • Infants with congenital cyanotic heart disease or other major congenital anomalies affecting cardiopulmonary function
  • Infants with severe upper airway obstruction or craniofacial anomalies preventing proper nasal cannula application
  • Infants requiring immediate invasive mechanical ventilation or non-invasive positive pressure ventilation (NIPPV/CPAP)
  • Infants with hemodynamic instability requiring high-dose vasopressors
  • Infants participating in another interventional clinical trial
  • Infants deemed unsuitable for participation by the principal investigator

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: Gerandomiseerd
  • Interventioneel model: Crossover-opdracht
  • Masker: Dubbele

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Group A (TSF ON followed by TSF OFF)

Participants receive automated FiO2 adjustment (TSF function ON) via the HFT750 device for 6 hours, followed by a 30-minute wash-out period, and then receive manual FiO2 adjustment (TSF function OFF) by healthcare providers for 6 hours.

Automated FiO2 adjustment using the Target SpO2 Feedback (TSF) algorithm on the HFT750 universal ventilator to maintain target SpO2 between 90% and 95% (FiO2 increases by 1% every 10-20 sec if SpO2 <90%; decreases by 1% every 30-120 sec if SpO2 >95%; FiO2 limit = baseline + 0.2, max 0.5) for 6 hours.

Standard manual FiO2 adjustment by healthcare providers using the HFT750 universal ventilator with the TSF function turned OFF to maintain target SpO2 between 90% and 95% (FiO2 limit = baseline + 0.2, max 0.5) for 6 hours.

Experimenteel: Group B (TSF OFF followed by TSF ON)

Participants receive standard manual FiO2 adjustment (TSF function OFF) by healthcare providers via the HFT750 device for 6 hours, followed by a 30-minute wash-out period, and then receive automated FiO2 adjustment (TSF function ON) for 6 hours.

Automated FiO2 adjustment using the Target SpO2 Feedback (TSF) algorithm on the HFT750 universal ventilator to maintain target SpO2 between 90% and 95% (FiO2 increases by 1% every 10-20 sec if SpO2 <90%; decreases by 1% every 30-120 sec if SpO2 >95%; FiO2 limit = baseline + 0.2, max 0.5) for 6 hours.

Standard manual FiO2 adjustment by healthcare providers using the HFT750 universal ventilator with the TSF function turned OFF to maintain target SpO2 between 90% and 95% (FiO2 limit = baseline + 0.2, max 0.5) for 6 hours.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Incidence of Composite Safety Events
Tijdsspanne: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application

Number of participants experiencing at least one composite safety event, defined as: (1) severe hypoxemia (SpO2 <85% lasting >60 seconds), (2) severe hyperoxemia (SpO2 >98% lasting >60 seconds while FiO2 >0.21), or (3) requirement for urgent medical intervention (manual FiO2 override due to bradycardia/apnea, or escalation to noninvasive/invasive mechanical ventilation)

During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Percentage of Time Spent Within the Target SpO2 Range (90% to 95%)
Tijdsspanne: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application

Proportion of total valid recording time during which peripheral oxygen saturation (SpO2) is maintained within the target range of 90% to 95%.

During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
Percentage of Time Spent in Hypoxemia (SpO2 <90%) and Hyperoxemia (SpO2 >95%)
Tijdsspanne: During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application

Proportion of total valid recording time spent below the target range (SpO2 <90%, and severe hypoxemia <85%) and above the target range (SpO2 >95% while FiO2 >0.21, and severe hyperoxemia >98%).

During each 6-hour intervention period (excluding the first 30-minute wash-out period after mode switching), up to 12.5 hours of total device application
Frequency of Manual FiO2 Adjustments and Signs of Respiratory Distress
Tijdsspanne: During each 6-hour intervention period and up to the completion of the first feeding immediately following the 12.5-hour device application period

Number of manual FiO2 adjustments performed by healthcare providers, mean delivered FiO2, and clinical assessment of respiratory distress signs (such as tachypnea, chest wall retractions, or nasal flaring) and feeding tolerance.

During each 6-hour intervention period and up to the completion of the first feeding immediately following the 12.5-hour device application period
Incidence of Adverse Events (AEs) and Adverse Device Effects (ADEs)
Tijdsspanne: From the start of device application until the completion of the first feeding immediately following the 12.5-hour device application period

Number and severity of all adverse events, serious adverse events (SAEs), and device-related adverse effects (e.g., nasal mucosal irritation, device malfunction) observed during the study.

From the start of device application until the completion of the first feeding immediately following the 12.5-hour device application period

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

11 augustus 2026

Primaire voltooiing (Geschat)

31 december 2027

Studie voltooiing (Geschat)

31 december 2027

Studieregistratiedata

Eerst ingediend

17 november 2026

Eerst ingediend dat voldeed aan de QC-criteria

17 november 2026

Eerst geplaatst (Werkelijk)

23 november 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

23 november 2026

Laatste update ingediend die voldeed aan QC-criteria

17 november 2026

Laatst geverifieerd

1 november 2026

Meer informatie

Termen gerelateerd aan deze studie

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren