Ultrasound-Calibrated Pulse Contour Cardiac Output Monitoring: Value and Accuracy in Hemodynamic Care

Accuracy and Decision-Making Value of Ultrasound-Calibrated Pulse Contour Cardiac Output Monitoring for Hemodynamic Management in Critically Ill Patients Compared With PiCCO

This is a single-center prospective clinical study in the intensive care unit (ICU). Thirty adult patients with sepsis requiring hemodynamic monitoring and fluid resuscitation will be enrolled. All participants will undergo two hemodynamic monitoring modalities simultaneously: pulse indicator continuous cardiac output (PiCCO) (reference standard) and arterial pulse contour cardiac output (APCO, ultrasound-calibrated pulse-contour cardiac output). The study aims to evaluate the consistency of clinical decision-making for fluid resuscitation guided by APCO compared with PiCCO, and to verify the accuracy of cardiac output (CO) measured by APCO. No investigational drugs will be administered. A standardized fluid challenge will be performed based on clinical indications. Hemodynamic parameters, laboratory results, and adverse events will be collected throughout the ICU stay.

Studieoversigt

Status

Ikke rekrutterer endnu

Detaljeret beskrivelse

Accurate hemodynamic assessment and fluid management are critical for septic patients in ICU. PiCCO using transpulmonary thermodilution is widely regarded as reference standard for cardiac output measurement, but it has limitations including high cost and potential catheter-related complications. APCO combines transthoracic echocardiography calibration with radial-artery pulse contour analysis, which may provide continuous cardiac output monitoring with lower complication risks.

In this prospective self-controlled study, 30 eligible adult sepsis patients fulfilling Sepsis-3 criteria will be recruited at Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. Each subject will undergo concurrent PiCCO and APCO measurements. Study visits include baseline screening (V1), dynamic monitoring when fluid infusion indication appears (V2), post-resuscitation assessment (V3), and follow-up until PiCCO catheter removal (V4). 500-mL crystalloid fluid challenge over 15 minutes will be conducted for patients with positive fluid responsiveness assessed by passive leg-raising test. Cardiac output (CO), stroke volume index (SVI) and other hemodynamic variables will be recorded synchronously after each fluid challenge. The primary endpoint is consistency of fluid-resuscitation clinical decisions between APCO and PiCCO, assessed by overall agreement rate and Kappa statistics. Secondary endpoints include bias, percentage error, Bland-Altman limits of agreement and correlation of CO values between two techniques. Safety outcomes include incidence of adverse events and device-related complications. Statistical analyses will be performed using SPSS and R software.

Undersøgelsestype

Observationel

Tilmelding (Anslået)

30

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiekontakt

Undersøgelse Kontakt Backup

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

18 år og ældre (Voksen, Ældre voksen)

Tager imod sunde frivillige

Ingen

Køn, der er berettiget til at studere

Alle

Prøveudtagningsmetode

Ikke-sandsynlighedsprøve

Studiebefolkning

Adult critically ill patients with sepsis admitted to the intensive care unit, who require hemodynamic monitoring with PiCCO and APCO for fluid resuscitation and volume assessment.

Beskrivelse

Inclusion Criteria:

  1. Age ≥18 years, admitted to the intensive care unit;
  2. Diagnosed with sepsis according to Sepsis-3 criteria: suspected or confirmed infection with an acute increase in SOFA score ≥2 points from baseline;
  3. Requiring hemodynamic monitoring for volume assessment and fluid resuscitation due to sepsis;
  4. Successfully inserted with central venous catheter and radial arterial catheter, meeting the technical requirements for PiCCO and APCO measurements;
  5. Informed consent signed by the patient or legal surrogate.

Exclusion Criteria:

  1. Severe arrhythmia (e.g., atrial fibrillation, ventricular tachycardia) precluding pulse contour analysis;
  2. Abnormal arterial access site (e.g., severe radial artery disease, arterial malformation) compromising the quality of invasive blood pressure measurement;
  3. Contraindications to ultrasound examination (e.g., severe thoracic deformity, massive pleural effusion, large intrathoracic space-occupying lesion) that prevents acquisition of clear ultrasound images;
  4. Contraindications for PiCCO monitoring (e.g., hemorrhagic disorders, aortic aneurysm, large-vessel vasculitis, pulmonary embolism, during extracorporeal circulation);
  5. Expected ICU stay ≤ 48 hours;
  6. Any other conditions judged unsuitable for enrollment by the investigator.

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Observationsmodeller: COHORT
  • Tidsperspektiver: PROSPECTIVE

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Agreement in fluid resuscitation decision-making between PiCCO and APCO
Tidsramme: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the overall agreement rate and Kappa value for consistency of clinical fluid resuscitation decision-making guided by PiCCO and APCO measurements.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Percentage error of cardiac output (CO) between PiCCO and APCO
Tidsramme: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the percentage error of paired cardiac output measurements obtained by PiCCO transpulmonary thermodilution and APCO monitoring.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Consistency of cardiac output trend changes after hemodynamic intervention
Tidsramme: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the consistency of cardiac output trend changes detected by PiCCO transpulmonary thermodilution and APCO monitoring after hemodynamic intervention.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Agreement of cardiac output evaluated by Bland-Altman analysis
Tidsramme: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Bland-Altman analysis is used to evaluate the agreement of cardiac output measured by PiCCO and APCO.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Pearson correlation of CO between PiCCO and APCO
Tidsramme: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the Pearson correlation coefficient of paired cardiac output measurements obtained by PiCCO transpulmonary thermodilution and APCO monitoring.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Efterforskere

  • Ledende efterforsker: Yuan Yu, MD, PhD, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Anslået)

1. november 2026

Primær færdiggørelse (Anslået)

1. juni 2027

Studieafslutning (Anslået)

1. december 2027

Datoer for studieregistrering

Først indsendt

13. oktober 2026

Først indsendt, der opfyldte QC-kriterier

3. november 2026

Først opslået (Faktiske)

5. november 2026

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

5. november 2026

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

3. november 2026

Sidst verificeret

1. oktober 2026

Mere information

Begreber relateret til denne undersøgelse

Andre undersøgelses-id-numre

  • ICU2026016
  • UHCT-IEC-SOP-016-03-01 (Anden identifikator: Medical Ethics Committee, Un)

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

INGEN

IPD-planbeskrivelse

Individual participant data will not be shared publicly due to patient privacy protection and local institutional data regulations. De-identified aggregate study results may be available upon reasonable request to the principal investigator.

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .