Ultrasound-Calibrated Pulse Contour Cardiac Output Monitoring: Value and Accuracy in Hemodynamic Care

Accuracy and Decision-Making Value of Ultrasound-Calibrated Pulse Contour Cardiac Output Monitoring for Hemodynamic Management in Critically Ill Patients Compared With PiCCO

This is a single-center prospective clinical study in the intensive care unit (ICU). Thirty adult patients with sepsis requiring hemodynamic monitoring and fluid resuscitation will be enrolled. All participants will undergo two hemodynamic monitoring modalities simultaneously: pulse indicator continuous cardiac output (PiCCO) (reference standard) and arterial pulse contour cardiac output (APCO, ultrasound-calibrated pulse-contour cardiac output). The study aims to evaluate the consistency of clinical decision-making for fluid resuscitation guided by APCO compared with PiCCO, and to verify the accuracy of cardiac output (CO) measured by APCO. No investigational drugs will be administered. A standardized fluid challenge will be performed based on clinical indications. Hemodynamic parameters, laboratory results, and adverse events will be collected throughout the ICU stay.

Studie Overzicht

Toestand

Nog niet aan het werven

Gedetailleerde beschrijving

Accurate hemodynamic assessment and fluid management are critical for septic patients in ICU. PiCCO using transpulmonary thermodilution is widely regarded as reference standard for cardiac output measurement, but it has limitations including high cost and potential catheter-related complications. APCO combines transthoracic echocardiography calibration with radial-artery pulse contour analysis, which may provide continuous cardiac output monitoring with lower complication risks.

In this prospective self-controlled study, 30 eligible adult sepsis patients fulfilling Sepsis-3 criteria will be recruited at Union Hospital, Tongji Medical College, Huazhong University of Science and Technology. Each subject will undergo concurrent PiCCO and APCO measurements. Study visits include baseline screening (V1), dynamic monitoring when fluid infusion indication appears (V2), post-resuscitation assessment (V3), and follow-up until PiCCO catheter removal (V4). 500-mL crystalloid fluid challenge over 15 minutes will be conducted for patients with positive fluid responsiveness assessed by passive leg-raising test. Cardiac output (CO), stroke volume index (SVI) and other hemodynamic variables will be recorded synchronously after each fluid challenge. The primary endpoint is consistency of fluid-resuscitation clinical decisions between APCO and PiCCO, assessed by overall agreement rate and Kappa statistics. Secondary endpoints include bias, percentage error, Bland-Altman limits of agreement and correlation of CO values between two techniques. Safety outcomes include incidence of adverse events and device-related complications. Statistical analyses will be performed using SPSS and R software.

Studietype

Observationeel

Inschrijving (Geschat)

30

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Contact Back-up

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Bemonsteringsmethode

Niet-waarschijnlijkheidssteekproef

Studie Bevolking

Adult critically ill patients with sepsis admitted to the intensive care unit, who require hemodynamic monitoring with PiCCO and APCO for fluid resuscitation and volume assessment.

Beschrijving

Inclusion Criteria:

  1. Age ≥18 years, admitted to the intensive care unit;
  2. Diagnosed with sepsis according to Sepsis-3 criteria: suspected or confirmed infection with an acute increase in SOFA score ≥2 points from baseline;
  3. Requiring hemodynamic monitoring for volume assessment and fluid resuscitation due to sepsis;
  4. Successfully inserted with central venous catheter and radial arterial catheter, meeting the technical requirements for PiCCO and APCO measurements;
  5. Informed consent signed by the patient or legal surrogate.

Exclusion Criteria:

  1. Severe arrhythmia (e.g., atrial fibrillation, ventricular tachycardia) precluding pulse contour analysis;
  2. Abnormal arterial access site (e.g., severe radial artery disease, arterial malformation) compromising the quality of invasive blood pressure measurement;
  3. Contraindications to ultrasound examination (e.g., severe thoracic deformity, massive pleural effusion, large intrathoracic space-occupying lesion) that prevents acquisition of clear ultrasound images;
  4. Contraindications for PiCCO monitoring (e.g., hemorrhagic disorders, aortic aneurysm, large-vessel vasculitis, pulmonary embolism, during extracorporeal circulation);
  5. Expected ICU stay ≤ 48 hours;
  6. Any other conditions judged unsuitable for enrollment by the investigator.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Observatiemodellen: COHORT
  • Tijdsperspectieven: PROSPECTIVE

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Agreement in fluid resuscitation decision-making between PiCCO and APCO
Tijdsspanne: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the overall agreement rate and Kappa value for consistency of clinical fluid resuscitation decision-making guided by PiCCO and APCO measurements.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Percentage error of cardiac output (CO) between PiCCO and APCO
Tijdsspanne: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the percentage error of paired cardiac output measurements obtained by PiCCO transpulmonary thermodilution and APCO monitoring.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Consistency of cardiac output trend changes after hemodynamic intervention
Tijdsspanne: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the consistency of cardiac output trend changes detected by PiCCO transpulmonary thermodilution and APCO monitoring after hemodynamic intervention.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Agreement of cardiac output evaluated by Bland-Altman analysis
Tijdsspanne: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Bland-Altman analysis is used to evaluate the agreement of cardiac output measured by PiCCO and APCO.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.
Pearson correlation of CO between PiCCO and APCO
Tijdsspanne: Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

This outcome assesses the Pearson correlation coefficient of paired cardiac output measurements obtained by PiCCO transpulmonary thermodilution and APCO monitoring.

Baseline (enrollment); dynamic monitoring (fluid responsiveness assessment); fluid resuscitation phase (end of early resuscitation); and ≥3 times daily in follow-up until PiCCO catheter removal (expected average 7 days), whichever come first.

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Onderzoekers

  • Hoofdonderzoeker: Yuan Yu, MD, PhD, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Geschat)

1 november 2026

Primaire voltooiing (Geschat)

1 juni 2027

Studie voltooiing (Geschat)

1 december 2027

Studieregistratiedata

Eerst ingediend

13 oktober 2026

Eerst ingediend dat voldeed aan de QC-criteria

3 november 2026

Eerst geplaatst (Werkelijk)

5 november 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

5 november 2026

Laatste update ingediend die voldeed aan QC-criteria

3 november 2026

Laatst geverifieerd

1 oktober 2026

Meer informatie

Termen gerelateerd aan deze studie

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Beschrijving IPD-plan

Individual participant data will not be shared publicly due to patient privacy protection and local institutional data regulations. De-identified aggregate study results may be available upon reasonable request to the principal investigator.

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .