A Comparison Between Ischial-Containment and Sub-Ischial Prosthetic Sockets to Determine Their Effects on Secondary Health Conditions
Tutkimuksen yleiskatsaus
Tila
Yksityiskohtainen kuvaus
The purpose of this study is to investigate and compare the effects of the standard ischial containment (IC) socket to a sub-ischial (SI) socket design on the spinal and pelvic motions and leg length (LLD) discrepancies of transfemoral prosthesis users. Data will be collected from 2 groups of transfemoral prosthesis users (15 SI socket wearers and 15 IC socket wearers) and a group of 15 able-bodied (AB) control participants. Gait and standing analyses will be conducted on each participant using a marker based motion capture system. Prosthesis users will wear their own conventional sockets for the gait and standing analyses. Kinematic and kinetic data will be analyzed for differences between the 2 groups of socket wearers and will also be compared to the control group.
The specific aims and hypotheses of this project are:
Aim 1: To measure dynamic leg length and kinematics of the pelvis and spinal segments in unilateral, transfemoral prosthesis users during gait. Quantitative motion analyses will be performed on two groups of 15 research participants with lower limb amputation using 2 different socket designs (IC and SI) walking over level ground at three self-selected walking speeds (normal, moderately fast, fastest). Data will be compared between socket type and to an AB control group. Hypothesis 1a-IC socket wearers will have a greater anterior pelvic tilt compared to SI socket wearers and AB controls, and they will exhibit considerably greater rotations of the pelvis and spinal segments during walking. Hypothesis 1b-The EMG activity of the lumbar erector spinae for SI socket users will better approximate the patterns of AB controls as compared to IC socket wearers. Hypothesis 1c-The dynamic LLD will be greater for IC socket users compared to SI socket wearers and AB controls.
Aim 2: To measure pelvic orientation and static leg length in unilateral, transfemoral prosthesis users during standing. A multi-segment model of the spine and pelvis will be used to determine neutral spinal, pelvis, and prosthetic alignment during standing for the research and AB control participants. Hypothesis 2a-IC socket wearers will demonstrate greater anterior pelvic tilt during standing compared to SI socket wearers and control participants. Hypothesis 2b-The prosthetic and sound leg lengths for both prosthesis user groups will be within 1.0 cm of each other during standing.
Aim 3: To estimate joint contact forces in the legs and lower back of unilateral, transfemoral prosthesis users and controls during gait. Computational musculoskeletal simulations will be used to comprehensively evaluate the data from Aim 1 to compare the lumbar joint loads between IC and SI socket wearers. Hypothesis 3-IC socket wearers will exhibit greater lumbar joint loads than SI socket wearers.
Opintotyyppi
Ilmoittautuminen (Arvioitu)
Yhteystiedot ja paikat
Opiskeluyhteys
- Nimi: Rebecca Stine, MS
- Puhelinnumero: (312) 503-5726
- Sähköposti: rebecca.stine@va.gov
Tutki yhteystietojen varmuuskopiointi
- Nimi: Steven A Gard, PhD
- Puhelinnumero: (312) 503-5718
- Sähköposti: steven.gard@va.gov
Opiskelupaikat
-
-
Illinois
-
Chicago, Illinois, Yhdysvallat, 60612
- Jesse Brown VA Medical Center, Chicago, IL
-
Ottaa yhteyttä:
- Rebecca Stine, MS
- Puhelinnumero: (312) 503-5726
- Sähköposti: rebecca.stine@va.gov
-
Ottaa yhteyttä:
- Steven A Gard, PhD
- Puhelinnumero: (312) 503-5718
- Sähköposti: steven.gard@va.gov
-
Päätutkija:
- Rebecca Stine, MS
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
- Aikuinen
- Vanhempi Aikuinen
Hyväksyy terveitä vapaaehtoisia
Näytteenottomenetelmä
Tutkimusväestö
Kuvaus
Inclusion Criteria:
For all participants:
- Ability to complete the experimental protocol without undue fatigue or health risks
- Low Back Pain scale assessment (0-10) less than 4 (activities of daily living or quality of life not impacted)
In addition for persons with lower limb amputation:
- Prosthesis user for a minimum of 2 years with their current socket design
- Good fit with either an ischial containment (IC) or sub-ischial (SI) socket
- Good skin integrity on their residual limb upon visual inspection
- Does not require the use of assistive devices to walk
Exclusion Criteria:
- Previously diagnosed with a spine condition (e.g., pinched nerve, bulging disc, etc.)
- Previously diagnosed with an orthopedic or neuromuscular condition that may affect walking or balance (e.g., ankle fusion, Parkinson's disease, stroke, etc.)
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
Kohortit ja interventiot
Ryhmä/Kohortti |
|---|
|
Ischial Containment Socket user
Person with unilateral transfemoral amputation wearing an ischial containment socket
|
|
Sub-Ischial Socket user
Person with unilateral transfemoral amputation wearing a sub-ischial socket
|
|
Able-bodied control
Person without an amputation
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Anterior Pelvic Tilt Angle Between the IC Socket wearers and the SI Socket wearers/Able-Bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Anterior Pelvic Tilt will be calculated for all participants during the walking and standing trials.
Participants will walk along a 10 meter walkway at varying speeds.
Quantitative motion analysis using a marker based motion measurement system will be used to calculate anterior pelvic tilt.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Electromyography (EMG) Activity of the Lumbar Erector Spinae Between the IC Socket wearers and the SI Socket wearers/Able-bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
EMG activity will be assessed using surface electrodes attached bilaterally to the erector spinae at and below T12/L1 and on the external oblique muscle.
Data will be collected during the quantitative motion analysis.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
|
Lumbar Flexion Between the IC Socket wearers and the SI Socket wearers/Able-bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Lumbar flexion will be calculated for all participants during the walking and standing trials.
Participants will walk along a 10 meter walkway at varying speeds.
Quantitative motion analysis using a marker based motion measurement system will be used to calculate anterior pelvic tilt.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
|
Hip Joint Range of Motion (ROM) between the IC Socket wearers and the SI Socket wearers/Able-bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Hip Joint ROM will be calculated for all participants during the walking trials.
Participants will walk along a 10 meter walkway at varying speeds.
Quantitative motion analysis using a marker based motion measurement system will be used to calculate hip joint ROM.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
|
Dynamic Leg Length Discrepancy (LLD) Between the IC Socket wearers and the SI Socket wearers/Able-bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Dynamic LLD will be calculated for all participants during the walking trials.
Participants will walk along a 10 meter walkway at varying speeds.
Quantitative motion analysis using a marker based motion measurement system will be used to calculated dynamic LLD.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
|
Peak Lumbar Joint Loading Between the IC Socket wearers and the SI Socket wearers/Able-bodied Controls
Aikaikkuna: To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Lumbar joint loading during walking will be analyzed by performing musculoskeletal modeling simulations using data collected during the quantitative motion analysis.
|
To be measured on day 1 of the single study visit, within an estimated 3-hour timeframe
|
Yhteistyökumppanit ja tutkijat
Tutkijat
- Päätutkija: Rebecca Stine, MS, Jesse Brown VA Medical Center, Chicago, IL
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Arvioitu)
Ensisijainen valmistuminen (Arvioitu)
Opintojen valmistuminen (Arvioitu)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- RRD2-014-25W
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Yhdysvalloissa valmistettu ja sieltä viety tuote
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .