Cette page a été traduite automatiquement et l'exactitude de la traduction n'est pas garantie. Veuillez vous référer au version anglaise pour un texte source.

A PROSPECTIVE, RANDOMIZED, TWO-ARM, CONTROLLED, SINGLE-BLIND, MULTI-CENTER STUDY TO EVALUATE THE SAFETY AND EFFICACY OF ARTIFIX® DURAL REPAIR SEALANT GRAFT IN CRANIAL AND SPINE SURGERY (NEOFIX STUDY) (NEOFIX)

10 août 2026 mis à jour par: Nurami Medical Ltd
The purpose of the NEOFIX study is to evaluate the safety and efficacy of ArtiFix® dural sealant graft in comparison to usual treatment for Dural repair (membrane that covers the brain and spinal cord) following cranial and spine surgeries. The investigated device will be implanted during the neurosurgical procedure and participants will be evaluated at 16, 36 and 120 days after the procedure.

Aperçu de l'étude

Type d'étude

Interventionnel

Inscription (Estimé)

228

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Sauvegarde des contacts de l'étude

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte
  • Adulte plus âgé

Accepte les volontaires sains

Non

La description

Inclusion Criteria:

Pre-operative inclusion criteria:

  1. Male or female aged ≥18 and <80 years.
  2. Subject is scheduled for an elective cranial or spine surgery with a potential[EG4.1] dural defect that can be repaired by ArtiFix® graft or Control. The maximal estimated dural defect is 25cm2 with an intact dural margins of at least 5 mm surrounding the defect.
  3. Subject has undergone MRI or CT that includes the planned implantation area in the past 6 months before screening.
  4. Surgical wound is expected to be Class I/clean.
  5. Subject understands the study requirements and procedures and provides written Informed Consent before any study-specific tests or procedures. Subject is able and willing to adhere to the follow-up visits and tests.

    Intra-operative Inclusion criteria:

  6. CSF leakage is observed from the dural defect, either spontaneously or during a Valsalva maneuver.
  7. The ArtiFix® or control can cover a maximum dural defect size of 25 cm². The dural defect must be suitable for coverage with at least 5 mm of intact dural surface surrounding the defect.

Exclusion Criteria:

Pre-operative exclusion criteria:

  1. Dural gap is ≤ 2 mm in cranial procedures. In spine procedures the minimal dural gap is not limited.
  2. Subject is unable to undergo MRI after the surgery (only closed bore MRI is allowed; open bore is not allowed).
  3. Subject is an acute cranial or spine traumatic surgical case or diagnosed as nontraumatic hemorrhagic CVA that requires urgent craniectomy.
  4. Subject is operated to repair Chiari malformation, or with a transsphenoidal approach, or that their surgery involves penetration of air cells, of any kind.
  5. Subject had a previous surgery in the same anatomical site.
  6. Subject has known hydrocephalus or other CSF obstruction disorder or has been treated for hydrocephalus with shunt insertion in the past 10 years.
  7. Subject is currently or was diagnosed in the past 5 years with malignancy, excluding basal cell carcinoma or low-grade tumors (i.e., meningioma, ependymoma, schwannoma, hemangioblastoma and other tumors defined as low grade, are eligible to participate).
  8. Intra-operative Valsalva maneuver poses a risk to the subject, as determined by the surgeon or the anesthesiologist.
  9. Subject has a local or systemic infection (e.g. urinary tract infection, active pneumonia) or evidence of any surgical site infection, fever > 38.3℃, or positive blood culture.
  10. Subject has a compromised immune system or an active autoimmune disease, or is on immunosuppressant agents, long term anti-inflammatory biological agents or is on chronic systemic steroid treatment (excluding peri-operative steroid treatment).
  11. Subjected is diagnosed with sepsis, or systemic collagen disease.
  12. Subject at high risk of delayed or failed wound healing due to past treatments or conditions.
  13. Subject has been diagnosed with kidney failure regardless of grade, as evident by serum creatinine levels >1.3 mg/dL.
  14. Subject was diagnosed with liver disease evident by any of the following test results: ALT > 55 U/L, AST > 48 U/L, ALP > 129 U/L, GGT > 61 U/L, Total bilirubin > 1.3 mg/dL, Albumin < 3.5 g/Dl, INR > 1.2, or PT > 15 sec.
  15. Subject has been diagnosed with an uncontrolled diabetes mellitus, as evident by HbA1c > 7%, or suffered recurrent episodes of hypoglycemia in the past 2 months.
  16. Subject had chemotherapy and/or radiotherapy in the past 8 weeks before surgery or is planned to have chemotherapy or radiotherapy less than 8 weeks after surgery.
  17. Subject requires an external ventricular or lumbar CSF drain that must be left in place after surgery.
  18. Subject has congenital or acquired coagulopathy.
  19. The subject was classified as exhibiting drug abuse.
  20. Subject has any condition that, in the Principal Investigator's judgment, may compromise the subject's safety or preclude compliance with the study investigation plan and/or device instructions.
  21. Subject is participating in another clinical trial that tests any other drug or device.
  22. Subject is pregnant or breast-feeding at screening or intends to become pregnant during the study period.
  23. Subject has known history of sensitivity or allergy to any component of the ArtiFix®.

    Intra-operative Exclusion criteria:

  24. Dural gap is ≤ 2 mm in cranial procedures. In spine procedures the minimal dural gap is not limited.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation parallèle
  • Masquage: Double

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Comparateur actif: Traitement standard
Standard-of-care for dural repair in spine or cranial procedures (only FDA cleared graft devices and approved sealants will be used)
Expérimental: ArtiFix the investigational device
ArtiFix® is a sterile, synthetic, single use, adhesive, resorbable sealant graft intended to seal and repair dural defects following cranial or spinal surgeries.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Délai
The absence of device-related CSF leakage or re-operation of the implantation site within 120 days after implantation.
Délai: at 120 days after ArtiFix® or control implantation
at 120 days after ArtiFix® or control implantation
The absence of device-related CSF leakage or re-operation of the implantation site within 120 days after implantation.
Délai: at 120 days after ArtiFix® implantation.
at 120 days after ArtiFix® implantation.

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Les enquêteurs

  • Directeur d'études: Joseph Milbaum, MD, Nurami Medical

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Estimé)

1 janvier 2027

Achèvement primaire (Estimé)

1 janvier 2029

Achèvement de l'étude (Estimé)

1 janvier 2029

Dates d'inscription aux études

Première soumission

10 août 2026

Première soumission répondant aux critères de contrôle qualité

10 août 2026

Première publication (Réel)

14 août 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

14 août 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

10 août 2026

Dernière vérification

1 août 2026

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • PLN-0000265

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

INDÉCIS

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .