Feasibility Study of Navigated Endoscopy for the Placement of High Dose Rate Brachytherapy Applicators in the Esophagus and Lung
調査の概要
詳細な説明
High dose rate brachytherapy is used in a limited number of esophagus and lung patients, often in combination with external beam radiation therapy. The procedure uses an applicator inserted into the lumen with the applicator connected to an automated "afterloader", a device with a radioactive brachytherapy source attached to a thick guide wire that positions the source within the applicator at set positions and times based on a treatment plan. Identification of the ideal applicator position is determined by white light endoscopy. With the endoscope in the lumen, fluoroscopic imaging can visualize the endoscope position. Temporary radio-opaque skin markers are placed on the patient's anterior surface under fluoroscopy imaging so that they align with the endoscope tip. The proximal and the distal ends of the target volume are marked in this manner by the surgeon stopping the endoscope at these positions. The endoscope is removed and the applicator inserted, with the positions of the applicator and "dummy" seeds (i.e. non-radioactive seeds used only for positioning purposes) aligned under fluoroscopy to match the skin markers. The procedure has 2 disadvantages: i. There is extra radiation dose to the clinical staff due to the fluoroscopy used to identify the source position.
ii. Visualization is only in 2D, which is inaccurate since it does not account for separations in the vertical direction and the angle of projection between the applicator and the so patient surface. Furthermore, any movement of the fluoroscopy device during the insertion leads to errors in visualization.
Medical applications of navigated endoscopy continue to grow, especially when used in combination with volumetricimaging for image-guided procedures. This feasibility study is aimed to use navigation endoscopy technology to overcome the disadvantages of the current practice and improve the overall process.
In the proposed protocol, applicator insertion, tracking and recording of the endoscope and applicator positions would enable placement of the applicator using 3D information, with greater accuracy and without extra radiation dose to the clinical staff. The tracking technology would be used to replace the use of radio-opaque skin markers; rather than placing these markers on the patient's skin, the location of the target volume boundaries would be noted by the tracker position within the endoscope. Applicator insertion would also be tracked, with the applicator positioning based on the location of the applicator relative to the previously recorded endoscope positions, including the locations of the distal and proximal ends of the target volume.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Ontario
-
Toronto、Ontario、カナダ、M5G 2M9
- Princess Margaret Cancer Centre
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Age ≥ 18 years
- Primary or metastatic cancer including endoluminal disease of either the lung or the esophagus
- Intention to treat using high dose rate brachytherapy as part of standard radiotherapy.
- Ability to provide written informed consent to participate in the study
Exclusion Criteria:
- Contraindications to standard radiation therapy including pregnancy, lactation, connective tissue disorders, serious co-morbid illness
- Concurrent illness or condition that precludes subject from undergoing endoscopy
- Patient is unable to tolerate an hour and a half long procedure. Standard of care practice is no more than 1 hour. However, with the additional tracking component, the procedure could be at most 1.5 hours long.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:デバイスの実現可能性
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
proportion of navigated endoscopy procedures that were successfully completed
時間枠:18 months
|
To assess clinical feasibility of inserting intraluminal (esophagus or lung) HDR applicators using navigation technologies for tracking endoscope and applicator insertion. This study is designed to assess the workflow when using the EM tracking technology and the accuracy of the procedure will be measured by comparison with current standard of care. |
18 months
|
|
Positional accuracy
時間枠:day 1
|
To compare the positional accuracy of the navigation procedure versus the current standard of care using 2D fluoroscopy using analysis of tracking data.
All tracking data will be recorded for post-procedure analysis and comparison with the fluoroscopic imaging performed during current standard of care
|
day 1
|
|
Optimize the clinical protocol using this technology
時間枠:6 months
|
The procedure will be assessed in detail (based on time and workflow) after three patients.
Participants in the review of the procedure will include surgeon, radiation oncologist, brachy-therapists and physicists.
Workflow improvements will be implemented at this time.
|
6 months
|
協力者と研究者
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。