Feasibility Evaluation of an Artificial Pancreas With Glucose Prediction Trust Index
Feasibility Evaluation of an Artificial Pancreas With an Enhanced Model Predictive Control (eMPC) Controller and a Health Monitoring System (HMS) With a Glucose Prediction Trust Index
調査の概要
詳細な説明
This is a feasibility study of an artificial pancreas (AP) system with our previously validated target eMPC and HMS algorithms, with the addition of a trust index of the predicted glucose value integrated into the portable Artificial Pancreas System (pAPS). The system will be evaluated on up to 20 subjects, with the goal of 10 subjects completing a 48-hour closed-loop session at one clinical site (William Sansum Diabetes Center). During the session subjects will bolus for all meals and snacks and perform a 45-minute walking session.
The purpose of this pilot study is to establish that closed-loop insulin delivery with a target enchanted model predictive control (eMPC)/Health Monitoring System (HMS) algorithms with a trust index of the predicted glucose value is safe and effective, to analyze and learn to improve upon the accuracy of the predicted glucose values, and to collect efficacy data to inform a future larger study.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
California
-
Santa Barbara、California、アメリカ、93105
- William Sansum Diabetes Center
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Age ≥ 18 and ≤ 75 years at the time of screening.
- Clinical diagnosis of type 1 diabetes for at least one year.
- Has been using an insulin pump for at least 6 months at the time of screening.
- HbA1c < 10%, as performed by point of care or central laboratory testing. A1c will be assessed at the screening visit, or if already completed within 2 months of the screening visit, the prior lab value may be used in lieu of repeating this assessment.
- Serum creatinine of < 2.0 mg/dL. Serum creatinine will be assessed at the screening visit, or if already completed within 2 months of the screening visit, the prior lab value may be used in lieu of repeating this assessment.
- Bolus for all meals and snacks that contain ≥ 5 grams of carbohydrate.
- Willing to perform at least 7 fingerstick blood glucose tests a day.
- Willing to refrain from taking acetaminophen products for the duration of the clinical trial. If acetaminophen is taken, subject is to avoid making any insulin dosing decisions based on CGM for at least 12 hours.
- Willing to abide by the study protocol and use study-provided devices, including the Omnipod, Dexcom CGM, glucometer, ketone meter and pAPS tablet.
Exclusion Criteria:
- Pregnancy
- One or more episodes of hypoglycemia requiring an emergency room visit or hospitalization in the past 6 months.
- One or more episodes of hyperglycemia requiring an emergency room visit or hospitalization in the past 6 months.
- Known unstable cardiac disease or untreated cardiac disease, as revealed by history or physical examination.
- Dermatological conditions that would preclude wearing a CGM sensor or Pod.
- One or more seizures in the past year.
- Screening A1c ≥ 10% or serum creatinine ≥ 2.0 mg/dL.
- Any condition that could interfere with participating in the trial, based on investigator judgment.
- Concurrent use of any non-insulin glucose-lowering agent (including GLP-1 agonists, Symlin, DPP-4 inhibitors, SGLT-2 inhibitors, biguanides, sulfonylureas and naturaceuticals), defined as using any of these medications within 30 days of the screening visit or during the study.
- Participation in another pharmaceutical or device trial at the time of enrollment or during the study.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:処理
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Artificial Pancreas
The artificial pancreas device will employ its enhanced Model Predictive Control (MPC) algorithm with a target glucose level of 110 mg/dL with a trust index for MPC-predicted glucose values, weighing future glucose predictions and only acting on predictions with higher weight in the trust index.
The Health Monitoring System algorithm uses the same CGM data as the MPC control algorithm but utilizes a separate algorithm for trending and predictions of future glucose values.
Using a redundant and independent algorithm is an important safety feature of the overall AP device.
|
Artificial Pancreas Device
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Percent time within the target glucose range of 70-180 mg/dl overall as assessed by continuous glucose monitoring (CGM).
時間枠:48 hours
|
The primary endpoint for this pilot study will be time within the target glucose range of 70-180 mg/dl overall as assessed by continuous glucose monitoring (CGM), during closed-loop use, determining if the combination of eMPC and the assessment of trust index can provide safe and effective glucose control.
|
48 hours
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Assessment of the trust index predictions vs. standard eMPC glucose predictions.
時間枠:48 hours
|
During closed-loop use, assessment of the trust index predictions vs. standard eMPC glucose predictions (frequency of moving away from nominal weights of predicted glucose).
|
48 hours
|
|
Frequency of hypoglycemia during closed-loop use.
時間枠:48 hours
|
Frequency of hypoglycemia during closed-loop use, defined as percent time glucose is below predefined thresholds of 70 mg/dl, 60 mg/dl and 50 mg/dL.
|
48 hours
|
|
Frequency of hyperglycemia during closed-loop use.
時間枠:48 hours
|
Frequency of hyperglycemia during closed-loop use, defined as percent time glucose is above predefined thresholds of 180 mg/dl, 250 mg/dl, and 300 mg/dL.
|
48 hours
|
|
Percent time within the target glucose range of 70-140 mg/dl during closed-loop use.
時間枠:48 hours
|
Percent time within the target glucose range of 70-140 mg/dl overall as assessed by continuous glucose monitoring (CGM), during closed-loop use.
|
48 hours
|
|
Percent time glucose is within range 70-150 mg/dl postprandial within 5 hours following meals, during closed-loop use.
時間枠:48 hours
|
Percent time glucose is within range 70-150 mg/dl postprandial within 5 hours following meals, during closed-loop use.
|
48 hours
|
|
Standard Deviation and Coefficient of Variation of glucose values
時間枠:48 hours
|
To measure glycemic variability, the standard deviation and coefficient of variation of glucose values (as measured by continuous glucose monitoring) will be determined during the 48 hours of closed-loop use.
|
48 hours
|
|
Markers of hypo- and hyperglycemia
時間枠:48 hours
|
Markers of hypo- and hyperglycemia, to include LBGI (Low blood glucose index) and HBGI (High blood glucose index), during closed-loop use.
|
48 hours
|
|
Insulin Doses Given
時間枠:48 hours
|
Change in total daily doses and post prandial insulin doses given during the 48 hours of closed-loop use.
|
48 hours
|
|
Number of treatments given for hypoglycemia (defined as carbohydrates given for treatment of a documented or suspected low blood sugar), during the 48 hours of closed-loop use.
時間枠:48 hours
|
Treatments for hypoglycemia given during the 48 hours of closed-loop use.
Hypoglycemia treatments may be given in response to alerts from the HMS, if blood glucose < 70 mg/dL, or at the request of the subject or investigator.
|
48 hours
|
|
Number of alerts given by the HMS to prevent hypoglycemia
時間枠:48 hours
|
Number of alerts given by the HMS to prevent hypoglycemia during closed-loop use.
|
48 hours
|
|
Failure analysis of the devices/connectivity issues that may occur (# disconnects and device restarts).
時間枠:48 hours
|
Failure analysis of the devices/connectivity issues that may occur during closed-loop use.
This includes number of CGM communication losses with 3 or more missed points, and number of times the entire system required a restart.
|
48 hours
|
協力者と研究者
捜査官
- 主任研究者:Eyal Dassau, PhD、Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
- 主任研究者:Francis J Doyle III, PhD、Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
- 主任研究者:Alejandro Leguna, PhD、Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
出版物と役立つリンク
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。