- ICH GCP
- 미국 임상 시험 레지스트리
- 임상시험 NCT03092310
Feasibility Evaluation of an Artificial Pancreas With Glucose Prediction Trust Index
Feasibility Evaluation of an Artificial Pancreas With an Enhanced Model Predictive Control (eMPC) Controller and a Health Monitoring System (HMS) With a Glucose Prediction Trust Index
연구 개요
상세 설명
This is a feasibility study of an artificial pancreas (AP) system with our previously validated target eMPC and HMS algorithms, with the addition of a trust index of the predicted glucose value integrated into the portable Artificial Pancreas System (pAPS). The system will be evaluated on up to 20 subjects, with the goal of 10 subjects completing a 48-hour closed-loop session at one clinical site (William Sansum Diabetes Center). During the session subjects will bolus for all meals and snacks and perform a 45-minute walking session.
The purpose of this pilot study is to establish that closed-loop insulin delivery with a target enchanted model predictive control (eMPC)/Health Monitoring System (HMS) algorithms with a trust index of the predicted glucose value is safe and effective, to analyze and learn to improve upon the accuracy of the predicted glucose values, and to collect efficacy data to inform a future larger study.
연구 유형
등록 (실제)
단계
- 해당 없음
연락처 및 위치
연구 장소
-
-
California
-
Santa Barbara, California, 미국, 93105
- William Sansum Diabetes Center
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
설명
Inclusion Criteria:
- Age ≥ 18 and ≤ 75 years at the time of screening.
- Clinical diagnosis of type 1 diabetes for at least one year.
- Has been using an insulin pump for at least 6 months at the time of screening.
- HbA1c < 10%, as performed by point of care or central laboratory testing. A1c will be assessed at the screening visit, or if already completed within 2 months of the screening visit, the prior lab value may be used in lieu of repeating this assessment.
- Serum creatinine of < 2.0 mg/dL. Serum creatinine will be assessed at the screening visit, or if already completed within 2 months of the screening visit, the prior lab value may be used in lieu of repeating this assessment.
- Bolus for all meals and snacks that contain ≥ 5 grams of carbohydrate.
- Willing to perform at least 7 fingerstick blood glucose tests a day.
- Willing to refrain from taking acetaminophen products for the duration of the clinical trial. If acetaminophen is taken, subject is to avoid making any insulin dosing decisions based on CGM for at least 12 hours.
- Willing to abide by the study protocol and use study-provided devices, including the Omnipod, Dexcom CGM, glucometer, ketone meter and pAPS tablet.
Exclusion Criteria:
- Pregnancy
- One or more episodes of hypoglycemia requiring an emergency room visit or hospitalization in the past 6 months.
- One or more episodes of hyperglycemia requiring an emergency room visit or hospitalization in the past 6 months.
- Known unstable cardiac disease or untreated cardiac disease, as revealed by history or physical examination.
- Dermatological conditions that would preclude wearing a CGM sensor or Pod.
- One or more seizures in the past year.
- Screening A1c ≥ 10% or serum creatinine ≥ 2.0 mg/dL.
- Any condition that could interfere with participating in the trial, based on investigator judgment.
- Concurrent use of any non-insulin glucose-lowering agent (including GLP-1 agonists, Symlin, DPP-4 inhibitors, SGLT-2 inhibitors, biguanides, sulfonylureas and naturaceuticals), defined as using any of these medications within 30 days of the screening visit or during the study.
- Participation in another pharmaceutical or device trial at the time of enrollment or during the study.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 치료
- 할당: 해당 없음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: Artificial Pancreas
The artificial pancreas device will employ its enhanced Model Predictive Control (MPC) algorithm with a target glucose level of 110 mg/dL with a trust index for MPC-predicted glucose values, weighing future glucose predictions and only acting on predictions with higher weight in the trust index.
The Health Monitoring System algorithm uses the same CGM data as the MPC control algorithm but utilizes a separate algorithm for trending and predictions of future glucose values.
Using a redundant and independent algorithm is an important safety feature of the overall AP device.
|
Artificial Pancreas Device
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Percent time within the target glucose range of 70-180 mg/dl overall as assessed by continuous glucose monitoring (CGM).
기간: 48 hours
|
The primary endpoint for this pilot study will be time within the target glucose range of 70-180 mg/dl overall as assessed by continuous glucose monitoring (CGM), during closed-loop use, determining if the combination of eMPC and the assessment of trust index can provide safe and effective glucose control.
|
48 hours
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Assessment of the trust index predictions vs. standard eMPC glucose predictions.
기간: 48 hours
|
During closed-loop use, assessment of the trust index predictions vs. standard eMPC glucose predictions (frequency of moving away from nominal weights of predicted glucose).
|
48 hours
|
|
Frequency of hypoglycemia during closed-loop use.
기간: 48 hours
|
Frequency of hypoglycemia during closed-loop use, defined as percent time glucose is below predefined thresholds of 70 mg/dl, 60 mg/dl and 50 mg/dL.
|
48 hours
|
|
Frequency of hyperglycemia during closed-loop use.
기간: 48 hours
|
Frequency of hyperglycemia during closed-loop use, defined as percent time glucose is above predefined thresholds of 180 mg/dl, 250 mg/dl, and 300 mg/dL.
|
48 hours
|
|
Percent time within the target glucose range of 70-140 mg/dl during closed-loop use.
기간: 48 hours
|
Percent time within the target glucose range of 70-140 mg/dl overall as assessed by continuous glucose monitoring (CGM), during closed-loop use.
|
48 hours
|
|
Percent time glucose is within range 70-150 mg/dl postprandial within 5 hours following meals, during closed-loop use.
기간: 48 hours
|
Percent time glucose is within range 70-150 mg/dl postprandial within 5 hours following meals, during closed-loop use.
|
48 hours
|
|
Standard Deviation and Coefficient of Variation of glucose values
기간: 48 hours
|
To measure glycemic variability, the standard deviation and coefficient of variation of glucose values (as measured by continuous glucose monitoring) will be determined during the 48 hours of closed-loop use.
|
48 hours
|
|
Markers of hypo- and hyperglycemia
기간: 48 hours
|
Markers of hypo- and hyperglycemia, to include LBGI (Low blood glucose index) and HBGI (High blood glucose index), during closed-loop use.
|
48 hours
|
|
Insulin Doses Given
기간: 48 hours
|
Change in total daily doses and post prandial insulin doses given during the 48 hours of closed-loop use.
|
48 hours
|
|
Number of treatments given for hypoglycemia (defined as carbohydrates given for treatment of a documented or suspected low blood sugar), during the 48 hours of closed-loop use.
기간: 48 hours
|
Treatments for hypoglycemia given during the 48 hours of closed-loop use.
Hypoglycemia treatments may be given in response to alerts from the HMS, if blood glucose < 70 mg/dL, or at the request of the subject or investigator.
|
48 hours
|
|
Number of alerts given by the HMS to prevent hypoglycemia
기간: 48 hours
|
Number of alerts given by the HMS to prevent hypoglycemia during closed-loop use.
|
48 hours
|
|
Failure analysis of the devices/connectivity issues that may occur (# disconnects and device restarts).
기간: 48 hours
|
Failure analysis of the devices/connectivity issues that may occur during closed-loop use.
This includes number of CGM communication losses with 3 or more missed points, and number of times the entire system required a restart.
|
48 hours
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Eyal Dassau, PhD, Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
- 수석 연구원: Francis J Doyle III, PhD, Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
- 수석 연구원: Alejandro Leguna, PhD, Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University
간행물 및 유용한 링크
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .