Real Time ST-segment Deviation Detection in High-risk Patients Detected by Wireless Single-lead ECG
調査の概要
詳細な説明
Background:
Assesment of wireless ECG-monitoring devices is essential for patient monitoring as it enables activity and out-of-hospital monitoring. Former studies have found a random variation for ST-deviations between wireless single-lead ECG and 12-lead ECG, which suggests that clinical monitoring of ST-deviations by single-lead ECG must be corrected for this to minimize false positive cases. Thus, we aimed to assess how a corrected limit of agreement for ST-deviations affects the ability to detect true myocardial ischemia by comparing thresholds to troponin elevations.
Methods:
This study sought to analyze data from two WARD cohort studies (NCT03491137) and (NCT03660501). Patients were included at Rigshospitalet and Bispebjerg Hospital from February 2018 to September 2020. The Surgery and COPD studies consisted of 505 and 202 patients, respectively, who were hospitalized with AECOPD or scheduled for elective major abdominal surgery lasting > 2 hours. All patients were monitored with wireless equipment, which continuously measured vital signs. The continuous wireless monitoring used in this study consisted of a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s. Another monitoring device was the plasma concentration of high-sensitive cardiac troponin T (hsTnT, Cobas 8000, e801 module, Diagnostics Roche). In this study, patients were divided into a myocardial injury group with TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
Analysis:
Baseline characteristics of patients undergoing major abdominal surgery and patients admitted with AECOPD with and without myocardial injury will be summarized in frequency tables including Using descriptive statistics, we will describe the occurrence of ST-segment deviations in the single-lead (Lifetouch) ECG in patients with and without myocardial injury as well as the association between ST-deviation and subsequent myocardial injury, which will be adjusted for known risk factors; age, gender, BMI, daily smoker (never-, previous-, current smoker), alcohol consumption (none, below- or above recommendations), ASA classification (ASA 1,2,3,4), history of myocardial infarction, history of diabetes mellitus, history of chronic heart failure. Furthermore, the sensitivity and specificity of continuous wireless single-lead ECG with the applied threshold of <-0.255 mV and > 0.245 mV will be determined. Receiver operating characteristics (ROC) analyses will be made to determine which type of ST-deviation and which specific threshold value was most sensitive for TnT elevations. Through ROC analyses it will also be examined how different variations of durations affect the thresholds' sensitivity and specificity. Statistical analysis will be performed using statistical software SPSS.
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
-
Copenhagen、デンマーク、2100
- Rigshospitalet, Department of Anesthesiology, Centre for Cancer and Organ Diseases, University of Copenhagen, Blegdamsvej 9, 2200 Copenhagen, Denmark
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
WARD-COPD:
- ≥ 18 years
- Patients admitted to emergency departments or pulmonary wards with AECOPD as admission diagnosis with suspected acute exacerbation of COPD
- The diagnosis had to be sustained in the patient record at the time of inclusion
- Expected admittance longer than 24 hours
- Possibility of an investigator to include the patient within 24 hours from admission
WARD-Surgery:
- ≥ 60 years
- Elective major abdominal cancer surgery
- Estimated surgical duration ≥ 2 hours.
Exclusion Criteria:
WARD-COPD:
- Patients who were not expected to cooperate to wear the monitoring equipment
- Unable to give informed consent
- Patients who were withheld active treatment
- Implanted cardioverter defibrillator or pacemaker
- Severe allergy for plaster/silicone
WARD-Surgery:
- Implanted cardioverter defibrillator or pacemaker
- Allergy to study devices
- Severe cognitive impairment assessed by MMSE
- Inability to cooperate wearing the wireless monitoring equipment
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
no myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
|
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax.
The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
|
|
Myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
|
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax.
The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV on single-lead ECG
時間枠:30 minutes
|
The primary outcome is the frequency of patients with ST-deviation <-0.255 and/or >0.245 mV (0.1 mV + LoA) on single-lead ECG monitoring lasting ≥ 30 minutes.
|
30 minutes
|
二次結果の測定
結果測定 |
時間枠 |
|---|---|
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 1 minute.
時間枠:1 minute
|
1 minute
|
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 10 minutes.
時間枠:10 minutes
|
10 minutes
|
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 20 minutes.
時間枠:20 minutes
|
20 minutes
|
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 30 minutes.
時間枠:30 minutes
|
30 minutes
|
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 1 minute.
時間枠:1 minute
|
1 minute
|
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 10 minutes.
時間枠:10 minutes
|
10 minutes
|
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 20 minutes.
時間枠:20 minutes
|
20 minutes
|
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 30 minutes and myocardial injury within three days of the first continuous monitoring day.
時間枠:3 days
|
3 days
|
協力者と研究者
スポンサー
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- ST-Deviation WARD
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。