Real Time ST-segment Deviation Detection in High-risk Patients Detected by Wireless Single-lead ECG

2022年1月3日 更新者:Jonathan Attilla Koefoed Starling、Rigshospitalet, Denmark
This study primarily aims to describe the frequency of significant ST-deviations, defined as ECG-ST-deviations <-0.255 or >0.245 mV for a minimum duration of 30 minutes as measured by a single-lead ECG in patients admitted with AECOPD or following major abdominal surgery. Secondarily we will describe the frequency of ST-deviations <-0.255 or >0.245 mV for a minimum duration of 1, 10 and 20 minutes, respectively, as well as for patients with ST-deviations <-0.1 or >0.1 mV for a minimum duration of 1, 10, 20 and 30 minutes, respectively. Lastly, we will investigate the association between ST-deviations and subsequent myocardial injury while adjusting for known risk factors.

調査の概要

詳細な説明

Background:

Assesment of wireless ECG-monitoring devices is essential for patient monitoring as it enables activity and out-of-hospital monitoring. Former studies have found a random variation for ST-deviations between wireless single-lead ECG and 12-lead ECG, which suggests that clinical monitoring of ST-deviations by single-lead ECG must be corrected for this to minimize false positive cases. Thus, we aimed to assess how a corrected limit of agreement for ST-deviations affects the ability to detect true myocardial ischemia by comparing thresholds to troponin elevations.

Methods:

This study sought to analyze data from two WARD cohort studies (NCT03491137) and (NCT03660501). Patients were included at Rigshospitalet and Bispebjerg Hospital from February 2018 to September 2020. The Surgery and COPD studies consisted of 505 and 202 patients, respectively, who were hospitalized with AECOPD or scheduled for elective major abdominal surgery lasting > 2 hours. All patients were monitored with wireless equipment, which continuously measured vital signs. The continuous wireless monitoring used in this study consisted of a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s. Another monitoring device was the plasma concentration of high-sensitive cardiac troponin T (hsTnT, Cobas 8000, e801 module, Diagnostics Roche). In this study, patients were divided into a myocardial injury group with TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.

Analysis:

Baseline characteristics of patients undergoing major abdominal surgery and patients admitted with AECOPD with and without myocardial injury will be summarized in frequency tables including Using descriptive statistics, we will describe the occurrence of ST-segment deviations in the single-lead (Lifetouch) ECG in patients with and without myocardial injury as well as the association between ST-deviation and subsequent myocardial injury, which will be adjusted for known risk factors; age, gender, BMI, daily smoker (never-, previous-, current smoker), alcohol consumption (none, below- or above recommendations), ASA classification (ASA 1,2,3,4), history of myocardial infarction, history of diabetes mellitus, history of chronic heart failure. Furthermore, the sensitivity and specificity of continuous wireless single-lead ECG with the applied threshold of <-0.255 mV and > 0.245 mV will be determined. Receiver operating characteristics (ROC) analyses will be made to determine which type of ST-deviation and which specific threshold value was most sensitive for TnT elevations. Through ROC analyses it will also be examined how different variations of durations affect the thresholds' sensitivity and specificity. Statistical analysis will be performed using statistical software SPSS.

研究の種類

観察的

入学 (実際)

698

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

      • Copenhagen、デンマーク、2100
        • Rigshospitalet, Department of Anesthesiology, Centre for Cancer and Organ Diseases, University of Copenhagen, Blegdamsvej 9, 2200 Copenhagen, Denmark

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

サンプリング方法

非確率サンプル

調査対象母集団

Patients from the two WARD studies: WARD COPD and WARD Surgery. Patients were included at Rigshospitalet (department of abdominal surgery) and Bispebjerg Hospital (departments of emergency medicine, pulmonary medicine, and abdominal surgery) in Denmark from February 2018 to September 2020 after obtained written informed consent. Eligibility was determined by the investigator. All patients were monitored with wireless equipment, which continuously measures vital signs.

説明

Inclusion Criteria:

WARD-COPD:

  • ≥ 18 years
  • Patients admitted to emergency departments or pulmonary wards with AECOPD as admission diagnosis with suspected acute exacerbation of COPD
  • The diagnosis had to be sustained in the patient record at the time of inclusion
  • Expected admittance longer than 24 hours
  • Possibility of an investigator to include the patient within 24 hours from admission

WARD-Surgery:

  • ≥ 60 years
  • Elective major abdominal cancer surgery
  • Estimated surgical duration ≥ 2 hours.

Exclusion Criteria:

WARD-COPD:

  • Patients who were not expected to cooperate to wear the monitoring equipment
  • Unable to give informed consent
  • Patients who were withheld active treatment
  • Implanted cardioverter defibrillator or pacemaker
  • Severe allergy for plaster/silicone

WARD-Surgery:

  • Implanted cardioverter defibrillator or pacemaker
  • Allergy to study devices
  • Severe cognitive impairment assessed by MMSE
  • Inability to cooperate wearing the wireless monitoring equipment

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
no myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
Myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV on single-lead ECG
時間枠:30 minutes
The primary outcome is the frequency of patients with ST-deviation <-0.255 and/or >0.245 mV (0.1 mV + LoA) on single-lead ECG monitoring lasting ≥ 30 minutes.
30 minutes

二次結果の測定

結果測定
時間枠
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 1 minute.
時間枠:1 minute
1 minute
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 10 minutes.
時間枠:10 minutes
10 minutes
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 20 minutes.
時間枠:20 minutes
20 minutes
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 30 minutes.
時間枠:30 minutes
30 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 1 minute.
時間枠:1 minute
1 minute
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 10 minutes.
時間枠:10 minutes
10 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 20 minutes.
時間枠:20 minutes
20 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 30 minutes and myocardial injury within three days of the first continuous monitoring day.
時間枠:3 days
3 days

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2018年2月20日

一次修了 (実際)

2020年6月9日

研究の完了 (実際)

2020年6月9日

試験登録日

最初に提出

2021年12月10日

QC基準を満たした最初の提出物

2021年12月10日

最初の投稿 (実際)

2021年12月29日

学習記録の更新

投稿された最後の更新 (実際)

2022年1月12日

QC基準を満たした最後の更新が送信されました

2022年1月3日

最終確認日

2022年1月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • ST-Deviation WARD

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する