Real Time ST-segment Deviation Detection in High-risk Patients Detected by Wireless Single-lead ECG
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
Background:
Assesment of wireless ECG-monitoring devices is essential for patient monitoring as it enables activity and out-of-hospital monitoring. Former studies have found a random variation for ST-deviations between wireless single-lead ECG and 12-lead ECG, which suggests that clinical monitoring of ST-deviations by single-lead ECG must be corrected for this to minimize false positive cases. Thus, we aimed to assess how a corrected limit of agreement for ST-deviations affects the ability to detect true myocardial ischemia by comparing thresholds to troponin elevations.
Methods:
This study sought to analyze data from two WARD cohort studies (NCT03491137) and (NCT03660501). Patients were included at Rigshospitalet and Bispebjerg Hospital from February 2018 to September 2020. The Surgery and COPD studies consisted of 505 and 202 patients, respectively, who were hospitalized with AECOPD or scheduled for elective major abdominal surgery lasting > 2 hours. All patients were monitored with wireless equipment, which continuously measured vital signs. The continuous wireless monitoring used in this study consisted of a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s. Another monitoring device was the plasma concentration of high-sensitive cardiac troponin T (hsTnT, Cobas 8000, e801 module, Diagnostics Roche). In this study, patients were divided into a myocardial injury group with TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
Analysis:
Baseline characteristics of patients undergoing major abdominal surgery and patients admitted with AECOPD with and without myocardial injury will be summarized in frequency tables including Using descriptive statistics, we will describe the occurrence of ST-segment deviations in the single-lead (Lifetouch) ECG in patients with and without myocardial injury as well as the association between ST-deviation and subsequent myocardial injury, which will be adjusted for known risk factors; age, gender, BMI, daily smoker (never-, previous-, current smoker), alcohol consumption (none, below- or above recommendations), ASA classification (ASA 1,2,3,4), history of myocardial infarction, history of diabetes mellitus, history of chronic heart failure. Furthermore, the sensitivity and specificity of continuous wireless single-lead ECG with the applied threshold of <-0.255 mV and > 0.245 mV will be determined. Receiver operating characteristics (ROC) analyses will be made to determine which type of ST-deviation and which specific threshold value was most sensitive for TnT elevations. Through ROC analyses it will also be examined how different variations of durations affect the thresholds' sensitivity and specificity. Statistical analysis will be performed using statistical software SPSS.
Tipo di studio
Iscrizione (Effettivo)
Contatti e Sedi
Luoghi di studio
-
-
-
Copenhagen, Danimarca, 2100
- Rigshospitalet, Department of Anesthesiology, Centre for Cancer and Organ Diseases, University of Copenhagen, Blegdamsvej 9, 2200 Copenhagen, Denmark
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
WARD-COPD:
- ≥ 18 years
- Patients admitted to emergency departments or pulmonary wards with AECOPD as admission diagnosis with suspected acute exacerbation of COPD
- The diagnosis had to be sustained in the patient record at the time of inclusion
- Expected admittance longer than 24 hours
- Possibility of an investigator to include the patient within 24 hours from admission
WARD-Surgery:
- ≥ 60 years
- Elective major abdominal cancer surgery
- Estimated surgical duration ≥ 2 hours.
Exclusion Criteria:
WARD-COPD:
- Patients who were not expected to cooperate to wear the monitoring equipment
- Unable to give informed consent
- Patients who were withheld active treatment
- Implanted cardioverter defibrillator or pacemaker
- Severe allergy for plaster/silicone
WARD-Surgery:
- Implanted cardioverter defibrillator or pacemaker
- Allergy to study devices
- Severe cognitive impairment assessed by MMSE
- Inability to cooperate wearing the wireless monitoring equipment
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Coorti e interventi
Gruppo / Coorte |
Intervento / Trattamento |
---|---|
no myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
|
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax.
The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
|
Myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
|
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax.
The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
---|---|---|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV on single-lead ECG
Lasso di tempo: 30 minutes
|
The primary outcome is the frequency of patients with ST-deviation <-0.255 and/or >0.245 mV (0.1 mV + LoA) on single-lead ECG monitoring lasting ≥ 30 minutes.
|
30 minutes
|
Misure di risultato secondarie
Misura del risultato |
Lasso di tempo |
---|---|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 1 minute.
Lasso di tempo: 1 minute
|
1 minute
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 10 minutes.
Lasso di tempo: 10 minutes
|
10 minutes
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 20 minutes.
Lasso di tempo: 20 minutes
|
20 minutes
|
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 30 minutes.
Lasso di tempo: 30 minutes
|
30 minutes
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 1 minute.
Lasso di tempo: 1 minute
|
1 minute
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 10 minutes.
Lasso di tempo: 10 minutes
|
10 minutes
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 20 minutes.
Lasso di tempo: 20 minutes
|
20 minutes
|
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 30 minutes and myocardial injury within three days of the first continuous monitoring day.
Lasso di tempo: 3 days
|
3 days
|
Collaboratori e investigatori
Sponsor
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Altri numeri di identificazione dello studio
- ST-Deviation WARD
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Lesioni al miocardio
-
Hospital Universitari Son DuretaEspen; This research prize was funded by Nestle Nutrition Institute and by Fresenius...CompletatoTrauma da moderato a grave, come definito da an | Injury Severity Score (ISS) > 12 punti sono stati inclusi nello studio.Spagna
Prove cliniche su Lifetouch single-lead ECG monitoring
-
Memorial Sloan Kettering Cancer CenterTerminatoFibrillazione | Pazienti oncologici postoperatori con esperienza atrialeStati Uniti