Real Time ST-segment Deviation Detection in High-risk Patients Detected by Wireless Single-lead ECG

January 3, 2022 updated by: Jonathan Attilla Koefoed Starling, Rigshospitalet, Denmark
This study primarily aims to describe the frequency of significant ST-deviations, defined as ECG-ST-deviations <-0.255 or >0.245 mV for a minimum duration of 30 minutes as measured by a single-lead ECG in patients admitted with AECOPD or following major abdominal surgery. Secondarily we will describe the frequency of ST-deviations <-0.255 or >0.245 mV for a minimum duration of 1, 10 and 20 minutes, respectively, as well as for patients with ST-deviations <-0.1 or >0.1 mV for a minimum duration of 1, 10, 20 and 30 minutes, respectively. Lastly, we will investigate the association between ST-deviations and subsequent myocardial injury while adjusting for known risk factors.

Study Overview

Detailed Description

Background:

Assesment of wireless ECG-monitoring devices is essential for patient monitoring as it enables activity and out-of-hospital monitoring. Former studies have found a random variation for ST-deviations between wireless single-lead ECG and 12-lead ECG, which suggests that clinical monitoring of ST-deviations by single-lead ECG must be corrected for this to minimize false positive cases. Thus, we aimed to assess how a corrected limit of agreement for ST-deviations affects the ability to detect true myocardial ischemia by comparing thresholds to troponin elevations.

Methods:

This study sought to analyze data from two WARD cohort studies (NCT03491137) and (NCT03660501). Patients were included at Rigshospitalet and Bispebjerg Hospital from February 2018 to September 2020. The Surgery and COPD studies consisted of 505 and 202 patients, respectively, who were hospitalized with AECOPD or scheduled for elective major abdominal surgery lasting > 2 hours. All patients were monitored with wireless equipment, which continuously measured vital signs. The continuous wireless monitoring used in this study consisted of a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s. Another monitoring device was the plasma concentration of high-sensitive cardiac troponin T (hsTnT, Cobas 8000, e801 module, Diagnostics Roche). In this study, patients were divided into a myocardial injury group with TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.

Analysis:

Baseline characteristics of patients undergoing major abdominal surgery and patients admitted with AECOPD with and without myocardial injury will be summarized in frequency tables including Using descriptive statistics, we will describe the occurrence of ST-segment deviations in the single-lead (Lifetouch) ECG in patients with and without myocardial injury as well as the association between ST-deviation and subsequent myocardial injury, which will be adjusted for known risk factors; age, gender, BMI, daily smoker (never-, previous-, current smoker), alcohol consumption (none, below- or above recommendations), ASA classification (ASA 1,2,3,4), history of myocardial infarction, history of diabetes mellitus, history of chronic heart failure. Furthermore, the sensitivity and specificity of continuous wireless single-lead ECG with the applied threshold of <-0.255 mV and > 0.245 mV will be determined. Receiver operating characteristics (ROC) analyses will be made to determine which type of ST-deviation and which specific threshold value was most sensitive for TnT elevations. Through ROC analyses it will also be examined how different variations of durations affect the thresholds' sensitivity and specificity. Statistical analysis will be performed using statistical software SPSS.

Study Type

Observational

Enrollment (Actual)

698

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Copenhagen, Denmark, 2100
        • Rigshospitalet, Department of Anesthesiology, Centre for Cancer and Organ Diseases, University of Copenhagen, Blegdamsvej 9, 2200 Copenhagen, Denmark

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Patients from the two WARD studies: WARD COPD and WARD Surgery. Patients were included at Rigshospitalet (department of abdominal surgery) and Bispebjerg Hospital (departments of emergency medicine, pulmonary medicine, and abdominal surgery) in Denmark from February 2018 to September 2020 after obtained written informed consent. Eligibility was determined by the investigator. All patients were monitored with wireless equipment, which continuously measures vital signs.

Description

Inclusion Criteria:

WARD-COPD:

  • ≥ 18 years
  • Patients admitted to emergency departments or pulmonary wards with AECOPD as admission diagnosis with suspected acute exacerbation of COPD
  • The diagnosis had to be sustained in the patient record at the time of inclusion
  • Expected admittance longer than 24 hours
  • Possibility of an investigator to include the patient within 24 hours from admission

WARD-Surgery:

  • ≥ 60 years
  • Elective major abdominal cancer surgery
  • Estimated surgical duration ≥ 2 hours.

Exclusion Criteria:

WARD-COPD:

  • Patients who were not expected to cooperate to wear the monitoring equipment
  • Unable to give informed consent
  • Patients who were withheld active treatment
  • Implanted cardioverter defibrillator or pacemaker
  • Severe allergy for plaster/silicone

WARD-Surgery:

  • Implanted cardioverter defibrillator or pacemaker
  • Allergy to study devices
  • Severe cognitive impairment assessed by MMSE
  • Inability to cooperate wearing the wireless monitoring equipment

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
no myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).
Myocardial injury
In this study, patients were divided into groups of myocardial injury, where patients had TnT elevations (of 20-64 ng/L with an absolute change of ≥ 5 ng/L, or a hsTnT level ≥65 ng/L, both evaluated as due to ischemic etiology) and a group with no myocardial injury.
The continuous wireless monitoring is a single-lead ECG patch (Isansys Life Touch patch) which consists of two ECG electrodes placed on the left side of the patients' thorax. The patch measures respiratory rate (RR), heart rate (HR), heart rate variability (HRV) and electrocardiogram (ECG)s.
Another monitoring device in the study was the plasma concentration of high-sensitive cardiac troponin T (hsTnT).

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV on single-lead ECG
Time Frame: 30 minutes
The primary outcome is the frequency of patients with ST-deviation <-0.255 and/or >0.245 mV (0.1 mV + LoA) on single-lead ECG monitoring lasting ≥ 30 minutes.
30 minutes

Secondary Outcome Measures

Outcome Measure
Time Frame
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 1 minute.
Time Frame: 1 minute
1 minute
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 10 minutes.
Time Frame: 10 minutes
10 minutes
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 20 minutes.
Time Frame: 20 minutes
20 minutes
Frequency of patients with significant ST-deviations <-0.1 and/or >0.1 mV lasting ≥ 30 minutes.
Time Frame: 30 minutes
30 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 1 minute.
Time Frame: 1 minute
1 minute
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 10 minutes.
Time Frame: 10 minutes
10 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 20 minutes.
Time Frame: 20 minutes
20 minutes
Frequency of patients with ST-deviation <-0.255 and/or >0.245 mV lasting ≥ 30 minutes and myocardial injury within three days of the first continuous monitoring day.
Time Frame: 3 days
3 days

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

February 20, 2018

Primary Completion (Actual)

June 9, 2020

Study Completion (Actual)

June 9, 2020

Study Registration Dates

First Submitted

December 10, 2021

First Submitted That Met QC Criteria

December 10, 2021

First Posted (Actual)

December 29, 2021

Study Record Updates

Last Update Posted (Actual)

January 12, 2022

Last Update Submitted That Met QC Criteria

January 3, 2022

Last Verified

January 1, 2022

More Information

Terms related to this study

Other Study ID Numbers

  • ST-Deviation WARD

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Myocardial Injury

Clinical Trials on Lifetouch single-lead ECG monitoring

3
Subscribe