Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System

10 september 2015 bijgewerkt door: DexCom, Inc.

PTL-900590 Effectiveness and Safety Study of the Dexcom G4™ Continuous Glucose Monitoring System

The purpose of this study is to evaluate the effectiveness and safety of the Dexcom G4 System when used as an adjuvant to blood glucose testing over a 7-day period in people >18 years-old with diabetes mellitus.

The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI.

Safety of the G4 System will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants.

Studie Overzicht

Toestand

Voltooid

Conditie

Gedetailleerde beschrijving

The primary objective is to characterize the System performance with respect to laboratory reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the G4 System in reference to YSI. The trend accuracy of the device performance, i.e., continuous glucose error grid analysis, will be evaluated as well as the temporal G4 System accuracy at different glucose rates of changes and different glucose ranges (hypoglycemic, euglycemic, and hyperglycemic ranges). The primary matched paired (Sensor-YSI) measurements will be collected during multiple in-clinic days.

Studietype

Ingrijpend

Inschrijving (Werkelijk)

72

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

    • California
      • Chula Vista, California, Verenigde Staten, 91911
        • Profil Institute for Clinical Research
      • Escondido, California, Verenigde Staten, 92026
        • AMCR Institute, Inc.
      • Walnut Creek, California, Verenigde Staten, 94598
        • Diablo Clinical Research
    • Idaho
      • Idaho Falls, Idaho, Verenigde Staten, 83404
        • Rocky Mountain Diabetes and Osteoporosis Center

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  • Age 18 years or older;
  • Diagnosed with diabetes mellitus;
  • Use one of the following for their diabetes management:

    • intensive insulin therapy (IIT) or
    • non-intensive insulin therapy (non-IIT)
  • Intensive-insulin using (IIT) subjects only - During each in-clinic session, willing to have their blood glucose levels manipulated into high and low glucose levels. Subjects not on IIT will only be observed during each in-clinic session.
  • For subjects that exercise routinely (at least 3 times per week), willing to exercise on each in-clinic session, if asked;
  • Willing to take a minimum of 7 fingersticks per day during home use days;
  • Willing to refrain from the use of acetaminophen during sensor wear period and the day prior to sensor insertion;

Exclusion Criteria:

  • Have extensive skin changes/diseases that preclude wearing the required number of devices on normal skin (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites;
  • Have a known allergy to medical-grade adhesives;
  • Are pregnant as demonstrated by a positive pregnancy test within 72 hours of sensor insertion,
  • Are on dialysis;
  • Have a hematocrit that is less than 36% or greater than 55% at screening visit;
  • Currently participating in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study);
  • Have a history of cardiovascular disease (including, but not limited to ischemic heart disease, peripheral vascular disease, cardiomyopathy, cerebrovascular disease, congenital heart disease, or significant arrhythmias), epilepsy, severe migraines in the past 6 months, adrenal disease, syncope, significant hypoglycemia unawareness, or a history of severe hypoglycemia (requiring emergency medical intervention) within the last 6 months;
  • Have any chronic infectious disease or intercurrent illness that would interfere with their participation in the study or pose an excessive risk to study staff handling venous or capillary blood samples (e.g. HIV/AIDS, Hepatitis B or C);
  • Have a MRI scan, CT scan, or diathermy scheduled during the week of the study. If any of these procedures are required urgently during the study, subjects will notify the study staff, end their CGM session, and remove their sensor.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: NVT
  • Interventioneel model: Opdracht voor een enkele groep
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Real Time Continuous Glucose Monitoring System
7 day use of real time continuous glucose monitoring system
Manipulation of glucose levels during multiple clinic days. Achieved through diet, exercise and insulin regimen defined in protocol.Evaluation of CGM performance characteristics in reference to a laboratory standard.
Andere namen:
  • CGM-systeem
  • G4 CGM System
  • Dexcom G4 CGM-systeem

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
CGM Relative Differences to Laboratory Reference
Tijdsspanne: 7 days
The outcome measure is measured as the relative differences (%) of the CGM glucose value in reference to a laboratory reference, yellow spring instrument( YSI) glucose measurements.
7 days

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Sponsor

Onderzoekers

  • Studie directeur: David Price, MD, DexCom, Inc.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start

1 december 2011

Primaire voltooiing (Werkelijk)

1 februari 2012

Studie voltooiing (Werkelijk)

1 februari 2012

Studieregistratiedata

Eerst ingediend

12 januari 2012

Eerst ingediend dat voldeed aan de QC-criteria

17 januari 2012

Eerst geplaatst (Schatting)

23 januari 2012

Updates van studierecords

Laatste update geplaatst (Schatting)

5 oktober 2015

Laatste update ingediend die voldeed aan QC-criteria

10 september 2015

Laatst geverifieerd

1 augustus 2015

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • PTL-900590

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren