Closed-Loop Stimulation in Left Bundle Branch Area Pacing

4 września 2026 zaktualizowane przez: Bartlomiej Wrobel

Closed-Loop Stimulation Combined With Left Bundle Branch Area Pacing in Patients With Atrioventricular Conduction Disorders

This prospective, single-center, crossover study evaluated the effects of closed-loop stimulation (CLS) compared with conventional accelerometer-based rate-adaptive pacing (VVIR) in patients with permanent atrial fibrillation undergoing left bundle branch area pacing (LBBAP). Chronotropic responses were evaluated during standardized mental and physical stress testing. Heart rate distribution during daily activity, functional capacity, quality of life, and device-related parameters were also assessed.

Przegląd badań

Typ studiów

Interwencyjne

Zapisy (Rzeczywisty)

33

Faza

  • Nie dotyczy

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Lokalizacje studiów

    • Podkarpackie Voivodeship
      • Rzeszów, Podkarpackie Voivodeship, Polska, 35-302
        • Clinical Provincial Hospital No. 2 in Rzeszow

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

  • Dorosły
  • Starszy dorosły

Akceptuje zdrowych ochotników

Nie

Opis

Inclusion Criteria:

  • Age ≥18 years
  • Permanent atrial fibrillation
  • Chronotropic incompetence
  • Permanent pacemaker with left bundle branch area pacing (LBBAP)
  • Ventricular pacing burden >50%
  • Ability to provide written informed consent

Exclusion Criteria:

  • Age <18 years
  • Inability or unwillingness to provide informed consent

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

  • Główny cel: Leczenie
  • Przydział: Nielosowe
  • Model interwencyjny: Zadanie krzyżowe
  • Maskowanie: Pojedynczy

Broń i interwencje

Grupa uczestników / Arm
Interwencja / Leczenie
Eksperymentalny: CLS followed by VVIR
Participants received closed-loop stimulation (CLS) during the first study period, followed by accelerometer-based VVIR pacing during the second study period.
The pacemaker was programmed to closed-loop stimulation (CLS), an impedance-based rate-adaptive pacing algorithm that adjusts the pacing rate in response to changes in intracardiac impedance associated with cardiac contractile dynamics.
The pacemaker was programmed to VVIR mode with accelerometer-based rate-adaptive pacing. The pacing rate was adjusted according to physical activity detected by the device's accelerometer.
Eksperymentalny: VVIR followed by CLS
Participants received accelerometer-based VVIR pacing during the first study period, followed by closed-loop stimulation (CLS) during the second study period.
The pacemaker was programmed to closed-loop stimulation (CLS), an impedance-based rate-adaptive pacing algorithm that adjusts the pacing rate in response to changes in intracardiac impedance associated with cardiac contractile dynamics.
The pacemaker was programmed to VVIR mode with accelerometer-based rate-adaptive pacing. The pacing rate was adjusted according to physical activity detected by the device's accelerometer.

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Change in Heart Rate During Mental Stress
Ramy czasowe: At the end of each 1-month pacing period
Change in heart rate (ΔHR) during the Paced Auditory Serial Addition Test (PASAT), calculated as the peak heart rate recorded during the test minus the baseline heart rate. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing and expressed in beats per minute (bpm).
At the end of each 1-month pacing period
Change in Heart Rate During Isometric Handgrip
Ramy czasowe: At the end of each 1-month pacing period
Change in heart rate (ΔHR) during the 5-minute isometric handgrip test performed at 30% of maximal voluntary contraction, calculated as the peak heart rate recorded during the test minus the baseline heart rate. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing and expressed in beats per minute (bpm).
At the end of each 1-month pacing period
Change in Heart Rate During Treadmill Exercise
Ramy czasowe: At the end of each 1-month pacing period
Change in heart rate (ΔHR) during the symptom-limited treadmill exercise test using the modified Bruce protocol, calculated as the peak heart rate achieved during exercise minus the baseline heart rate. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing and expressed in beats per minute (bpm).
At the end of each 1-month pacing period

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Exercise Duration During Treadmill Testing
Ramy czasowe: At the end of each 1-month pacing period
Total exercise duration during the symptom-limited treadmill exercise test performed using the modified Bruce protocol. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing and expressed in seconds.
At the end of each 1-month pacing period
Metabolic Equivalents Achieved During Treadmill Testing
Ramy czasowe: At the end of each 1-month pacing period
Exercise capacity during the symptom-limited treadmill exercise test using the modified Bruce protocol, expressed as metabolic equivalents (METs) achieved during the test. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing.
At the end of each 1-month pacing period
Heart Rate Score During Daily Activity
Ramy czasowe: At the end of each 1-month pacing period
Heart Rate Score (HRS), defined as the percentage of all sensed and paced ventricular beats occurring in the most populated 10-beats-per-minute heart rate histogram bin. HRS was assessed from pacemaker-derived heart rate histograms under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing. Higher HRS values indicate a more concentrated heart rate distribution and less heart rate variation across histogram bins. The score ranges from 0% to 100%.
At the end of each 1-month pacing period
Health-Related Quality of Life Assessed With the EQ-5D-5L Index
Ramy czasowe: At the end of each 1-month pacing period
Health-related quality of life was assessed using the EuroQol 5-Dimension 5-Level (EQ-5D-5L) descriptive system. Responses were converted to an EQ-5D-5L index value using the Polish value set. Index values range from -0.590 to 1.000, with higher values indicating better health-related quality of life. A value of 1 represents full health, 0 represents a health state equivalent to death, and negative values represent health states valued as worse than death. The outcome was compared between closed-loop stimulation (CLS) and accelerometer-based VVIR pacing.
At the end of each 1-month pacing period
Self-Rated Health Assessed With the EQ VAS
Ramy czasowe: At the end of each 1-month pacing period
Self-rated health was assessed using the EuroQol Visual Analogue Scale (EQ VAS), ranging from 0 to 100, where 0 represents the worst health the participant can imagine and 100 represents the best health the participant can imagine. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing.
At the end of each 1-month pacing period
Patient-Reported Pacing-Mode Preference
Ramy czasowe: At the end of the study, after completion of both 1-month pacing periods
At the end of the study, participants were asked during which of the two pacing periods they experienced better overall subjective well-being, or whether they had no preference. Responses were subsequently categorized according to the pacing mode programmed during the preferred period as preference for closed-loop stimulation (CLS), preference for accelerometer-based VVIR pacing, or no preference.
At the end of the study, after completion of both 1-month pacing periods
Ventricular Pacing Percentage
Ramy czasowe: At the end of each 1-month pacing period
Ventricular pacing percentage was obtained from pacemaker interrogation and defined as the proportion of ventricular events that were paced during each study period. The outcome was assessed under both closed-loop stimulation (CLS) and accelerometer-based VVIR pacing and expressed as a percentage from 0% to 100%.
At the end of each 1-month pacing period

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Rzeczywisty)

23 czerwca 2025

Zakończenie podstawowe (Rzeczywisty)

27 lipca 2026

Ukończenie studiów (Rzeczywisty)

27 lipca 2026

Daty rejestracji na studia

Pierwszy przesłany

29 sierpnia 2026

Pierwszy przesłany, który spełnia kryteria kontroli jakości

4 września 2026

Pierwszy wysłany (Rzeczywisty)

10 września 2026

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

10 września 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

4 września 2026

Ostatnia weryfikacja

1 września 2026

Więcej informacji

Terminy związane z tym badaniem

Plan dla danych uczestnika indywidualnego (IPD)

Planujesz udostępniać dane poszczególnych uczestników (IPD)?

NIE

Opis planu IPD

There is currently no plan to make individual participant data publicly available.

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Nie

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .