Comparison of the Concordance Between Digital Marking and Manual Marking for Toric Intraocular Lens Positioning
Comparison of the Concordance Between Digital Image-Guided Navigation and Manual Peripheral Corneal Marking for Toric Intraocular Lens Positioning
研究概览
详细说明
Patients with visually significant cataract and regular corneal astigmatism who were scheduled for phacoemulsification were screened for eligibility. Inclusion criteria were adults (18 -90 years of age) that were candidates for uncomplicated cataract surgery. The exclusion criteria were irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems.
Preoperative assessment included best-corrected visual acuity (BCVA), which were determined by Snellen's chart and converted to logMAR. Spherical equivalent (SE) and astigmatic refraction (AR) were examined by Full Auto Ref-Keratometer (RK-F2, Canon). Evaluation for corneal astigmatism and axial length with ARGOS® Biometer. Slit lamp examination (BP 900®, Hagg-Streit) intraocular pressure measurement (Canon TX-20P), and dilated retinal examination was performed on all of the patients.
For manual technique a slit-lamp was used. Patients were seated with their head fixed with the horizontal supporting band and their chin placed in the center of the chinrest. A thin horizontal slit beam is projected onto the cornea and moved vertically (up and down) until the area of maximum light concentration (brightest reflection) on the corneal surface is identified. For each eye, after identifying the point of maximum light reflex, under topical anesthesia with 0.5% proparacaine hydrochloride eye drops, the corneal limbus was marked at the 0° and 180° meridian (3 and 9 o'clock) positions using a blunt tip of the Sinskey hook coated with sterile gentian violet ink, using light pressure as a single touch of approximately one second, in order to standardize ink deposition and minimize ink spread. The marking was maintained horizontally without any movement during the marking. The thickness of marks was routinely less than 0.25 mm
For the digital technique, preoperative biometry data obtained from the ARGOS Biometer were entered into the Alcon Vision Planner v1.6, which was paired with the VERION Image-Guided System (Alcon Laboratories, Inc., Fort Worth, Texas, USA) to project the digital overlays directly into the operating microscope. This module captures a high-resolution preoperative reference image of the eye which can be used to document the center of the undilated pupil, corneal reflex position or eccentricity of the visual axis, scleral vessels and iris structures. All marking was performed by the same experienced surgeon (W. Tattiyakul) who was experienced with using all techniques, performed all procedures and measurements. No complications had occurred.
With both the manual and digital marks visible intraoperatively, the angular deviation along the horizontal meridian (in degrees) between the manual mark and the digital reference was recorded from the operating-microscope display for each eye by two trained observers independently, each blinded to the other's readings, using the same standardized display. The mean of the two observers' readings was used for all subsequent analyses, and inter-observer agreement was assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analysis
研究类型
注册 (估计的)
参与标准
资格标准
适合学习的年龄
- 成人
- 年长者
接受健康志愿者
取样方法
研究人群
描述
Inclusion Criteria:
- candidates for uncomplicated cataract surgery
Exclusion Criteria:
- irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems
学习计划
研究是如何设计的?
设计细节
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Mean signed angular deviation in degrees of the manual slit-lamp horizontal reference mark from the VERION digital reference meridian
大体时间:6 months
|
For each eye, the signed angular difference (degrees) between the horizontal reference meridian (0-180°) marked by the manual slit-lamp peripheral corneal marking technique and the corresponding meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
Clockwise deviation is coded positive and counterclockwise negative.
Agreement is summarized across eyes as the mean signed angular deviation with 95% Bland-Altman limits of agreement.
Reported in a single unit (degrees).
|
6 months
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Proportion of eyes within ±5° of the VERION digital reference meridian, measured intraoperatively
大体时间:6 months
|
For each eye, the absolute angular difference (degrees) between the horizontal reference meridian marked by the manual slit-lamp peripheral corneal marking technique and the meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
The outcome is the proportion of eyes with an absolute deviation of 5° or less, reported as a percentage (%) of eyes assessed.
±5° is the pre-specified threshold of clinically acceptable agreement for the marking step.
|
6 months
|
合作者和调查者
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- MTU-EC-OP-0-162/69
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.