Comparison of the Concordance Between Digital Marking and Manual Marking for Toric Intraocular Lens Positioning
Comparison of the Concordance Between Digital Image-Guided Navigation and Manual Peripheral Corneal Marking for Toric Intraocular Lens Positioning
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Descrição detalhada
Patients with visually significant cataract and regular corneal astigmatism who were scheduled for phacoemulsification were screened for eligibility. Inclusion criteria were adults (18 -90 years of age) that were candidates for uncomplicated cataract surgery. The exclusion criteria were irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems.
Preoperative assessment included best-corrected visual acuity (BCVA), which were determined by Snellen's chart and converted to logMAR. Spherical equivalent (SE) and astigmatic refraction (AR) were examined by Full Auto Ref-Keratometer (RK-F2, Canon). Evaluation for corneal astigmatism and axial length with ARGOS® Biometer. Slit lamp examination (BP 900®, Hagg-Streit) intraocular pressure measurement (Canon TX-20P), and dilated retinal examination was performed on all of the patients.
For manual technique a slit-lamp was used. Patients were seated with their head fixed with the horizontal supporting band and their chin placed in the center of the chinrest. A thin horizontal slit beam is projected onto the cornea and moved vertically (up and down) until the area of maximum light concentration (brightest reflection) on the corneal surface is identified. For each eye, after identifying the point of maximum light reflex, under topical anesthesia with 0.5% proparacaine hydrochloride eye drops, the corneal limbus was marked at the 0° and 180° meridian (3 and 9 o'clock) positions using a blunt tip of the Sinskey hook coated with sterile gentian violet ink, using light pressure as a single touch of approximately one second, in order to standardize ink deposition and minimize ink spread. The marking was maintained horizontally without any movement during the marking. The thickness of marks was routinely less than 0.25 mm
For the digital technique, preoperative biometry data obtained from the ARGOS Biometer were entered into the Alcon Vision Planner v1.6, which was paired with the VERION Image-Guided System (Alcon Laboratories, Inc., Fort Worth, Texas, USA) to project the digital overlays directly into the operating microscope. This module captures a high-resolution preoperative reference image of the eye which can be used to document the center of the undilated pupil, corneal reflex position or eccentricity of the visual axis, scleral vessels and iris structures. All marking was performed by the same experienced surgeon (W. Tattiyakul) who was experienced with using all techniques, performed all procedures and measurements. No complications had occurred.
With both the manual and digital marks visible intraoperatively, the angular deviation along the horizontal meridian (in degrees) between the manual mark and the digital reference was recorded from the operating-microscope display for each eye by two trained observers independently, each blinded to the other's readings, using the same standardized display. The mean of the two observers' readings was used for all subsequent analyses, and inter-observer agreement was assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analysis
Tipo de estudo
Inscrição (Estimado)
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
- Adulto
- Adulto mais velho
Aceita Voluntários Saudáveis
Método de amostragem
População do estudo
Descrição
Inclusion Criteria:
- candidates for uncomplicated cataract surgery
Exclusion Criteria:
- irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Mean signed angular deviation in degrees of the manual slit-lamp horizontal reference mark from the VERION digital reference meridian
Prazo: 6 months
|
For each eye, the signed angular difference (degrees) between the horizontal reference meridian (0-180°) marked by the manual slit-lamp peripheral corneal marking technique and the corresponding meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
Clockwise deviation is coded positive and counterclockwise negative.
Agreement is summarized across eyes as the mean signed angular deviation with 95% Bland-Altman limits of agreement.
Reported in a single unit (degrees).
|
6 months
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Proportion of eyes within ±5° of the VERION digital reference meridian, measured intraoperatively
Prazo: 6 months
|
For each eye, the absolute angular difference (degrees) between the horizontal reference meridian marked by the manual slit-lamp peripheral corneal marking technique and the meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
The outcome is the proportion of eyes with an absolute deviation of 5° or less, reported as a percentage (%) of eyes assessed.
±5° is the pre-specified threshold of clinically acceptable agreement for the marking step.
|
6 months
|
Colaboradores e Investigadores
Patrocinador
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Estimado)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Outros números de identificação do estudo
- MTU-EC-OP-0-162/69
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .