Comparison of the Concordance Between Digital Marking and Manual Marking for Toric Intraocular Lens Positioning

13 juli 2026 uppdaterad av: Pirapat Chareonboonsit, Thammasat University

Comparison of the Concordance Between Digital Image-Guided Navigation and Manual Peripheral Corneal Marking for Toric Intraocular Lens Positioning

This cross-sectional comparative study evaluated patients from a single cataract center. In each eye, manual horizontal meridian axis line was marked on the cornea using the manual peripheral corneal marking technique (Horizontal Slit Beam Marking ), while the Verion digital image-guided navigation system provided a reference axis fixed at the horizontal meridian (0°/180°). The primary outcome measure was the difference in degrees between the digital marker and the manual marker horizontal meridian as measured intraoperatively. Agreement was assessed using Bland-Altman analysis, reporting the mean bias (systematic error) and 95% limits of agreement (LoA, mean ± 1.96 SD). A misalignment of ±10° was defined as the threshold of clinical significance.

Studieöversikt

Status

Har inte rekryterat ännu

Betingelser

Intervention / Behandling

Detaljerad beskrivning

Patients with visually significant cataract and regular corneal astigmatism who were scheduled for phacoemulsification were screened for eligibility. Inclusion criteria were adults (18 -90 years of age) that were candidates for uncomplicated cataract surgery. The exclusion criteria were irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems.

Preoperative assessment included best-corrected visual acuity (BCVA), which were determined by Snellen's chart and converted to logMAR. Spherical equivalent (SE) and astigmatic refraction (AR) were examined by Full Auto Ref-Keratometer (RK-F2, Canon). Evaluation for corneal astigmatism and axial length with ARGOS® Biometer. Slit lamp examination (BP 900®, Hagg-Streit) intraocular pressure measurement (Canon TX-20P), and dilated retinal examination was performed on all of the patients.

For manual technique a slit-lamp was used. Patients were seated with their head fixed with the horizontal supporting band and their chin placed in the center of the chinrest. A thin horizontal slit beam is projected onto the cornea and moved vertically (up and down) until the area of maximum light concentration (brightest reflection) on the corneal surface is identified. For each eye, after identifying the point of maximum light reflex, under topical anesthesia with 0.5% proparacaine hydrochloride eye drops, the corneal limbus was marked at the 0° and 180° meridian (3 and 9 o'clock) positions using a blunt tip of the Sinskey hook coated with sterile gentian violet ink, using light pressure as a single touch of approximately one second, in order to standardize ink deposition and minimize ink spread. The marking was maintained horizontally without any movement during the marking. The thickness of marks was routinely less than 0.25 mm

For the digital technique, preoperative biometry data obtained from the ARGOS Biometer were entered into the Alcon Vision Planner v1.6, which was paired with the VERION Image-Guided System (Alcon Laboratories, Inc., Fort Worth, Texas, USA) to project the digital overlays directly into the operating microscope. This module captures a high-resolution preoperative reference image of the eye which can be used to document the center of the undilated pupil, corneal reflex position or eccentricity of the visual axis, scleral vessels and iris structures. All marking was performed by the same experienced surgeon (W. Tattiyakul) who was experienced with using all techniques, performed all procedures and measurements. No complications had occurred.

With both the manual and digital marks visible intraoperatively, the angular deviation along the horizontal meridian (in degrees) between the manual mark and the digital reference was recorded from the operating-microscope display for each eye by two trained observers independently, each blinded to the other's readings, using the same standardized display. The mean of the two observers' readings was used for all subsequent analyses, and inter-observer agreement was assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analysis

Studietyp

Observationell

Inskrivning (Beräknad)

100

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

  • Vuxen
  • Äldre vuxen

Tar emot friska volontärer

Nej

Testmetod

Icke-sannolikhetsprov

Studera befolkning

Cataract Patients at Thammasat University

Beskrivning

Inclusion Criteria:

  • candidates for uncomplicated cataract surgery

Exclusion Criteria:

  • irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Mean signed angular deviation in degrees of the manual slit-lamp horizontal reference mark from the VERION digital reference meridian
Tidsram: 6 months
For each eye, the signed angular difference (degrees) between the horizontal reference meridian (0-180°) marked by the manual slit-lamp peripheral corneal marking technique and the corresponding meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively. Clockwise deviation is coded positive and counterclockwise negative. Agreement is summarized across eyes as the mean signed angular deviation with 95% Bland-Altman limits of agreement. Reported in a single unit (degrees).
6 months

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Proportion of eyes within ±5° of the VERION digital reference meridian, measured intraoperatively
Tidsram: 6 months
For each eye, the absolute angular difference (degrees) between the horizontal reference meridian marked by the manual slit-lamp peripheral corneal marking technique and the meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively. The outcome is the proportion of eyes with an absolute deviation of 5° or less, reported as a percentage (%) of eyes assessed. ±5° is the pre-specified threshold of clinically acceptable agreement for the marking step.
6 months

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Beräknad)

10 juli 2026

Primärt slutförande (Beräknad)

30 september 2026

Avslutad studie (Beräknad)

30 oktober 2026

Studieregistreringsdatum

Först inskickad

8 juli 2026

Först inskickad som uppfyllde QC-kriterierna

13 juli 2026

Första postat (Faktisk)

17 juli 2026

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

17 juli 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

13 juli 2026

Senast verifierad

1 juli 2026

Mer information

Termer relaterade till denna studie

Andra studie-ID-nummer

  • MTU-EC-OP-0-162/69

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .