- ICH GCP
- USA klinikai vizsgálatok nyilvántartása
- Klinikai vizsgálat NCT07710963
Comparison of the Concordance Between Digital Marking and Manual Marking for Toric Intraocular Lens Positioning
Comparison of the Concordance Between Digital Image-Guided Navigation and Manual Peripheral Corneal Marking for Toric Intraocular Lens Positioning
A tanulmány áttekintése
Állapot
Körülmények
Beavatkozás / kezelés
Részletes leírás
Patients with visually significant cataract and regular corneal astigmatism who were scheduled for phacoemulsification were screened for eligibility. Inclusion criteria were adults (18 -90 years of age) that were candidates for uncomplicated cataract surgery. The exclusion criteria were irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems.
Preoperative assessment included best-corrected visual acuity (BCVA), which were determined by Snellen's chart and converted to logMAR. Spherical equivalent (SE) and astigmatic refraction (AR) were examined by Full Auto Ref-Keratometer (RK-F2, Canon). Evaluation for corneal astigmatism and axial length with ARGOS® Biometer. Slit lamp examination (BP 900®, Hagg-Streit) intraocular pressure measurement (Canon TX-20P), and dilated retinal examination was performed on all of the patients.
For manual technique a slit-lamp was used. Patients were seated with their head fixed with the horizontal supporting band and their chin placed in the center of the chinrest. A thin horizontal slit beam is projected onto the cornea and moved vertically (up and down) until the area of maximum light concentration (brightest reflection) on the corneal surface is identified. For each eye, after identifying the point of maximum light reflex, under topical anesthesia with 0.5% proparacaine hydrochloride eye drops, the corneal limbus was marked at the 0° and 180° meridian (3 and 9 o'clock) positions using a blunt tip of the Sinskey hook coated with sterile gentian violet ink, using light pressure as a single touch of approximately one second, in order to standardize ink deposition and minimize ink spread. The marking was maintained horizontally without any movement during the marking. The thickness of marks was routinely less than 0.25 mm
For the digital technique, preoperative biometry data obtained from the ARGOS Biometer were entered into the Alcon Vision Planner v1.6, which was paired with the VERION Image-Guided System (Alcon Laboratories, Inc., Fort Worth, Texas, USA) to project the digital overlays directly into the operating microscope. This module captures a high-resolution preoperative reference image of the eye which can be used to document the center of the undilated pupil, corneal reflex position or eccentricity of the visual axis, scleral vessels and iris structures. All marking was performed by the same experienced surgeon (W. Tattiyakul) who was experienced with using all techniques, performed all procedures and measurements. No complications had occurred.
With both the manual and digital marks visible intraoperatively, the angular deviation along the horizontal meridian (in degrees) between the manual mark and the digital reference was recorded from the operating-microscope display for each eye by two trained observers independently, each blinded to the other's readings, using the same standardized display. The mean of the two observers' readings was used for all subsequent analyses, and inter-observer agreement was assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analysis
Tanulmány típusa
Beiratkozás (Becsült)
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
- Felnőtt
- Idősebb felnőtt
Egészséges önkénteseket fogad
Mintavételi módszer
Tanulmányi populáció
Leírás
Inclusion Criteria:
- candidates for uncomplicated cataract surgery
Exclusion Criteria:
- irregular corneal astigmatism, lens subluxation, uveitis, glaucoma, trauma, history of retinal disease or optic neuropathy, previous intraocular or corneal surgery and corneal disease that could not be evaluated by the ARGOS or VERION systems
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Mean signed angular deviation in degrees of the manual slit-lamp horizontal reference mark from the VERION digital reference meridian
Időkeret: 6 months
|
For each eye, the signed angular difference (degrees) between the horizontal reference meridian (0-180°) marked by the manual slit-lamp peripheral corneal marking technique and the corresponding meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
Clockwise deviation is coded positive and counterclockwise negative.
Agreement is summarized across eyes as the mean signed angular deviation with 95% Bland-Altman limits of agreement.
Reported in a single unit (degrees).
|
6 months
|
Másodlagos eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Proportion of eyes within ±5° of the VERION digital reference meridian, measured intraoperatively
Időkeret: 6 months
|
For each eye, the absolute angular difference (degrees) between the horizontal reference meridian marked by the manual slit-lamp peripheral corneal marking technique and the meridian identified by the VERION Image-Guided System (reference standard) is measured intraoperatively.
The outcome is the proportion of eyes with an absolute deviation of 5° or less, reported as a percentage (%) of eyes assessed.
±5° is the pre-specified threshold of clinically acceptable agreement for the marking step.
|
6 months
|
Együttműködők és nyomozók
Szponzor
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete (Becsült)
Elsődleges befejezés (Becsült)
A tanulmány befejezése (Becsült)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Tényleges)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
Egyéb vizsgálati azonosító számok
- MTU-EC-OP-0-162/69
Gyógyszer- és eszközinformációk, tanulmányi dokumentumok
Egy amerikai FDA által szabályozott gyógyszerkészítményt tanulmányoz
Egy amerikai FDA által szabályozott eszközterméket tanulmányoz
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .