KV-PREDICT Diagnostic Device in Patients With Heart Failure (PREDICT_HF)
Single-Centre, Interventional, Non-Invasive Clinical Study to Evaluate the Performance of the KV-PREDICT Diagnostic Medical Device in Patients With Heart Failure
研究概览
详细说明
Background and rationale Heart failure is a chronic condition in which periodic clinical assessment is important for follow-up. Non-invasive tools that can be used quickly during routine outpatient visits are of interest as a complement to established assessment methods. KV-PREDICT is an investigational diagnostic medical device (Class IIa, not CE-marked) developed to acquire cardiac electrical and mechanical signals non-invasively. Before any evaluation of clinical or predictive performance, it is first necessary to confirm that the device, in its current prototype configuration, can reliably acquire signals of adequate quality across patients with different degrees of heart failure severity.
Objective and nature of the study This is a single-center, non-invasive, interventional pilot study of exploratory nature. Its purpose is limited to assessing the feasibility, quality and reproducibility of the signals acquired by KV-PREDICT. The study does not evaluate the diagnostic or predictive performance of the device, is not designed to support any clinical decision, and does not modify the clinical management of participants in any way.
Study conduct Eligible patients are enrolled during a scheduled outpatient follow-up visit, after providing written informed consent. Each participant undergoes a single, brief signal-acquisition session performed with the device applied externally to the chest, over clothing, without any invasive procedure and without interfering with the standard clinical visit. No dedicated follow-up visits are required for the purposes of this study.
Data handling and analysis The acquired signals are used exclusively for research purposes and are analyzed offline, after the acquisition session, outside the clinical setting. No automated interpretation or clinical output is generated by the device during the visit. Analyses are descriptive and exploratory, consistent with the pilot nature of the study.
Expected value and risks No direct clinical benefit to individual participants is expected. Because the procedure is non-invasive and brief, the anticipated risks and burden for participants are minimal. The results are intended to inform the subsequent development of the device.
研究类型
注册 (估计的)
阶段
- 不适用
联系人和位置
学习联系方式
- 姓名:Gabriele Pieraccini, MSc / Eng.
- 电话号码:+39 351 5548005
- 邮箱:gabriele.pieraccini@kardiovis.com
学习地点
-
-
Emilia-Romagna
-
Modena、Emilia-Romagna、意大利、41124
- S.C. Cardiologia, Azienda Ospedaliero-Universitaria Policlinico di Modena
-
接触:
- Study Principal Investigator
- 电话号码:+39 059 4225463
- 邮箱:giuseppe.boriani@unimore.it
-
-
参与标准
资格标准
适合学习的年龄
- 成人
- 年长者
接受健康志愿者
描述
Inclusion Criteria:
- Age 18 years or older
- Clinical diagnosis of heart failure
- NYHA functional class I, II, or III
- Able to understand the study information and provide written informed consent
- Attending a scheduled outpatient follow-up visit at the site
Exclusion Criteria:
- Acute clinical instability at the time of the visit
- Conditions that, in the investigator's judgment, could interfere with correct performance of the acquisition procedure
- Inability to provide valid informed consent
- Concurrent participation in other studies that could interfere with the aims of this investigation
学习计划
研究是如何设计的?
设计细节
- 主要用途:诊断
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:KV-PREDICT signal acquisition
Single group of adult heart failure patients undergoing one non-invasive KV-PREDICT signal acquisition during a scheduled outpatient visit.
|
KV-PREDICT is applied externally to the chest, over clothing (no skin contact required), during a scheduled outpatient visit.
It performs a single acquisition of about 10 seconds, simultaneously recording a single-lead ECG and cardiac mechanical vibration signals.
The embedded software only acquires, filters, synchronizes and transmits the signals; it produces no automated clinical output or interpretation.
During the same session, an external, standalone reference pulse oximeter (photoplethysmograph), which is not connected to and not part of the investigational device, is recorded separately and used only for offline comparison during signal analysis.
All signals are analyzed offline for research purposes only.
|
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Quality and suitability of KV-PREDICT recordings for extraction of mechanical acceleration signals
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number and corresponding percentage of technically valid and processable acceleration signal acquisitions measured in mG, across patients with different heart-failure severity
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Quality and suitability of KV-PREDICT recordings for extraction of mechanical gyroscopic signals
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number and corresponding percentage of technically valid and processable gyroscopic signal acquisitions measured in mDeg/s, across patients with different heart-failure severity
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Variability of acquired acceleration signals' amplitude across NYHA classes and left-ventricular function profiles
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of amplitude of acceleration signals measured in mG across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired gyroscopic signals' amplitude across NYHA classes and left-ventricular function profiles
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of amplitude of gyroscopic signals' amplitude measured in mDeg/s across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired acceleration signals' timing across NYHA classes and left-ventricular function profiles
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of timing of acceleration signals measured in ms across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired gyroscopic signals' timing across NYHA classes and left-ventricular function profiles
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of timing of gyroscopic signals measured in ms across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Adverse events or complaints related to the procedure for data acquisition
大体时间:During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number of adverse events or complaints collected during the recording sessions
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
合作者和调查者
调查人员
- 首席研究员:Giuseppe Boriani, MD, PhD、Azienda Ospedaliero-Universitaria di Modena
出版物和有用的链接
一般刊物
- Shah SJ, Nakamura K, Marcus GM, Gerber IL, McKeown BH, Jordan MV, Huddleston M, Foster E, Michaels AD. Association of the fourth heart sound with increased left ventricular end-diastolic stiffness. J Card Fail. 2008 Jun;14(5):431-6. doi: 10.1016/j.cardfail.2008.01.010. Epub 2008 May 27.
- Jafari Tadi M, Lehtonen E, Saraste A, Tuominen J, Koskinen J, Teras M, Airaksinen J, Pankaala M, Koivisto T. Gyrocardiography: A New Non-invasive Monitoring Method for the Assessment of Cardiac Mechanics and the Estimation of Hemodynamic Variables. Sci Rep. 2017 Jul 28;7(1):6823. doi: 10.1038/s41598-017-07248-y.
- Alhakak AS, Olsen FJ, Skaarup KG, Lassen MCH, Johansen ND, Jorgensen PG, Abildgaard U, Jensen GB, Schnohr P, Sogaard P, Mogelvang R, Biering-Sorensen T. Age- and sex-based normal reference ranges of the cardiac time intervals: the Copenhagen City Heart Study. Clin Res Cardiol. 2025 Apr;114(4):430-442. doi: 10.1007/s00392-023-02269-2. Epub 2023 Jul 31.
- Cao M, Gardner RS, Hariharan R, Nair DG, Schulze C, An Q, Thakur PH, Kwan B, Zhang Y, Boehmer JP. Ambulatory Monitoring of Heart Sounds via an Implanted Device Is Superior to Auscultation for Prediction of Heart Failure Events. J Card Fail. 2020 Feb;26(2):151-159. doi: 10.1016/j.cardfail.2019.10.006. Epub 2019 Oct 18.
- Boriani G, Imberti JF, Bonini N, Carriere C, Mei DA, Zecchin M, Piccinin F, Vitolo M, Sinagra G. Remote multiparametric monitoring and management of heart failure patients through cardiac implantable electronic devices. Eur J Intern Med. 2023 Sep;115:1-9. doi: 10.1016/j.ejim.2023.04.011. Epub 2023 Apr 17.
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新