KV-PREDICT Diagnostic Device in Patients With Heart Failure (PREDICT_HF)
Single-Centre, Interventional, Non-Invasive Clinical Study to Evaluate the Performance of the KV-PREDICT Diagnostic Medical Device in Patients With Heart Failure
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Descrição detalhada
Background and rationale Heart failure is a chronic condition in which periodic clinical assessment is important for follow-up. Non-invasive tools that can be used quickly during routine outpatient visits are of interest as a complement to established assessment methods. KV-PREDICT is an investigational diagnostic medical device (Class IIa, not CE-marked) developed to acquire cardiac electrical and mechanical signals non-invasively. Before any evaluation of clinical or predictive performance, it is first necessary to confirm that the device, in its current prototype configuration, can reliably acquire signals of adequate quality across patients with different degrees of heart failure severity.
Objective and nature of the study This is a single-center, non-invasive, interventional pilot study of exploratory nature. Its purpose is limited to assessing the feasibility, quality and reproducibility of the signals acquired by KV-PREDICT. The study does not evaluate the diagnostic or predictive performance of the device, is not designed to support any clinical decision, and does not modify the clinical management of participants in any way.
Study conduct Eligible patients are enrolled during a scheduled outpatient follow-up visit, after providing written informed consent. Each participant undergoes a single, brief signal-acquisition session performed with the device applied externally to the chest, over clothing, without any invasive procedure and without interfering with the standard clinical visit. No dedicated follow-up visits are required for the purposes of this study.
Data handling and analysis The acquired signals are used exclusively for research purposes and are analyzed offline, after the acquisition session, outside the clinical setting. No automated interpretation or clinical output is generated by the device during the visit. Analyses are descriptive and exploratory, consistent with the pilot nature of the study.
Expected value and risks No direct clinical benefit to individual participants is expected. Because the procedure is non-invasive and brief, the anticipated risks and burden for participants are minimal. The results are intended to inform the subsequent development of the device.
Tipo de estudo
Inscrição (Estimado)
Estágio
- Não aplicável
Contactos e Locais
Contato de estudo
- Nome: Gabriele Pieraccini, MSc / Eng.
- Número de telefone: +39 351 5548005
- E-mail: gabriele.pieraccini@kardiovis.com
Locais de estudo
-
-
Emilia-Romagna
-
Modena, Emilia-Romagna, Itália, 41124
- S.C. Cardiologia, Azienda Ospedaliero-Universitaria Policlinico di Modena
-
Contato:
- Study Principal Investigator
- Número de telefone: +39 059 4225463
- E-mail: giuseppe.boriani@unimore.it
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
- Adulto
- Adulto mais velho
Aceita Voluntários Saudáveis
Descrição
Inclusion Criteria:
- Age 18 years or older
- Clinical diagnosis of heart failure
- NYHA functional class I, II, or III
- Able to understand the study information and provide written informed consent
- Attending a scheduled outpatient follow-up visit at the site
Exclusion Criteria:
- Acute clinical instability at the time of the visit
- Conditions that, in the investigator's judgment, could interfere with correct performance of the acquisition procedure
- Inability to provide valid informed consent
- Concurrent participation in other studies that could interfere with the aims of this investigation
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Diagnóstico
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Experimental: KV-PREDICT signal acquisition
Single group of adult heart failure patients undergoing one non-invasive KV-PREDICT signal acquisition during a scheduled outpatient visit.
|
KV-PREDICT is applied externally to the chest, over clothing (no skin contact required), during a scheduled outpatient visit.
It performs a single acquisition of about 10 seconds, simultaneously recording a single-lead ECG and cardiac mechanical vibration signals.
The embedded software only acquires, filters, synchronizes and transmits the signals; it produces no automated clinical output or interpretation.
During the same session, an external, standalone reference pulse oximeter (photoplethysmograph), which is not connected to and not part of the investigational device, is recorded separately and used only for offline comparison during signal analysis.
All signals are analyzed offline for research purposes only.
|
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Quality and suitability of KV-PREDICT recordings for extraction of mechanical acceleration signals
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number and corresponding percentage of technically valid and processable acceleration signal acquisitions measured in mG, across patients with different heart-failure severity
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Quality and suitability of KV-PREDICT recordings for extraction of mechanical gyroscopic signals
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number and corresponding percentage of technically valid and processable gyroscopic signal acquisitions measured in mDeg/s, across patients with different heart-failure severity
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Variability of acquired acceleration signals' amplitude across NYHA classes and left-ventricular function profiles
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of amplitude of acceleration signals measured in mG across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired gyroscopic signals' amplitude across NYHA classes and left-ventricular function profiles
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of amplitude of gyroscopic signals' amplitude measured in mDeg/s across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired acceleration signals' timing across NYHA classes and left-ventricular function profiles
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of timing of acceleration signals measured in ms across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Variability of acquired gyroscopic signals' timing across NYHA classes and left-ventricular function profiles
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Variability of timing of gyroscopic signals measured in ms across NYHA classes and left-ventricular function profiles
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
|
Adverse events or complaints related to the procedure for data acquisition
Prazo: During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Number of adverse events or complaints collected during the recording sessions
|
During the single signal-acquisition session at the scheduled study visit (about 10 seconds of recording)
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Giuseppe Boriani, MD, PhD, Azienda Ospedaliero-Universitaria di Modena
Publicações e links úteis
Publicações Gerais
- Shah SJ, Nakamura K, Marcus GM, Gerber IL, McKeown BH, Jordan MV, Huddleston M, Foster E, Michaels AD. Association of the fourth heart sound with increased left ventricular end-diastolic stiffness. J Card Fail. 2008 Jun;14(5):431-6. doi: 10.1016/j.cardfail.2008.01.010. Epub 2008 May 27.
- Jafari Tadi M, Lehtonen E, Saraste A, Tuominen J, Koskinen J, Teras M, Airaksinen J, Pankaala M, Koivisto T. Gyrocardiography: A New Non-invasive Monitoring Method for the Assessment of Cardiac Mechanics and the Estimation of Hemodynamic Variables. Sci Rep. 2017 Jul 28;7(1):6823. doi: 10.1038/s41598-017-07248-y.
- Alhakak AS, Olsen FJ, Skaarup KG, Lassen MCH, Johansen ND, Jorgensen PG, Abildgaard U, Jensen GB, Schnohr P, Sogaard P, Mogelvang R, Biering-Sorensen T. Age- and sex-based normal reference ranges of the cardiac time intervals: the Copenhagen City Heart Study. Clin Res Cardiol. 2025 Apr;114(4):430-442. doi: 10.1007/s00392-023-02269-2. Epub 2023 Jul 31.
- Cao M, Gardner RS, Hariharan R, Nair DG, Schulze C, An Q, Thakur PH, Kwan B, Zhang Y, Boehmer JP. Ambulatory Monitoring of Heart Sounds via an Implanted Device Is Superior to Auscultation for Prediction of Heart Failure Events. J Card Fail. 2020 Feb;26(2):151-159. doi: 10.1016/j.cardfail.2019.10.006. Epub 2019 Oct 18.
- Boriani G, Imberti JF, Bonini N, Carriere C, Mei DA, Zecchin M, Piccinin F, Vitolo M, Sinagra G. Remote multiparametric monitoring and management of heart failure patients through cardiac implantable electronic devices. Eur J Intern Med. 2023 Sep;115:1-9. doi: 10.1016/j.ejim.2023.04.011. Epub 2023 Apr 17.
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Estimado)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- PREDICT-HF
- 26-06-058156 (Outro identificador: EUDAMED (European Database on Medical Devices))
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .