Pediatric Lead Extractability and Survival Evaluation (PLEASE)

March 17, 2026 updated by: Charles Berul

Pediatric Lead Extractability and Survival Evaluation (PLEASE)Study: A Prospective Assessment of Implantable Cardioverter Defibrillator (ICD) Lead Extractability and Survival in Children and Congenital Heart Disease Patients

This is a randomized, prospective clinical trial comparing 2 different types of implantable cardioverter defibrillator (ICD) leads in children and patients with congenital heart disease. ICD lead survival in this patient group is particularly suboptimal, and lead extraction is technically difficult and carries a substantial morbidity risk. Recently, improved ICD lead designs have been released and are currently being utilized in patients. The main aim of the study is to determine if either type of lead performs better in terms of implantation electrical characteristics, long-term survival without breaking, and ease of extractability.

Study Overview

Detailed Description

The recognition that ICD therapy in children can be life-saving, as well as the development of smaller devices and leads, are resulting in increased ICD implantation in the pediatric and young adult populations. However, ICD lead survival in this patient group is suboptimal. Lead extraction is technically difficult and has substantial morbidity. Recently, improved ICD lead designs have been released and are currently being utilized. While the development of transvenous ICD leads has allowed for less invasive implantation, the long-term presence of these leads carries some risks. When a lead becomes infected or causes patient morbidity, it must be removed. Because of the larger caliber of ICD leads and the presence of uninsulated shocking coils, non-functional ICD leads usually warrant removal to avoid future mechanical or electrical lead interactions. The larger size of ICD leads is of concern in this younger population, particularly with regard to avoiding the accumulation of implanted hardware. The principal indication for lead revision in children and young adults is lead malfunction, commonly related to patient growth and increased physical activity in this younger population. The removal of chronic indwelling pacer or ICD leads is not trivial. Scar tissue begins to bind transvenous leads to the endothelial surface several weeks after implantation, and this fibrous adherence progresses over time, exacerbated by high-energy shock delivery. In children and younger adult ICD recipients, exuberant lead scarring is prevalent, particularly in the regions of venous entry, high voltage shocking coils, and the lead tip. Transvenous lead extraction has been facilitated by the use of telescoping sheaths and powered sheaths that can be advanced over the lead to disrupt scar tissue and free the lead from the endothelium.

The 2 ICD lead types to be compared (thin ICD leads versus ePTFE-coated leads) in the proposed study are at clinical equipoise. Both types are believed to be improved over prior generation ICD leads. However, a direct comparison has not been performed. As the pediatric and congenital heart disease patient subgroup is younger than the average ICD patient population, they will have greater likelihood of long-term survival, and therefore, the issues of lead survival, durability, and extractability are critically important. While these themes are important for all ICD patients, they are particularly germane for younger patients who have more active lifestyles and are, in general, expected to outlive their ICD leads.

This study will prospectively assess the improvements in ICD lead design, specifically comparing a Gore ePTFE coating and thinner caliber ICD leads in pediatric and congenital heart disease population.

Specific Aim #1: To prospectively evaluate the long-term performance of modern-generation ICD leads in pediatric and congenital heart disease patients.

Specific Aim #2: To determine the potential beneficial effects of two improvements in ICD lead design - Gore ePTFE coil coating and thinner lead diameter in pediatric and congenital heart disease patients.

Specific Aim #3: To directly compare the safety and efficacy of these new ICD lead technologies in terms of extractability in pediatric and congenital heart disease patients.

The primary hypothesis is that improved lead designs and smaller diameter leads will demonstrate benefit in survival and extractability. The primary outcome variables will include: a) ICD lead functionality and performance by subtype, b) inappropriate shocks due to lead issues, c) lead extractability by subtype (Gore-coated versus thin leads).

Secondary outcomes include comparison of implant electrical parameters, patient morbidity and mortality, comparing each new lead type with each other and against historical pediatric controls with standard transvenous ICD leads.

Study Type

Interventional

Enrollment (Actual)

748

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Massachusetts
      • Boston, Massachusetts, United States, 02115
        • Children's Hospital Boston

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

1 year to 55 years (Child, Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Patients will be eligible for study enrollment at the time of implantation of a new ICD and transvenous ICD lead.
  • All patients under the age of 21 years at the time of initial ICD implantation, or those patients with congenital heart disease and an indication for ICD implantation are eligible for participation.

Exclusion Criteria:

  • Patients who have existing ICD leads in place.
  • Patients with expected survival less than 1 year.
  • Patients who cannot be expected to participate in follow-up visits.
  • Patients receiving an epicardial or subcutaneous ICD (without transvenous ICD lead).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Thin leads
Thin (less than or equal to 7 French introducer) isodiametric ICD leads
Thin (less than or equal to 7 French introducer) isodiametric ICD leads
Active Comparator: Gore PTFE-coated
ICD lead with PTFE-coated coils
Randomization of ICD lead type at implant

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
ICD Lead Functionality and Performance by Subtype
Time Frame: 5 years
Lead survival
5 years
Compare Ease of Lead Extractability by Subtype
Time Frame: 5 years
Extraction techniques include simple traction (no locking stylet or sheath) and advanced extraction (locking stylet with a sheath). Compared number of extractions using each technique for single and ePTFE-coated coils.
5 years

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Patients With Major Complications
Time Frame: 5 years
Major complications experienced by patients undergoing lead extraction.
5 years
Comparison of Inappropriate Shocks
Time Frame: 5 years
Comparison of inappropriate shocks between thin leads and PTFE-coated leads.
5 years

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Charles I Berul, MD, Boston Children's Hospital

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

June 1, 2006

Primary Completion (Actual)

January 1, 2009

Study Completion (Actual)

January 1, 2009

Study Registration Dates

First Submitted

June 6, 2006

First Submitted That Met QC Criteria

June 6, 2006

First Posted (Estimated)

June 8, 2006

Study Record Updates

Last Update Posted (Actual)

March 19, 2026

Last Update Submitted That Met QC Criteria

March 17, 2026

Last Verified

March 1, 2026

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Subscribe