Comparison of Stent Graft, Sirolimus Stent, and Bare Metal Stent Implanted in Patients With Acute Coronary Syndrome

March 26, 2007 updated by: University of Zagreb

Comparison of Stent Graft, Sirolimus Stent, and Bare Metal Stent Implanted in Patients With Acute Coronary Syndrome: Clinical and Angiographic Follow-up

During the 6-month period 119 patients with acute coronary syndrome, were randomized to either stent graft group (n=40), sirolimus eluting stent group (n=39), or bare metal stent group (n=40). Demographic, angiographic and procedural characteristics were similar for all three groups. The incidence of 6-month major adverse coronary events were analysed.

Study Overview

Status

Completed

Detailed Description

Background: Percutaneous coronary intervention with stent implantation is a standard therapy for patients with acute coronary syndrome. The in-stent restenosis is still a problem. Recently, drug eluting stents reduce the incidence of this unfavorable event. The primary role of the polytetrafluoroethylene stent graft (PTFE) is management of coronary perforations, closure of coronary aneurysms, and in degenerated saphenous vein grafts. We compared these stents in native coronary vessels in patients with acute coronary syndrome with sirolimus and bare metal stents, for possible reduction of in-stent restenosis.

Methods and results: During the 6-month period 119 patients with acute coronary syndrome, were randomized to either stent graft group (n=40), sirolimus eluting stent group (n=39), or bare metal stent group (n=40). Demographic, angiographic and procedural characteristics were similar for all three groups. The incidence of 6-month major adverse coronary events was similar in all three groups. The target lesion revascularisation was higher in the bare metal stent group (P=0.044). The primary end-point, restenosis rate at six-month follow-up was higher in the bare metal stent group, compared with the stent graft and sirolimus eluting stent groups. The percent diameter stenosis in follow-up was significantly higher in bare metal stent group (P=0.005). The late loss was significantly lower in the sirolimus eluting stent group (0.23 mm), compared with the bare metal stent group (P= 0.034). There was a trend of lower late loss in the stent graft group, compared with bare metal stent group.

Conclusion: Three groups of stents implanted in patients with acute coronary syndrome (stent-graft, sirolimus and bare metal), did not differ regarding the incidence of major adverse cardiac events. Sirolimus-eluting stents had a lower incidence of in-stent restenosis in comparison with bare metal stent group. Stent graft implanted in native coronary arteries appears to be safe and efficient in patients with acute coronary syndrome, but a significant reduction of in-stent restenosis was not achieved.

Study Type

Interventional

Enrollment

100

Phase

  • Phase 4

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Zagreb, Croatia, 10000
        • University Hospital Zagreb

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

20 years to 80 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • stent implantation in acute coronary syndrome

Exclusion Criteria:

  • previous percutaneous coronary intervention or coronary artery bypass graft surgery
  • multivessel, diffuse disease, tortuous vessel
  • arteries less than 3 mm in diameter
  • distal stenosis location
  • left main and bifurcation lesions

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Three groups of stents implanted in patients with acute coronary syndrome (stent-graft, sirolimus and bare metal), did not differ regarding the incidence of major adverse cardiac events.

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Maja Strozzi, MD, University Hospital Zagreb

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

March 1, 2004

Study Completion (Actual)

November 1, 2004

Study Registration Dates

First Submitted

March 26, 2007

First Submitted That Met QC Criteria

March 26, 2007

First Posted (Estimate)

March 27, 2007

Study Record Updates

Last Update Posted (Estimate)

March 27, 2007

Last Update Submitted That Met QC Criteria

March 26, 2007

Last Verified

March 1, 2007

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Subscribe