Comparison of Two Rotational Alignment Methods for Astigmatism Correction in SMILE Surgery

Comparison of Astigmatic Correction Outcomes Following SMILE With OcuLign-Assisted Cyclotorsion Control and Manual Limbal Marking: A Prospective Randomized Contralateral-Eye Noninferiority Study

Small incision lenticule extraction (SMILE) is a commonly used corneal refractive procedure for correcting myopia and myopic astigmatism. During astigmatism correction, eye rotation between the upright examination position and the supine surgical position may affect the alignment of the astigmatic treatment axis and may contribute to residual astigmatism after surgery.

Manual limbal marking is a commonly used method to guide cyclotorsion adjustment during refractive surgery. OcuLign is an assisted cyclotorsion control function of the VISUMAX 800 platform that helps identify ocular rotation and guide axis alignment during SMILE. However, direct clinical evidence comparing OcuLign-assisted cyclotorsion control with manual limbal marking during SMILE is limited.

This study is a single-center, prospective, randomized, masked, contralateral-eye non-inferiority trial. Adults aged 18 to 40 years with stable myopia and astigmatism who plan to undergo bilateral SMILE surgery will be enrolled. In each participant, one eye will receive SMILE with OcuLign-assisted cyclotorsion control, and the fellow eye will receive SMILE with manual limbal marking-guided cyclotorsion adjustment. The treatment assigned to the right eye will be determined by randomization.

The primary objective is to compare residual refractive astigmatism at 3 months after surgery between the two methods. Secondary outcomes include visual acuity, refractive predictability and stability, astigmatism vector analysis, higher-order aberrations, contrast sensitivity, patient-reported quality of vision, overall satisfaction, and safety outcomes. Participants will be followed for approximately 6 months after surgery.

Studieoversigt

Detaljeret beskrivelse

Accurate alignment of the astigmatic treatment axis is an important factor in the correction of myopic astigmatism during small incision lenticule extraction (SMILE). Cyclotorsion may occur when a patient changes from the upright preoperative examination position to the supine surgical position, and may also be influenced by fixation, head position, and suction during surgery. Inadequate compensation for ocular rotation may reduce astigmatic correction accuracy and contribute to postoperative residual astigmatism.

Manual limbal marking is a commonly used method for cyclotorsion adjustment in corneal refractive surgery. This method is clinically feasible and widely used, but it depends on preoperative marking and intraoperative manual judgment. Marking clarity, ocular surface conditions, patient cooperation, and surgeon experience may affect the accuracy and reproducibility of the adjustment. OcuLign is an assisted cyclotorsion control function of the VISUMAX 800 platform that helps identify ocular rotation and guide axis alignment during SMILE. This study is designed to compare OcuLign-assisted cyclotorsion control with manual limbal marking-guided cyclotorsion adjustment during SMILE.

This trial uses a prospective, randomized, contralateral-eye non-inferiority design. Each participant will undergo bilateral SMILE surgery. One eye will be assigned to OcuLign-assisted cyclotorsion control, and the fellow eye will be assigned to manual limbal marking-guided cyclotorsion adjustment, according to a prespecified randomization sequence. This paired-eye design is intended to reduce inter-individual variability in refractive characteristics, healing response, and subjective visual perception. Apart from the method of cyclotorsion adjustment, surgical parameters, lenticule design, suction procedure, postoperative medication, and follow-up schedule will be kept consistent between the two eyes.

Participants and postoperative outcome assessors will be masked to the eye-level treatment assignment whenever feasible. The operating surgeon cannot be masked because the assigned cyclotorsion adjustment method must be performed during surgery. Postoperative assessments will be conducted at predefined follow-up visits through approximately 6 months after surgery.

The primary comparison will evaluate whether OcuLign-assisted cyclotorsion control is non-inferior to manual limbal marking-guided adjustment in terms of residual refractive astigmatism at 3 months after surgery. Secondary assessments will evaluate visual acuity, refractive predictability and stability, astigmatism vector parameters, higher-order aberrations, contrast sensitivity, patient-reported quality of vision, satisfaction, and safety. Safety monitoring will include loss of corrected distance visual acuity and postoperative adverse events.

Undersøgelsestype

Interventionel

Tilmelding (Anslået)

52

Fase

  • Ikke anvendelig

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiesteder

    • Guangdong
      • Guangzhou, Guangdong, Kina, 510060
        • Zhongshan Ophthalmic Center

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

  • Voksen

Tager imod sunde frivillige

Ingen

Beskrivelse

Inclusion Criteria:

  1. Age range: 18 - 40 years old. Patients with myopic astigmatism and stable refractive status for >= 1 year.
  2. Normal corneal morphology in both eyes, with corneal thickness and other ocular parameters meeting the requirements for SMILE surgery.
  3. Preoperative corrected distance visual acuity (CDVA) >= 1.0 in both eyes.
  4. Manifest refractive cylinder >= 0.75 D in both eyes, with an interocular difference in absolute cylinder power <= 1.00 D.
  5. Both eyes can provide reliable preoperative upright-position image information required for astigmatic axis recognition, and are judged by the investigator to meet the requirements for OcuLign cyclotorsion recognition and axis alignment on the VISUMAX 800 platform.
  6. Voluntary to undergo bilateral SMILE surgery on the VISUMAX 800 platform, understanding the random eye-assignment scheme and signing the informed consent form.

Exclusion Criteria:

  1. History of prior ocular surgery, significant ocular trauma, or presence of ocular diseases that may affect the safety of refractive surgery or the interpretation of study outcomes.
  2. Irregular astigmatism or poor repeatability of bilateral astigmatism measurements, making reliable astigmatic axis analysis impossible.
  3. Presence of strabismus, nystagmus, significant binocular vision abnormality, or any condition that may affect fixation or cooperation during examinations.
  4. Severe systemic diseases that may affect ocular healing or surgical safety.
  5. Females who are pregnant or lactating.
  6. Inability to complete the required postoperative follow-up visits or questionnaire assessments.

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Primært formål: Behandling
  • Tildeling: Randomiseret
  • Interventionel model: Parallel tildeling
  • Maskning: Dobbelt

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Eksperimentel: OcuLign-Assisted Cyclotorsion Control
One eye of each participant will undergo SMILE using OcuLign-assisted cyclotorsion control on the VISUMAX 800 platform. OcuLign will be used intraoperatively to assist in identifying ocular rotation and guiding astigmatic axis alignment. Apart from the cyclotorsion adjustment method, the surgical parameters, lenticule design, suction procedure, postoperative medication, and follow-up schedule will be the same as those used for the fellow eye.
One eye of each participant will undergo SMILE using device-assisted cyclotorsion control with the OcuLign function of the VISUMAX 800 platform. During surgery, OcuLign will assist in identifying ocular rotation and guiding astigmatic axis alignment. Apart from the cyclotorsion adjustment method, the surgical parameters, lenticule design, suction procedure, postoperative medication, and follow-up schedule will be the same as those used for the fellow eye.
Andre navne:
  • OcuLign-assisted cyclotorsion control
  • VISUMAX 800 OcuLign
Aktiv komparator: Manual Limbal Marking-Guided Adjustment
The fellow eye of each participant will undergo SMILE using manual limbal marking-guided cyclotorsion adjustment. Limbal reference marks will be made in the upright preoperative position, and intraoperative axis alignment will be manually adjusted according to the marks and the surgical reference direction. Apart from the cyclotorsion adjustment method, the surgical parameters, surgical procedure, postoperative medication, and follow-up schedule will be the same as those used for the contralateral eye.
The fellow eye of each participant will undergo SMILE using manual limbal marking-guided cyclotorsion adjustment. Limbal reference marks will be made in the upright preoperative position, and intraoperative axis alignment will be manually adjusted according to the limbal marks and the surgical reference direction. Apart from the cyclotorsion adjustment method, the surgical parameters, surgical procedure, postoperative medication, and follow-up schedule will be the same as those used for the contralateral eye.

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Residual refractive astigmatism
Tidsramme: 3 months after surgery
Residual refractive astigmatism will be measured as the absolute manifest cylinder power in diopters at 3 months after SMILE. The primary analysis will compare residual refractive astigmatism between eyes treated with OcuLign-assisted cyclotorsion control and eyes treated with manual limbal marking-guided cyclotorsion adjustment to evaluate non-inferiority.
3 months after surgery

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Astigmatic correction predictability
Tidsramme: 1 month, 3 months, and 6 months after surgery
Astigmatic correction predictability will be assessed as the proportion of eyes with postoperative residual manifest cylinder within predefined thresholds, including ≤0.25 D, ≤0.50 D, and ≤1.00 D.
1 month, 3 months, and 6 months after surgery
Astigmatism vector analysis parameters
Tidsramme: 3 months after surgery
Astigmatism correction will be evaluated using vector analysis based on the Alpins method. Parameters will include target-induced astigmatism, surgically induced astigmatism, difference vector, angle of error, magnitude error, correction index, and index of success.
3 months after surgery
Visual acuity
Tidsramme: 1 month, 3 months, and 6 months after surgery
Uncorrected distance visual acuity and corrected distance visual acuity will be measured using standardized visual acuity charts and recorded as logMAR visual acuity. Visual acuity-derived performance will also be summarized using the safety index and efficacy index, calculated from the same visual acuity measurements. The safety index is defined as postoperative corrected distance visual acuity divided by preoperative corrected distance visual acuity, and the efficacy index is defined as postoperative uncorrected distance visual acuity divided by preoperative corrected distance visual acuity.
1 month, 3 months, and 6 months after surgery
Manifest refraction
Tidsramme: 1 month, 3 months, and 6 months after surgery
Postoperative manifest refraction will include sphere, cylinder, and spherical equivalent measured by standard manifest refraction.
1 month, 3 months, and 6 months after surgery
Refractive accuracy
Tidsramme: 1 month, 3 months, and 6 months after surgery
Refractive accuracy will be assessed as the proportion of eyes with postoperative spherical equivalent within predefined ranges of the target refraction, including within ±0.25 D, ±0.50 D, and ±1.00 D.
1 month, 3 months, and 6 months after surgery
Refractive stability
Tidsramme: 1 month to 6 months after surgery
Refractive stability will be assessed as the change in manifest spherical equivalent refraction between postoperative follow-up visits.
1 month to 6 months after surgery
Higher-order aberrations
Tidsramme: 1 month, 3 months, and 6 months after surgery
Higher-order aberrations will be measured using wavefront aberrometry under standardized conditions. Parameters will include total higher-order aberrations, coma aberration, spherical aberration, and trefoil aberration.
1 month, 3 months, and 6 months after surgery
Contrast sensitivity
Tidsramme: 3 months after surgery
Contrast sensitivity will be measured under standardized conditions using the CSV-1000 contrast sensitivity testing system at multiple spatial frequencies, including 3, 6, 12, and 18 cycles per degree. The area under the log contrast sensitivity function may be calculated as a summary measure when appropriate.
3 months after surgery
Quality of Vision score
Tidsramme: 3 months after surgery
Subjective visual symptoms will be assessed using the Quality of Vision questionnaire, which evaluates the frequency, severity, and bothersome nature of visual disturbances such as glare, halos, and double vision. Higher scores indicate worse subjective visual quality.
3 months after surgery
Overall visual satisfaction score
Tidsramme: 3 months after surgery
Overall visual satisfaction will be assessed using a 5-point Likert rating scale. Scores range from 1 to 5, with higher scores indicating greater satisfaction.
3 months after surgery

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Publikationer og nyttige links

Den person, der er ansvarlig for at indtaste oplysninger om undersøgelsen, leverer frivilligt disse publikationer. Disse kan handle om alt relateret til undersøgelsen.

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Anslået)

13. juli 2026

Primær færdiggørelse (Anslået)

30. juni 2027

Studieafslutning (Anslået)

30. juni 2027

Datoer for studieregistrering

Først indsendt

6. juli 2026

Først indsendt, der opfyldte QC-kriterier

6. juli 2026

Først opslået (Faktiske)

10. juli 2026

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

10. juli 2026

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

6. juli 2026

Sidst verificeret

1. juli 2026

Mere information

Begreber relateret til denne undersøgelse

Yderligere relevante MeSH-vilkår

Andre undersøgelses-id-numre

  • IIT2026071

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

INGEN

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ingen

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .

Kliniske forsøg med Nærsynethed

3
Abonner