Fault Detection, Zone MPC and DiAs System in T1D (ZoneMPC)
A Randomized Crossover Study to Evaluate the Efficacy of Fault Detection Algorithms Using the ZoneMPC Algorithm and DiAs System in Adult Subjects With Type 1 Diabetes in the Outpatient Setting
Studienübersicht
Status
Bedingungen
Intervention / Behandlung
Detaillierte Beschreibung
Investigational Device:
Artificial Pancreas System with Fault Detection Algorithms: Roche Accu-Check Spirit Combo Insulin Pump, Dexcom G4P System with Share, Diabetes Assistant (DiAs) on Android phone, DiAs Web Monitoring (DWM)
- referred to as Remote Monitoring Server, MPC control algorithm, Health Monitoring System (HMS) algorithm. Sensor and infusion set fault detection algorithms will be applied offline with data obtained from server and notifications will be sent to the clinician.
Control Arm:
Sensor-augmented insulin pump therapy: Subject will use their personal insulin pump and Dexcom G4P System with Share.
Primary Objective:
To determine the efficacy of the fault detection algorithm. The primary outcome is based on the amount of time the sensor glucose is >250 mg/dL in the 4 hours preceding detection of the infusion set failure during sensor augmented pump therapy vs. closed-loop control with fault detection alerts.
Secondary Objectives:
To determine the effectiveness of the sensor fault detection algorithm. To determine the efficacy of the Zone MPC controller by evaluating glycemic outcomes
Number of Subjects:
There will be 20 subjects recruited: 10 at Stanford and 10 at Denver (up to 36 subjects will be enrolled to reach 20 subjects completing the study)
Diagnosis and Main Inclusion Criteria:
Adult subjects between 18 and 55 years of age inclusive, diagnosed with type 1 diabetes.
Trial Design:
This outpatient study will be conducted over 6 weeks as shown in the figure below. The 6-week period will consist of two 2-week blocks of prolonged infusion set wear with a 1-week sensor run-in period preceding each block. In each block, subjects will wear an infusion set for up to 7 days. A new infusion set will be inserted at the start of each week in the block. Following enrollment procedures, subjects will be randomized at a ratio of 1:1 to either use the AP system with fault detection algorithms (intervention) or sensor-augmented pump therapy (control) in the first block.
Studientyp
Einschreibung (Voraussichtlich)
Phase
- Unzutreffend
Kontakte und Standorte
Studienkontakt
- Name: David Maahs, MD, PhD
- E-Mail: David.Maahs@ucdenver.edu
Studieren Sie die Kontaktsicherung
- Name: Laurel Messer
- E-Mail: Laurel.Messer@ucdenver.edu
Teilnahmekriterien
Zulassungskriterien
Studienberechtigtes Alter
Akzeptiert gesunde Freiwillige
Studienberechtigte Geschlechter
Beschreibung
Inclusion Criteria:
- Clinical diagnosis of type 1 diabetes for at least 12 months
- Daily insulin therapy for at least 12 months
- Age between 18.0 to 55.0 years of age
- Use of an insulin pump for at least 3 months
- Subject comprehends English
- Females of childbearing potential must use an adequate method of contraception and have a negative pregnancy test
- Total daily insulin requirement ≥ 0.3 units/kg/day
- Subject has an adult companion, age > 18 years, who lives with the subject, has access to where they sleep, is willing to be in the house when the subject is sleeping and willing to attend to the subject if there are safety concerns -
Exclusion Criteria:
- Diabetic ketoacidosis in the past 6 months
- Hypoglycemic seizure or loss of consciousness in the past 6 months
- Subjects requiring an intermediate or long-acting insulin (such as NPH, detemir or glargine)
- Subjects using other anti-diabetic medications other than insulin (oral or injectable) at the time of enrollment. Any prior use of other anti-diabetic medications must be washed out for at least 8 weeks prior to enrollment.
- Current use of other medications, which in the judgment of the investigator would be a contraindication to participation in the study
- Subject has a medical disorder that in the judgment of the investigator will affect completion of any aspect of the protocol
- Subject is currently participating in another investigational device or drug study within 30 days or 5-half-lives of the drug.
- Subject has a history of any cardiac or vascular disorder including, but not limited to, myocardial infarction, unstable angina, coronary artery bypass surgery, coronary artery stenting, transient ischemic attack, cerebrovascular accident, angina, congestive heart failure, arrhythmia or thromboembolic disease
- Subject has a history of hepatic disease
- Subject has renal failure on dialysis
- Systolic blood pressure > 160 mmHg on screening visit
- Diastolic blood pressure > 90 mmHg on screening visit
- Subjects with inadequately treated thyroid disease or celiac disease
- Subject has a neurologic disorder that in the judgment of the investigator will affect completion of the protocol
- Subject has received inpatient psychiatric treatment in the past 6 months
- Subject consumes more than an average of 4 standard alcoholic drinks/day in the last 30 days
- Subject has an active skin condition that would affect sensor placement
- Subject is unable to avoid acetaminophen for the duration of the study
- Current use of oral/inhaled glucocorticoids or other medications, which in the judgment of the investigator would be a contraindication to participation in the study
- Subject is currently on beta blocker medication -
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
- Hauptzweck: Behandlung
- Zuteilung: Zufällig
- Interventionsmodell: Crossover-Aufgabe
- Maskierung: Keine (Offenes Etikett)
Waffen und Interventionen
Teilnehmergruppe / Arm |
Intervention / Behandlung |
---|---|
Kein Eingriff: Sensor Augmented Pump (control)
Use sensor augmented pump (SAP) for 3 weeks.
|
|
Experimental: Artificial Pancreas (intervention)
Artificial pancreas system (Algorithm + CGM + pump)--use the AP system for 3 weeks which consists of: (1) Fault detection and Zone MPC algorithm housed on the DiAs platform + (2) Roche insulin pump + (3) Dexcom CGM
|
The AP system using fault detection algorithms will determine whether insulin infusion problems are occurring and may prevent severe hyperglycemia due to its predictive nature
|
Was misst die Studie?
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
---|---|---|
Amount of time sensor glucose levels are >250 mg/dl
Zeitfenster: 4 hours after insulin infusion set failure
|
Amount of time sensor glucose levels are >250 mg/dl in the control arm versus the experimental arm
|
4 hours after insulin infusion set failure
|
Sekundäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
---|---|---|
Effectiveness of sensor fault detection algorithm as defined by % of sensor failures caught by the system
Zeitfenster: During 2 week intervention period
|
Effectiveness of sensor fault detection algorithm as defined by % of sensor failures caught by the system
|
During 2 week intervention period
|
Mean sensor glucose values
Zeitfenster: 2 week intervention period versus 2 week control period
|
mean sensor glucose values
|
2 week intervention period versus 2 week control period
|
Percent of time in range between 70-180 mg/dl
Zeitfenster: 2 week intervention period versus 2 week control period
|
Percent of time in range between 70-180 mg/dl
|
2 week intervention period versus 2 week control period
|
Mitarbeiter und Ermittler
Publikationen und hilfreiche Links
Allgemeine Veröffentlichungen
- Howsmon DP, Baysal N, Buckingham BA, Forlenza GP, Ly TT, Maahs DM, Marcal T, Towers L, Mauritzen E, Deshpande S, Huyett LM, Pinsker JE, Gondhalekar R, Doyle FJ 3rd, Dassau E, Hahn J, Bequette BW. Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas. J Diabetes Sci Technol. 2018 May;12(3):599-607. doi: 10.1177/1932296818755173. Epub 2018 Feb 1.
- Forlenza GP, Deshpande S, Ly TT, Howsmon DP, Cameron F, Baysal N, Mauritzen E, Marcal T, Towers L, Bequette BW, Huyett LM, Pinsker JE, Gondhalekar R, Doyle FJ 3rd, Maahs DM, Buckingham BA, Dassau E. Application of Zone Model Predictive Control Artificial Pancreas During Extended Use of Infusion Set and Sensor: A Randomized Crossover-Controlled Home-Use Trial. Diabetes Care. 2017 Aug;40(8):1096-1102. doi: 10.2337/dc17-0500. Epub 2017 Jun 5. Erratum In: Diabetes Care. 2017 Sep 8;:
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn
Primärer Abschluss (Voraussichtlich)
Studienabschluss (Voraussichtlich)
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Schätzen)
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Schätzen)
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Schlüsselwörter
Zusätzliche relevante MeSH-Bedingungen
Andere Studien-ID-Nummern
- IDE G150122
Plan für individuelle Teilnehmerdaten (IPD)
Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .
Klinische Studien zur Diabetes Typ 1
-
Eledon PharmaceuticalsZurückgezogenSpröder Diabetes mellitus Typ 1Vereinigte Staaten
-
Shanghai Jiao Tong University School of MedicineRekrutierung
-
Hadassah Medical OrganizationUnbekanntUnkontrollierte Typ-1-Diabetiker
-
National Institute of Allergy and Infectious Diseases...PPD; Rho Federal Systems Division, Inc.; Immune Tolerance Network (ITN)AbgeschlossenDiabetes mellitus Typ 1 | T1DM | T1D | Neu aufgetretener Diabetes mellitus Typ 1Vereinigte Staaten, Australien
-
Shanghai Changzheng HospitalRekrutierungSpröder Diabetes mellitus Typ 1China
-
Capillary Biomedical, Inc.AbgeschlossenDiabetes mellitus, Typ 1 | Diabetes Typ 1 | Diabetes mellitus Typ 1 | Diabetes mellitus, insulinabhängig, 1Australien
-
Spiden AGDCB Research AGRekrutierungDiabetes mellitus Typ 1 | Diabetes mellitus Typ 1 mit Hypoglykämie | Diabetes mellitus Typ 1 mit HyperglykämieSchweiz
-
Capillary Biomedical, Inc.BeendetDiabetes Typ 1 | Diabetes mellitus Typ 1 | Diabetes mellitus, Typ I | Diabetes mellitus, insulinabhängig, 1 | IDDMÖsterreich
-
Instytut Diabetologii Sp. z o.o.National Center for Research and Development, Poland; Nalecz Institute of Biocybernetics...UnbekanntDiabetes mellitus Typ 1 mit Hyperglykämie | Diabetes mellitus Typ 1 mit HypoglykämiePolen
-
Poznan University of Medical SciencesUnbekanntDiabetes mellitus Typ 1 | Remission von Typ-1-Diabetes | Chronische Komplikationen von DiabetesPolen