Effectiveness and Safety of the Dexcom™ G4 Continuous Glucose Monitoring System in Pediatric Subjects With Diabetes Mellitus
Effectiveness and Safety of the Dexcom™ G4 Continuous Glucose Monitoring System in Pediatric Subjects With Diabetes Mellitus.
The primary objective of the study is to evaluate the safety and effectiveness of the G4 System (System) when used as an adjuvant to blood glucose testing over a 7-day period in pediatric subjects with diabetes mellitus.
System effectiveness will be primarily evaluated with comparison to laboratory reference Yellow Springs Instrument (YSI) venous sample measurements. The system performance will be evaluated in terms of point and rate accuracy of the system in reference to YSI. The point accuracy is measured as the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at YSI glucose levels <80 mg/dL. The trend accuracy of the device performance, i.e., continuous glucose error grid analysis, will be evaluated as well as the temporal accuracy at different glucose rates of changes and different glucose ranges (hypoglycemic, euglycemic, and hyperglycemic ranges). The primary matched paired (Sensor-YSI) measurements will be collected during one in-clinic session. System effectiveness will also be evaluated with SMBG measurements similarly.
Safety data of the system will also be collected and safety will be characterized by the incidence of Adverse Device Effects (ADE) experienced by study participants.
Other objectives include the evaluation of device performance at an alternate site (the upper buttocks).
Aperçu de l'étude
Statut
Les conditions
Type d'étude
Inscription (Réel)
Contacts et emplacements
Lieux d'étude
-
-
California
-
Escondido, California, États-Unis, 92026
- Advanced Metabolic Care and Research
-
Orange, California, États-Unis, 92868
- CHOC Children's
-
Stanford, California, États-Unis, 94305
- Stanford University
-
-
Colorado
-
Aurora, Colorado, États-Unis, 80045
- Barbara Davis Center
-
-
Idaho
-
Idaho Falls, Idaho, États-Unis, 83404
- Rocky Mountain Diabetes and Osteoporosis Center, PA
-
-
Massachusetts
-
Boston, Massachusetts, États-Unis, 02215
- Joslin Diabetes Center
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Sexes éligibles pour l'étude
Méthode d'échantillonnage
Population étudiée
La description
Inclusion Criteria:
- Ages 2 to 17 years ;
- Diagnosis of diabetes mellitus;
- Willing to participate in one 7-day Sensor wear period;
- Willing to wear 2 systems simultaneously;
- Willing to take a minimum of 7 fingersticks per day during home use with a blood glucose meter provided;
- Willing to attend one in-clinic session of up to 6 hours duration. Clinic procedures where blood sampling via SMBG and/or venous YSI method, depending on age;
- Subject or guardian is able to speak, read, and write English.
Exclusion Criteria:
- Extensive skin changes/diseases at sensor wear site that preclude wearing the Sensor on normal skin (e.g. extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis);
- Known allergy to medical-grade adhesives;
- Require a magnetic resonance imaging (MRI) scan, computed tomography (CT) scan, or diathermy during the wear period. If any of these procedures are required urgently during the study, subjects will notify the study staff, end their continuous glucose monitoring (CGM) session, and remove their sensor.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
Cohortes et interventions
Groupe / Cohorte |
|---|
|
Pediatric subjects with diabetes mellitus
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
System Performance Effectiveness
Délai: 7 days
|
The primary objective is to characterize System effectiveness in comparison with the laboratory reference YSI venous sample measurements.
|
7 days
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
System Performance at Alternate Wear Site
Délai: 7 days
|
An additional objective is to evaluate System performance at an alternate wear site.
The System performance will be assessed, using the same statistical methodology as the primary wear site.
|
7 days
|
Collaborateurs et enquêteurs
Parrainer
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude
Achèvement primaire (Réel)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Estimation)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Estimation)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Mots clés
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- PTL900803
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .