Position and Angle Marking System (PAMS) for Ultrasound-guided Neuraxial Procedures

4 août 2026 mis à jour par: Sng Ban Leong, KK Women's and Children's Hospital

Epidural anaesthesia involves the administration of numbing medication (local anaesthetics) close to the spinal canal, and is regarded as the best method for providing pain relief during labour and childbirth. The use of ultrasound to guide epidural insertion and placement has been shown to improve pain relief and reduce the risk of complications.

However, after identifying the desired needle insertion site via ultrasound, current practice involves demarcating the insertion site using marker pens, which is time-consuming, inaccurate, and may be inadvertently removed by skin sterilisation. Furthermore, the desired needle angulation determined by ultrasound cannot be accurately measured and maintained during the epidural procedure. Both issues limit the benefits of using ultrasound to guide the epidural procedure.

To address these issues, the investigators developed a Position and Angular Marking System (PAMS) that attaches on to the ultrasound probe. When the desired insertion site is identified by ultrasound, the doctor gently presses PAMS into the patient's back to create skin indentations demarcating the needle insertion point. These indentations are not affected by skin sterilisation, and may improve accuracy and reduce the time taken to perform the epidural procedure. Next, the angle between the ultrasound probe and the patient's back can be measured using PAMS, and this angle can be maintained using a needle guide during the epidural procedure.

The purpose of this randomised study is to evaluate the usability of PAMS and to identify areas for further improvement.

This study will involve up to 10 healthy simulated patients, and up to 30 volunteer ultrasound operators. After undergoing a standardised training session with a mannequin, the ultrasound operators will be randomly assigned to perform skin marking on the simulated patients as though they are performing an ultrasound-guided epidural anaesthesia procedure (no needle puncture will be made) using either PAMS, or standard clinical practice with marker pens. After completing the procedure, the ultrasound operators will cross over and perform the procedure again using either PAMS or standard clinical practice with marker pens.

Both the ultrasound operators and simulated patients will be asked to complete satisfaction and usability surveys regarding their experience with the procedures.

Aperçu de l'étude

Description détaillée

The volunteer ultrasound operators are women's anaesthesiologists trained in the use of ultrasound-guided neuraxial procedures.

Simulated patients are healthy volunteers from other departments, without spine abnormalities.

The ultrasound operators will be asked to complete a questionnaire regarding your prior experience with neuraxial ultrasonography and undergo structured training on the use of PAMS using a mannequin. Next, they will be asked to perform the following two procedures (the order at which they will perform the procedures will be randomised):

  1. Ultrasound scan using PAMS

    • After attaching PAMS to the ultrasound probe, they will perform a lumbar ultrasound scan on a simulated patient to identify the desired needle insertion site and gently press PAMS into the simulated patient's back to create skin indentations demarcating the desired needle insertion site.
    • Next, adjust the angulation of the ultrasound probe to optimise the image of the simulated patient's lumbar spine, and read the angle between the ultrasound probe and patient's back off a graduated scale.
    • Remove the ultrasound probe, and position the needle-insertion guide corresponding to the skin indentations. Needle angulation can be simulated by mounting a blunt needle on to the insertion guide.
    • Finally, they will be asked to complete a usability questionnaire.
  2. Ultrasound scan without PAMS (standard clinical practice)

    • Perform a lumbar ultrasound scan to identify the desired needle insertion site, followed by demarcation using skin markers.
    • Next, adjust the angulation of the ultrasound probe to optimise the image of the simulated patient's lumbar spine.
    • Remove the ultrasound probe and simulate needle insertion by touching the tip of a blunt needle to the patient's back, at the desired angle.
    • Finally, they will then be asked to complete a usability questionnaire.

Both procedures are non-invasive (i.e. no skin punctures will be made). No follow up is required for this study.

No new research medications are involved in this study.

Additionally, video recordings may be made of the procedures. Any video or ultrasound images obtained during the course of this study will be stored and analysed only for the purposes of this study for a period not exceeding 7 years, and will be destroyed after completion of the study. Technical data such as the scanning duration and success rate will also be documented.

Type d'étude

Interventionnel

Inscription (Estimé)

30

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

      • Singapore, Singapour, 128038
        • KK Women's and Children's Hospital

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

18 ans et plus (Adulte, Adulte plus âgé)

Accepte les volontaires sains

Oui

La description

Ultrasound operators

Inclusion Criteria:

  • Anaesthesiologists
  • Trained in the use of ultrasound-guided neuraxial procedures

Exclusion Criteria:

  • No experience with ultrasound-guided neuraxial procedures

Simulated patients

Inclusion Criteria:

  • Healthy, no significant medical comorbidity (American Society of Anesthesiologists Physical Status I or II)
  • No spinal abnormalities, or previous spinal surgery

Exclusion Criteria:

  • Significant spinal abnormality
  • Previous spinal surgery

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation croisée
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Comparateur actif: Standard clinical practice
  • Lumbar ultrasound scan will be performed to identify the desired needle insertion site, followed by demarcation using skin markers.
  • The angulation of the ultrasound probe will be adjusted to optimise the image of the simulated patient's lumbar spine.
  • The ultrasound probe will be removed and needle insertion simulated by touching the tip of a blunt needle to the patient's back, at the desired angle.
  • A usability questionnaire regarding the procedure will be completed.
When the desired insertion site is identified by ultrasound, the doctor demarcate the needle insertion point using marker pens.
Expérimental: Position and Angle Marking System (PAMS)
  • PAMS will be attached to the ultrasound probe.
  • Lumbar ultrasound scan will be performed to identify the desired needle insertion site, followed by demarcation by gentle pressing PAMS into the simulated patient's back to create skin indentations.
  • The angulation of the ultrasound probe will be adjusted to optimise the image of the simulated patient's lumbar spine, and the angle read off a graduated scale.
  • The ultrasound probe will be removed and needle insertion simulated by touching the tip of a blunt needle to the patient's back, at the desired angle.
  • A usability questionnaire regarding the procedure will be completed.
Position and Angular Marking System (PAMS) attaches on to the ultrasound probe. When the desired insertion site is identified by ultrasound, the doctor gently presses PAMS into the patient's back to create skin indentations demarcating the needle insertion point. These indentations are not affected by skin sterilisation, and may improve accuracy and reduce the time taken to perform the epidural procedure. Next, the angle between the ultrasound probe and the patient's back can be measured using PAMS, and this angle can be maintained using a needle guide during the epidural procedure.
Autres noms:
  • PAMS

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Satisfaction survey of simulated patients and operators
Délai: Immediately after performing the procedure
Satisfaction and feedback regarding the respective procedures
Immediately after performing the procedure

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Time taken to perform procedure
Délai: Immediately after performing the procedure
Comparison of time taken to perform the interventional procedure (PAMS), versus standard clinical practice (control)
Immediately after performing the procedure

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Publications et liens utiles

La personne responsable de la saisie des informations sur l'étude fournit volontairement ces publications. Il peut s'agir de tout ce qui concerne l'étude.

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

7 septembre 2021

Achèvement primaire (Estimé)

31 décembre 2026

Achèvement de l'étude (Estimé)

31 décembre 2027

Dates d'inscription aux études

Première soumission

5 juillet 2021

Première soumission répondant aux critères de contrôle qualité

30 juillet 2021

Première publication (Réel)

2 août 2021

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

7 août 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

4 août 2026

Dernière vérification

1 août 2026

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • PAMS 2021/2159

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner