Quality and Safety of Real-Time Brachytherapy Treatments: Clinical Implementation of a Multipoint Scintillation Detector for Treatment Monitoring ((CurieInVivo))
In Vivo Surveillance of Brachytherapy Using Plastic Scintillation Dosimetry: A Clinical Feasibility Study
This project addresses a key safety gap in High-Dose-Rate (HDR) brachytherapy, where very high, localized radiation doses mean that small geometric deviations in catheter placement or radioactive source positioning can cause clinically significant errors. Incident reports from the Radiation Oncology Incident Learning System (RO-ILS) of the American Society for Radiation Oncology (ASTRO) and guidelines from the American Association of Physicists in Medicine Task Group 59 (AAPM TG-59) identify these events-often occurring between imaging and treatment delivery-as among the most critical. Unlike external beam radiotherapy, brachytherapy lacks a prior phantom-based verification step, making in vivo dosimetry the only real-time method to confirm delivered dose and detect anomalies in dwell time or source position. International bodies such as the International Atomic Energy Agency (IAEA) recommend this approach for quality management and accident prevention.
Multipoint plastic scintillation detectors (mPSD) offer water-equivalent, real-time measurements and can enable three-dimensional (3D) radioactive source tracking, but their clinical use remains limited due to integration challenges and the absence of well-defined alert thresholds.
The project aims to implement a real-time in vivo monitoring solution using an mPSD at the Centre régional intégré de cancérologie (CRIC) of Lévis, Québec (QC) for HDR gynecological and prostate brachytherapy. Objectives include: (1) demonstrating clinical feasibility and integration within existing clinical workflows; (2) validating the accuracy and precision of dose, dwell-time, and source-position measurements under clinical conditions; and (3) developing practical alert thresholds to identify common deviations. The central hypothesis is that mPSD technology can provide reliable, reproducible real-time measurements without lengthening procedures or disrupting clinical workflow.
The methodology relies on adapting the HYPERSCINT® platform for HDR in vivo use. A custom mPSD, consisting of several scintillators along a single optical fiber, will be insertable into a 6-French (6F) catheter.
Phase 1 focuses on feasibility and integration, including HDR-specific calibration, definition of detector insertion scenarios, integration of acquisition hardware, and staff training.
Phase 2 validates performance in at least 20 patients. Planned treatment parameters and time-stamped detector signals acquired at 1-10 hertz (Hz) will be used to reconstruct delivered dose, dwell times, and source-to-detector distances. Measurement accuracy and reproducibility will be quantified using mixed-effects statistical models.
Phase 3 establishes preliminary alert thresholds based on deviations in dose, time, and position, along with visualization tools such as Bland-Altman plots, although these thresholds will not be applied clinically during the study.
Expected outcomes include demonstrated feasibility of in vivo mPSD monitoring, quantitative performance metrics with confidence intervals, identification of influential factors and operational recommendations, preliminary alert thresholds with reporting templates, and a roadmap for sustainable deployment in Lévis and other centers. Scientific outputs will also be produced for dissemination across the Santé Québec network.
Aperçu de l'étude
Statut
Les conditions
Intervention / Traitement
Type d'étude
Inscription (Estimé)
Phase
- N'est pas applicable
Contacts et emplacements
Coordonnées de l'étude
- Nom: Mathieu Goulet, PhD
- Numéro de téléphone: 1-418-835-7121 ext 14539
- E-mail: mathieu_goulet@ssss.gouv.qc.ca
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
- Adulte
- Adulte plus âgé
Accepte les volontaires sains
La description
Inclusion Criteria:
- Patients receiving prostate brachytherapy treatment, or female user receiving gynecological brachytherapy treatment for cervical cancer.
Exclusion Criteria:
- Patients with a Clostridium difficile infection or any other high-risk positive infection.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Autre
- Répartition: N / A
- Modèle interventionnel: Affectation à un seul groupe
- Masquage: Aucun (étiquette ouverte)
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: In-vivo dosimetry
Patients in this treatment arm will receive the same treatment and care as patients outside of the protocol, but will have their treatment monitored by a plastic scintillation detector placed inside or at close proximity of the tumor.
|
Typical brachytherapy treatments are conducted without dose or source position monitoring in real-time.
This intervention will assess the feasibility of using a scintillation detector to that end.
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Proportion of brachytherapy treatments with successful scintillation detector-based measurement
Délai: From the patient anesthesia at the beginning of the brachytherapy procedure to the end of brachytherapy radiation delivery on the same day.
|
Feasibility will be assessed by determining whether a scintillation detector successfully acquires a measurement during brachytherapy treatment.
Radiation emitted from the brachytherapy source induces optical scintillation in the detector.
The emitted light is transmitted via optical fiber and collected by a spectrometer located outside the treatment room.
A successful measurement is defined as the acquisition of scintillation data that can be analyzed to infer clinically relevant parameters, including either the radiation dose rate as a function of time and/or the radioactive source position during treatment delivery.
The outcome will be reported as the number and proportion of treatments with successful measurements.
|
From the patient anesthesia at the beginning of the brachytherapy procedure to the end of brachytherapy radiation delivery on the same day.
|
Collaborateurs et enquêteurs
Parrainer
Publications et liens utiles
Publications générales
- Yamada Y, Rogers L, Demanes DJ, Morton G, Prestidge BR, Pouliot J, Cohen GN, Zaider M, Ghilezan M, Hsu IC; American Brachytherapy Society. American Brachytherapy Society consensus guidelines for high-dose-rate prostate brachytherapy. Brachytherapy. 2012 Jan-Feb;11(1):20-32. doi: 10.1016/j.brachy.2011.09.008.
- Fonseca GP, Johansen JG, Smith RL, Beaulieu L, Beddar S, Kertzscher G, Verhaegen F, Tanderup K. In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice. Phys Imaging Radiat Oncol. 2020 Sep 28;16:1-11. doi: 10.1016/j.phro.2020.09.002. eCollection 2020 Oct.
- Linares Rosales HM, Duguay-Drouin P, Archambault L, Beddar S, Beaulieu L. Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy. Med Phys. 2019 May;46(5):2412-2421. doi: 10.1002/mp.13498. Epub 2019 Apr 2.
- Linares Rosales HM, Johansen JG, Kertzscher G, Tanderup K, Beaulieu L, Beddar S. 3D source tracking and error detection in HDR using two independent scintillator dosimetry systems. Med Phys. 2021 May;48(5):2095-2107. doi: 10.1002/mp.14607. Epub 2021 Mar 28.
- Therriault-Proulx F, Beddar S, Beaulieu L. On the use of a single-fiber multipoint plastic scintillation detector for 192Ir high-dose-rate brachytherapy. Med Phys. 2013 Jun;40(6):062101. doi: 10.1118/1.4803510.
- Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson JF. High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys. 1998 Apr;25(4):375-403. doi: 10.1118/1.598232.
- Hoopes DJ, Dicker AP, Eads NL, Ezzell GA, Fraass BA, Kwiatkowski TM, Lash K, Patton GA, Piotrowski T, Tomlinson C, Ford EC. RO-ILS: Radiation Oncology Incident Learning System: A report from the first year of experience. Pract Radiat Oncol. 2015 Sep-Oct;5(5):312-318. doi: 10.1016/j.prro.2015.06.009. Epub 2015 Jun 25.
- Therriault-Proulx F, Archambault L, Beaulieu L, Beddar S. Development of a novel multi-point plastic scintillation detector with a single optical transmission line for radiation dose measurement. Phys Med Biol. 2012 Nov 7;57(21):7147-59. doi: 10.1088/0031-9155/57/21/7147. Epub 2012 Oct 12.
- Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Phys Med Biol. 2016 Oct 21;61(20):R305-R343. doi: 10.1088/0031-9155/61/20/R305. Epub 2016 Oct 3.
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Estimé)
Achèvement primaire (Estimé)
Achèvement de l'étude (Estimé)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- 2026-1305
- 2025 (Subvention/contrat des NIH des États-Unis: Faculty of Social Sciences Scientific Grant at the University of Gdańsk)
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Description du régime IPD
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
produit fabriqué et exporté des États-Unis.
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .