- ICH GCP
- Registro de ensaios clínicos dos EUA
- Ensaio Clínico NCT07580651
Quality and Safety of Real-Time Brachytherapy Treatments: Clinical Implementation of a Multipoint Scintillation Detector for Treatment Monitoring ((CurieInVivo))
In Vivo Surveillance of Brachytherapy Using Plastic Scintillation Dosimetry: A Clinical Feasibility Study
This project addresses a key safety gap in High-Dose-Rate (HDR) brachytherapy, where very high, localized radiation doses mean that small geometric deviations in catheter placement or radioactive source positioning can cause clinically significant errors. Incident reports from the Radiation Oncology Incident Learning System (RO-ILS) of the American Society for Radiation Oncology (ASTRO) and guidelines from the American Association of Physicists in Medicine Task Group 59 (AAPM TG-59) identify these events-often occurring between imaging and treatment delivery-as among the most critical. Unlike external beam radiotherapy, brachytherapy lacks a prior phantom-based verification step, making in vivo dosimetry the only real-time method to confirm delivered dose and detect anomalies in dwell time or source position. International bodies such as the International Atomic Energy Agency (IAEA) recommend this approach for quality management and accident prevention.
Multipoint plastic scintillation detectors (mPSD) offer water-equivalent, real-time measurements and can enable three-dimensional (3D) radioactive source tracking, but their clinical use remains limited due to integration challenges and the absence of well-defined alert thresholds.
The project aims to implement a real-time in vivo monitoring solution using an mPSD at the Centre régional intégré de cancérologie (CRIC) of Lévis, Québec (QC) for HDR gynecological and prostate brachytherapy. Objectives include: (1) demonstrating clinical feasibility and integration within existing clinical workflows; (2) validating the accuracy and precision of dose, dwell-time, and source-position measurements under clinical conditions; and (3) developing practical alert thresholds to identify common deviations. The central hypothesis is that mPSD technology can provide reliable, reproducible real-time measurements without lengthening procedures or disrupting clinical workflow.
The methodology relies on adapting the HYPERSCINT® platform for HDR in vivo use. A custom mPSD, consisting of several scintillators along a single optical fiber, will be insertable into a 6-French (6F) catheter.
Phase 1 focuses on feasibility and integration, including HDR-specific calibration, definition of detector insertion scenarios, integration of acquisition hardware, and staff training.
Phase 2 validates performance in at least 20 patients. Planned treatment parameters and time-stamped detector signals acquired at 1-10 hertz (Hz) will be used to reconstruct delivered dose, dwell times, and source-to-detector distances. Measurement accuracy and reproducibility will be quantified using mixed-effects statistical models.
Phase 3 establishes preliminary alert thresholds based on deviations in dose, time, and position, along with visualization tools such as Bland-Altman plots, although these thresholds will not be applied clinically during the study.
Expected outcomes include demonstrated feasibility of in vivo mPSD monitoring, quantitative performance metrics with confidence intervals, identification of influential factors and operational recommendations, preliminary alert thresholds with reporting templates, and a roadmap for sustainable deployment in Lévis and other centers. Scientific outputs will also be produced for dissemination across the Santé Québec network.
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Tipo de estudo
Inscrição (Estimado)
Estágio
- Não aplicável
Contactos e Locais
Contato de estudo
- Nome: Mathieu Goulet, PhD
- Número de telefone: 1-418-835-7121 ext 14539
- E-mail: mathieu_goulet@ssss.gouv.qc.ca
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
- Adulto
- Adulto mais velho
Aceita Voluntários Saudáveis
Descrição
Inclusion Criteria:
- Patients receiving prostate brachytherapy treatment, or female user receiving gynecological brachytherapy treatment for cervical cancer.
Exclusion Criteria:
- Patients with a Clostridium difficile infection or any other high-risk positive infection.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Outro
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Experimental: In-vivo dosimetry
Patients in this treatment arm will receive the same treatment and care as patients outside of the protocol, but will have their treatment monitored by a plastic scintillation detector placed inside or at close proximity of the tumor.
|
Typical brachytherapy treatments are conducted without dose or source position monitoring in real-time.
This intervention will assess the feasibility of using a scintillation detector to that end.
|
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Proportion of brachytherapy treatments with successful scintillation detector-based measurement
Prazo: From the patient anesthesia at the beginning of the brachytherapy procedure to the end of brachytherapy radiation delivery on the same day.
|
Feasibility will be assessed by determining whether a scintillation detector successfully acquires a measurement during brachytherapy treatment.
Radiation emitted from the brachytherapy source induces optical scintillation in the detector.
The emitted light is transmitted via optical fiber and collected by a spectrometer located outside the treatment room.
A successful measurement is defined as the acquisition of scintillation data that can be analyzed to infer clinically relevant parameters, including either the radiation dose rate as a function of time and/or the radioactive source position during treatment delivery.
The outcome will be reported as the number and proportion of treatments with successful measurements.
|
From the patient anesthesia at the beginning of the brachytherapy procedure to the end of brachytherapy radiation delivery on the same day.
|
Colaboradores e Investigadores
Patrocinador
Publicações e links úteis
Publicações Gerais
- Yamada Y, Rogers L, Demanes DJ, Morton G, Prestidge BR, Pouliot J, Cohen GN, Zaider M, Ghilezan M, Hsu IC; American Brachytherapy Society. American Brachytherapy Society consensus guidelines for high-dose-rate prostate brachytherapy. Brachytherapy. 2012 Jan-Feb;11(1):20-32. doi: 10.1016/j.brachy.2011.09.008.
- Fonseca GP, Johansen JG, Smith RL, Beaulieu L, Beddar S, Kertzscher G, Verhaegen F, Tanderup K. In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice. Phys Imaging Radiat Oncol. 2020 Sep 28;16:1-11. doi: 10.1016/j.phro.2020.09.002. eCollection 2020 Oct.
- Linares Rosales HM, Duguay-Drouin P, Archambault L, Beddar S, Beaulieu L. Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy. Med Phys. 2019 May;46(5):2412-2421. doi: 10.1002/mp.13498. Epub 2019 Apr 2.
- Linares Rosales HM, Johansen JG, Kertzscher G, Tanderup K, Beaulieu L, Beddar S. 3D source tracking and error detection in HDR using two independent scintillator dosimetry systems. Med Phys. 2021 May;48(5):2095-2107. doi: 10.1002/mp.14607. Epub 2021 Mar 28.
- Therriault-Proulx F, Beddar S, Beaulieu L. On the use of a single-fiber multipoint plastic scintillation detector for 192Ir high-dose-rate brachytherapy. Med Phys. 2013 Jun;40(6):062101. doi: 10.1118/1.4803510.
- Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson JF. High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys. 1998 Apr;25(4):375-403. doi: 10.1118/1.598232.
- Hoopes DJ, Dicker AP, Eads NL, Ezzell GA, Fraass BA, Kwiatkowski TM, Lash K, Patton GA, Piotrowski T, Tomlinson C, Ford EC. RO-ILS: Radiation Oncology Incident Learning System: A report from the first year of experience. Pract Radiat Oncol. 2015 Sep-Oct;5(5):312-318. doi: 10.1016/j.prro.2015.06.009. Epub 2015 Jun 25.
- Therriault-Proulx F, Archambault L, Beaulieu L, Beddar S. Development of a novel multi-point plastic scintillation detector with a single optical transmission line for radiation dose measurement. Phys Med Biol. 2012 Nov 7;57(21):7147-59. doi: 10.1088/0031-9155/57/21/7147. Epub 2012 Oct 12.
- Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Phys Med Biol. 2016 Oct 21;61(20):R305-R343. doi: 10.1088/0031-9155/61/20/R305. Epub 2016 Oct 3.
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Estimado)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- 2026-1305
- 2025 (Concessão/Contrato do NIH dos EUA: Faculty of Social Sciences Scientific Grant at the University of Gdańsk)
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .