Digital VS Conventional Altered Cast Technique for Mandibular Bilateral Distal Extension RPDs

7 mai 2026 mis à jour par: Kareem Mohammed Reda, Tanta University

Digital VS Conventional Altered Cast Technique for Mandibular Bilateral Distal Extension Removable Partial Denture

This study aims to compare outcomes of digitally created altered casts with conventional altered casts in mandibular bilateral distal-extension cases.

Aperçu de l'étude

Description détaillée

Partially dentate adults represent an increasing proportion of the population, driven by increased life expectancy and a demographic shift from total edentulism toward partial edentulism. Fabrication of removable partial dentures for distal extension edentulous ridges presents significant clinical complexity because the teeth and residual ridge possess fundamentally different degrees of displaceability. Optimum denture support re-quires recording teeth in anatomical form and the mucosa in functional form .

The altered cast impression technique , introduced by Applegate, remains one of the most widely accepted methods for achieving optimal tissue support and accurate denture base positioning relative to tooth abutments . removable partial dentures fabricated using altered cast impression technique demonstrate improved patient comfort and preservation of oral health . However, despite these advantages, debate persists regarding the clinical significance of altered cast impression technique over single-impression techniques .

Recent advances in computer-aided design and computer-aided manufacturing have transformed the removable partial dentures framework, opening new pathways in the fabrication of RPD through additive and subtractive manufacturing . Digital intraoral scanning offers advantages, including reduced dimensional error, improved efficiency, and real-time feedback on impression accuracy and extension [13-16]. However, intraoral scanning techniques are inherently mucostatic and image-based, rendering the recording of functional impression impossible, a limitation acknowledged as necessary for optimal support in distal extension removable partial dentures . Lab scanning of conventionally prepared altered casts pre-pared by traditional methods fails to eliminate inherited errors from cast separation and reassembly.

Recent studies have introduced hybrid digital workflows that combine functional registration of the edentulous ridge with mucostatic scanning of the remaining oral structures in an effort to reproduce conventional clinical procedures while improving efficiency and precision . Most investigations on digital technology in RPD have concentrated on methods for fabricating metal frameworks and evaluating their performance.

This study aims to compare outcomes of digitally created altered casts with conventional altered casts in mandibular bilateral distal-extension cases.

Type d'étude

Interventionnel

Inscription (Réel)

7

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

      • Tanta, Egypte
        • Tanta University

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte

Accepte les volontaires sains

Non

La description

Inclusion Criteria:

  • Patients presented with bilateral, well-formed mandibular distal extension residual ridges covered by keratinized mucosa, with a minimum thickness of 1.5-3 mm .
  • Each participant had a sound first or second premolar serving as the terminal abutment, with a crown-to-root ratio of ≥ 1:2.
  • Comprised fully or partially dentate maxillary arches, an Angle Class I molar relationship, and adequate interarch space (7-8 mm) to accommodate the prosthesis .

Exclusion Criteria:

  • The presence of systemic conditions affecting bone metabolism or healing
  • pathological conditions in the edentulous area (e.g., residual cysts, retained roots, or severe ridge re-sorption);
  • parafunctional habits such as bruxism, clenching, or tongue thrusting
  • pro-nounced anatomical undercuts in the lingual mandible or adjacent to abutment teeth that could impede denture fabrication or insertion.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation parallèle
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: Digital workflow and Digital altered cast group
Digital selective tissue placement was performed within the same Exocad design project using the free-form scan-data tool, paint and pull feature to apply the principles of selective tissue pressure technique . Using free form mesh editing tools, the buccal shelf area opposing teeth 37 and 47 was identified, and controlled tissue ward displacement of the mucosal surface was simulated while monitoring the magnitude of displacement on sectional views and with digital measurement tools. Figure 2B, F The modified mandibular dataset was saved as the digitally altered cast STL (DA), which served as the basis for the digital RPD design and subsequent denture base fabrication.
The RPD design features of fully digitally fabricated RPD illustrated in Figure 2A. Digital selective tissue placement was performed within the same Exocad design project using the free-form scan-data tool, paint and pull feature to apply the principles of selective tissue pressure technique . Using free form mesh editing tools, the buccal shelf area opposing teeth 37 and 47 was identified, and controlled tissue ward displacement of the mucosal surface was simulated while monitoring the magnitude of displacement on sectional views and with digital measurement tools.
Expérimental: Conventional workflow and conventional altered cast group
For the CA cast technique, one of the SLM frameworks (MF2) was adapted on the stone master cast and used as a functional impression tray (Acrostone Dental Manufacturing, Cairo, Egypt) over the distal extension areas. A selec-tive pressure impression was made according to the traditional altered cast protocol, with the distal saddle areas repoured in dental stone to obtain the CA cast. This CA was then digitized using a desktop scanner (Dove Technologies, Suzhou, China) to produce the CA cast STL. Wax rims were used to verify the jaw relationship, and the maxillary cast was mounted using a facebow transfer on a semi adjustable articulator (A7 Plus, Bio-Art Equi-pamentos Odontológicos Ltda., São Carlos, SP, Brazil), while the mandibular cast was mounted in maximum intercuspation.
For the CA cast technique, one of the SLM frameworks (MF2) was adapted on the stone master cast and used as a functional impression tray (Acrostone Dental Manufacturing, Cairo, Egypt) over the distal extension areas. A selec-tive pressure impression was made according to the traditional altered cast protocol, with the distal saddle areas repoured in dental stone to obtain the CA cast. This CA was then digitized using a desktop scanner (Dove Technologies, Suzhou, China) to produce the CA cast STL. Wax rims were used to verify the jaw relationship, and the maxillary cast was mounted using a facebow transfer on a semi adjustable articulator (A7 Plus, Bio-Art Equi-pamentos Odontológicos Ltda., São Carlos, SP, Brazil), while the mandibular cast was mounted in maximum intercuspation. Semi anatomic acrylic teeth (Acrostone Dental Manufacturing, Cairo, Egypt.) were arranged, and wax try ins were evaluated intraorally in a conventional manner prior to processing.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Délai
Assessment of the condition of ridge mucosal covering of the Removable Partial Denture constructed out of digitally altered model.
Délai: at 15 days, 3 months and 6 months .
at 15 days, 3 months and 6 months .

Mesures de résultats secondaires

Mesure des résultats
Délai
Denture base fit by Pressure Indicating Paste
Délai: one month
one month

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

30 mai 2021

Achèvement primaire (Réel)

30 novembre 2024

Achèvement de l'étude (Réel)

30 novembre 2024

Dates d'inscription aux études

Première soumission

1 mai 2026

Première soumission répondant aux critères de contrôle qualité

7 mai 2026

Première publication (Réel)

13 mai 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

13 mai 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

7 mai 2026

Dernière vérification

1 mai 2021

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • R-RP-5-21-10

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner