Cette page a été traduite automatiquement et l'exactitude de la traduction n'est pas garantie. Veuillez vous référer au version anglaise pour un texte source.

SpO2 Verification Study - Philips Nasal Alar With Nellcor OxiMax SpO2 Board (453564624031)

14 juillet 2026 mis à jour par: Philips Clinical & Medical Affairs Global
The primary purpose of the clinical investigation is to verify the SpO2 accuracy of the Philips AlarX5 SpO2 sensor with Nellcor OxiMax SpO2 Board.

Aperçu de l'étude

Statut

Complété

Les conditions

Intervention / Traitement

Description détaillée

To verify the SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) with the Nellcor OxiMax pulse oximetry technology during non-motion conditions over the range of 70-100% SaO2, when compared with the reference SaO2 obtained from CO-Oximetry analysis of arterial blood samples.

Type d'étude

Observationnel

Inscription (Réel)

43

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

    • Colorado
      • Louisville, Colorado, États-Unis, 80027
        • Element Materials Technology

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte

Accepte les volontaires sains

Oui

Méthode d'échantillonnage

Échantillon non probabiliste

Population étudiée

The study population will include a minimum of 36 to 72 healthy adult volunteers.

La description

Inclusion Criteria:

  • Participant must have the ability to understand and provide written informed consent
  • Participant is an adult between 18-50 years of age
  • Participant must be willing and able to comply with study procedures and duration
  • Participant is a non-smoker or who has not smoked within 2 days prior to the study

Exclusion Criteria:

  • Participant is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation (i.e., Raynaud's Syndrome), injury, or physical malformation of fingers, hands, nose, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos (e.g. permanent, Henna) or artificial dyes (e.g. spray tan, artificial tanning lotion) in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
  • Females who are pregnant
  • Females who are trying to get pregnant with confirmation of positive urine pregnancy test
  • Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Participants with known respiratory conditions such as: (self-reported)

    • uncontrolled / severe asthma
    • flu
    • pneumonia / bronchitis
    • shortness of breath / respiratory distress
    • unresolved respiratory or lung surgery
    • emphysema, COPD, lung disease
    • recent COVID with hospitalization
  • Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)

    • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
    • have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
    • chest pain (angina)
    • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
    • previous heart attack
    • blocked artery
    • unexplained shortness of breath
    • congestive heart failure (CHF)
    • history of stroke
    • transient ischemic attack
    • carotid artery disease
    • myocardial ischemia
    • myocardial infarction
    • cardiomyopathy
    • implantable active medical device such as pacemaker or automatic defibrillator
  • Participants with self-reported health conditions

    • diabetes
    • uncontrolled thyroid disease
    • kidney disease / chronic renal impairment,
    • history of seizures (except childhood febrile seizures)
    • epilepsy
    • history of unexplained syncope
    • recent history of frequent migraine headaches
    • recent symptomatic head injury (within the last 2 months) with continued symptoms
    • cancer requiring chemotherapy, radiation, or current treatment - Participants with known clotting disorders (self reported)
    • history of bleeding disorders or personal history of prolonged bleeding from injury
    • history of blood clots
    • hemophilia
    • current use of blood thinner: prescription or daily use of aspirin
    • Sickle Cell Trait or Disease - Participants with self-reported dermatological conditions at sensor application sites
    • severe dermatitis
    • hyperkeratosis
    • nail fungus - Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self reported)
  • Participants with severe allergies to iodine (only applicable if iodine is used)
  • Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine)
  • Participants with allergies to ultrasound gel
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
  • Unwillingness or inability to remove nail polish, nail jewelry, or artificial nails from test digits. Bruised nail bed of test digits.
  • Unwillingness or inability to remove makeup, creams, lotions, sunscreen, or powders from application sites on the nose or ear.
  • Unwillingness or inability to cut/trim fingernail(s) of test digits if determined length will interfere with correct application of the sensor
  • Piercing at application site of the ear or nose sensors that may interfere with correct placement of sensor
  • Participants received colored intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
  • Surgical hardware in pathway of Device Under Test
  • Other known health conditions should be considered upon disclosure

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

Cohortes et interventions

Groupe / Cohorte
Intervention / Traitement
Healthy Volunteers
Participants are healthy, non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years.

The experimental Philips AlarX5 SpO2 sensor will be tested with Nellcor OxiMax pulse oximetry technology. The AlarX5 SpO2 sensor will be connected via an adapter cable to an IntelliVue X3 Patient Monitor integrated with a Nellcor OxiMax SpO2 Board. Within the study, biomedical/health related outcomes as it relates to the healthy volunteers are not evaluated. The data collected will be used to evaluate how accurate the pulse oximeter (DUT) is to the reference.

Devices under test are listed below:

  • Nellcor OxiMax SpO2 Board (453564624031) integrated within Philips IntelliVue X3 Patient Monitor (867030)
  • Philips AlarX5, Experimental SpO2 sensor based on the Nasal Alar design (nasal ala and earlobe placement)
  • Philips M1943NL SpO2 Adapter Cable (3m)

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Délai: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Up to 6 weeks

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Délai: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Up to 6 weeks

Autres mesures de résultats

Mesure des résultats
Description de la mesure
Délai
Evaluation of the pigmentation bias of the AlarX5 SpO2 sensor (nasal ala and earlobe application).
Délai: Up to 6 weeks
Non-disparate performance for the AlarX5 SpO2 sensor.
Up to 6 weeks

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

28 mai 2026

Achèvement primaire (Réel)

29 juin 2026

Achèvement de l'étude (Réel)

29 juin 2026

Dates d'inscription aux études

Première soumission

16 juin 2026

Première soumission répondant aux critères de contrôle qualité

25 juin 2026

Première publication (Réel)

2 juillet 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

16 juillet 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

14 juillet 2026

Dernière vérification

1 juillet 2026

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • SEF-2025-301123

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

INDÉCIS

Description du régime IPD

This is currently under consideration.

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Oui

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner