SpO2 Verification Study - Philips Nasal Alar With Nellcor OxiMax SpO2 Board (453564624031)

14 luglio 2026 aggiornato da: Philips Clinical & Medical Affairs Global
The primary purpose of the clinical investigation is to verify the SpO2 accuracy of the Philips AlarX5 SpO2 sensor with Nellcor OxiMax SpO2 Board.

Panoramica dello studio

Stato

Completato

Condizioni

Intervento / Trattamento

Descrizione dettagliata

To verify the SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) with the Nellcor OxiMax pulse oximetry technology during non-motion conditions over the range of 70-100% SaO2, when compared with the reference SaO2 obtained from CO-Oximetry analysis of arterial blood samples.

Tipo di studio

Osservativo

Iscrizione (Effettivo)

43

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

    • Colorado
      • Louisville, Colorado, Stati Uniti, 80027
        • Element Materials Technology

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Adulto

Accetta volontari sani

Sì

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

The study population will include a minimum of 36 to 72 healthy adult volunteers.

Descrizione

Inclusion Criteria:

  • Participant must have the ability to understand and provide written informed consent
  • Participant is an adult between 18-50 years of age
  • Participant must be willing and able to comply with study procedures and duration
  • Participant is a non-smoker or who has not smoked within 2 days prior to the study

Exclusion Criteria:

  • Participant is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation (i.e., Raynaud's Syndrome), injury, or physical malformation of fingers, hands, nose, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos (e.g. permanent, Henna) or artificial dyes (e.g. spray tan, artificial tanning lotion) in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
  • Females who are pregnant
  • Females who are trying to get pregnant with confirmation of positive urine pregnancy test
  • Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Participants with known respiratory conditions such as: (self-reported)

    • uncontrolled / severe asthma
    • flu
    • pneumonia / bronchitis
    • shortness of breath / respiratory distress
    • unresolved respiratory or lung surgery
    • emphysema, COPD, lung disease
    • recent COVID with hospitalization
  • Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)

    • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
    • have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
    • chest pain (angina)
    • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
    • previous heart attack
    • blocked artery
    • unexplained shortness of breath
    • congestive heart failure (CHF)
    • history of stroke
    • transient ischemic attack
    • carotid artery disease
    • myocardial ischemia
    • myocardial infarction
    • cardiomyopathy
    • implantable active medical device such as pacemaker or automatic defibrillator
  • Participants with self-reported health conditions

    • diabetes
    • uncontrolled thyroid disease
    • kidney disease / chronic renal impairment,
    • history of seizures (except childhood febrile seizures)
    • epilepsy
    • history of unexplained syncope
    • recent history of frequent migraine headaches
    • recent symptomatic head injury (within the last 2 months) with continued symptoms
    • cancer requiring chemotherapy, radiation, or current treatment - Participants with known clotting disorders (self reported)
    • history of bleeding disorders or personal history of prolonged bleeding from injury
    • history of blood clots
    • hemophilia
    • current use of blood thinner: prescription or daily use of aspirin
    • Sickle Cell Trait or Disease - Participants with self-reported dermatological conditions at sensor application sites
    • severe dermatitis
    • hyperkeratosis
    • nail fungus - Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self reported)
  • Participants with severe allergies to iodine (only applicable if iodine is used)
  • Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine)
  • Participants with allergies to ultrasound gel
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
  • Unwillingness or inability to remove nail polish, nail jewelry, or artificial nails from test digits. Bruised nail bed of test digits.
  • Unwillingness or inability to remove makeup, creams, lotions, sunscreen, or powders from application sites on the nose or ear.
  • Unwillingness or inability to cut/trim fingernail(s) of test digits if determined length will interfere with correct application of the sensor
  • Piercing at application site of the ear or nose sensors that may interfere with correct placement of sensor
  • Participants received colored intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
  • Surgical hardware in pathway of Device Under Test
  • Other known health conditions should be considered upon disclosure

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

Coorti e interventi

Gruppo / Coorte
Intervento / Trattamento
Healthy Volunteers
Participants are healthy, non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years.

The experimental Philips AlarX5 SpO2 sensor will be tested with Nellcor OxiMax pulse oximetry technology. The AlarX5 SpO2 sensor will be connected via an adapter cable to an IntelliVue X3 Patient Monitor integrated with a Nellcor OxiMax SpO2 Board. Within the study, biomedical/health related outcomes as it relates to the healthy volunteers are not evaluated. The data collected will be used to evaluate how accurate the pulse oximeter (DUT) is to the reference.

Devices under test are listed below:

  • Nellcor OxiMax SpO2 Board (453564624031) integrated within Philips IntelliVue X3 Patient Monitor (867030)
  • Philips AlarX5, Experimental SpO2 sensor based on the Nasal Alar design (nasal ala and earlobe placement)
  • Philips M1943NL SpO2 Adapter Cable (3m)

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Lasso di tempo: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Up to 6 weeks

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Lasso di tempo: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Up to 6 weeks

Altre misure di risultato

Misura del risultato
Misura Descrizione
Lasso di tempo
Evaluation of the pigmentation bias of the AlarX5 SpO2 sensor (nasal ala and earlobe application).
Lasso di tempo: Up to 6 weeks
Non-disparate performance for the AlarX5 SpO2 sensor.
Up to 6 weeks

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

28 maggio 2026

Completamento primario (Effettivo)

29 giugno 2026

Completamento dello studio (Effettivo)

29 giugno 2026

Date di iscrizione allo studio

Primo inviato

16 giugno 2026

Primo inviato che soddisfa i criteri di controllo qualità

25 giugno 2026

Primo Inserito (Effettivo)

2 luglio 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

16 luglio 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

14 luglio 2026

Ultimo verificato

1 luglio 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • SEF-2025-301123

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

INDECISO

Descrizione del piano IPD

This is currently under consideration.

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

Sì

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Sottoscrivi