- ICH GCP
- US Clinical Trials Registry
- Klinisk utprøving NCT07680530
SpO2 Verification Study - Philips Nasal Alar With Nellcor OxiMax SpO2 Board (453564624031)
Studieoversikt
Detaljert beskrivelse
Studietype
Registrering (Faktiske)
Kontakter og plasseringer
Studiesteder
-
-
Colorado
-
Louisville, Colorado, Forente stater, 80027
- Element Materials Technology
-
-
Deltakelseskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Voksen
Tar imot friske frivillige
Prøvetakingsmetode
Studiepopulasjon
Beskrivelse
Inclusion Criteria:
- Participant must have the ability to understand and provide written informed consent
- Participant is an adult between 18-50 years of age
- Participant must be willing and able to comply with study procedures and duration
- Participant is a non-smoker or who has not smoked within 2 days prior to the study
Exclusion Criteria:
- Participant is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation (i.e., Raynaud's Syndrome), injury, or physical malformation of fingers, hands, nose, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos (e.g. permanent, Henna) or artificial dyes (e.g. spray tan, artificial tanning lotion) in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
- Females who are pregnant
- Females who are trying to get pregnant with confirmation of positive urine pregnancy test
- Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
Participants with known respiratory conditions such as: (self-reported)
- uncontrolled / severe asthma
- flu
- pneumonia / bronchitis
- shortness of breath / respiratory distress
- unresolved respiratory or lung surgery
- emphysema, COPD, lung disease
- recent COVID with hospitalization
Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
- high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
- have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
- chest pain (angina)
- heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- implantable active medical device such as pacemaker or automatic defibrillator
Participants with self-reported health conditions
- diabetes
- uncontrolled thyroid disease
- kidney disease / chronic renal impairment,
- history of seizures (except childhood febrile seizures)
- epilepsy
- history of unexplained syncope
- recent history of frequent migraine headaches
- recent symptomatic head injury (within the last 2 months) with continued symptoms
- cancer requiring chemotherapy, radiation, or current treatment - Participants with known clotting disorders (self reported)
- history of bleeding disorders or personal history of prolonged bleeding from injury
- history of blood clots
- hemophilia
- current use of blood thinner: prescription or daily use of aspirin
- Sickle Cell Trait or Disease - Participants with self-reported dermatological conditions at sensor application sites
- severe dermatitis
- hyperkeratosis
- nail fungus - Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self reported)
- Participants with severe allergies to iodine (only applicable if iodine is used)
- Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine)
- Participants with allergies to ultrasound gel
- Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
- Unwillingness or inability to remove nail polish, nail jewelry, or artificial nails from test digits. Bruised nail bed of test digits.
- Unwillingness or inability to remove makeup, creams, lotions, sunscreen, or powders from application sites on the nose or ear.
- Unwillingness or inability to cut/trim fingernail(s) of test digits if determined length will interfere with correct application of the sensor
- Piercing at application site of the ear or nose sensors that may interfere with correct placement of sensor
- Participants received colored intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
- Surgical hardware in pathway of Device Under Test
- Other known health conditions should be considered upon disclosure
Studieplan
Hvordan er studiet utformet?
Designdetaljer
Kohorter og intervensjoner
Gruppe / Kohort |
Intervensjon / Behandling |
|---|---|
|
Healthy Volunteers
Participants are healthy, non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years.
|
The experimental Philips AlarX5 SpO2 sensor will be tested with Nellcor OxiMax pulse oximetry technology. The AlarX5 SpO2 sensor will be connected via an adapter cable to an IntelliVue X3 Patient Monitor integrated with a Nellcor OxiMax SpO2 Board. Within the study, biomedical/health related outcomes as it relates to the healthy volunteers are not evaluated. The data collected will be used to evaluate how accurate the pulse oximeter (DUT) is to the reference. Devices under test are listed below:
|
Hva måler studien?
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Tidsramme: Up to 6 weeks
|
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
|
Up to 6 weeks
|
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Tidsramme: Up to 6 weeks
|
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
|
Up to 6 weeks
|
Andre resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Evaluation of the pigmentation bias of the AlarX5 SpO2 sensor (nasal ala and earlobe application).
Tidsramme: Up to 6 weeks
|
Non-disparate performance for the AlarX5 SpO2 sensor.
|
Up to 6 weeks
|
Samarbeidspartnere og etterforskere
Studierekorddatoer
Studer hoveddatoer
Studiestart (Faktiske)
Primær fullføring (Faktiske)
Studiet fullført (Faktiske)
Datoer for studieregistrering
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Faktiske)
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Andre studie-ID-numre
- SEF-2025-301123
Plan for individuelle deltakerdata (IPD)
Planlegger du å dele individuelle deltakerdata (IPD)?
IPD-planbeskrivelse
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert medikamentprodukt
Studerer et amerikansk FDA-regulert enhetsprodukt
produkt produsert i og eksportert fra USA
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .