Denne siden ble automatisk oversatt og nøyaktigheten av oversettelsen er ikke garantert. Vennligst referer til engelsk versjon for en kildetekst.

SpO2 Verification Study - Philips Nasal Alar With Nellcor OxiMax SpO2 Board (453564624031)

14. juli 2026 oppdatert av: Philips Clinical & Medical Affairs Global
The primary purpose of the clinical investigation is to verify the SpO2 accuracy of the Philips AlarX5 SpO2 sensor with Nellcor OxiMax SpO2 Board.

Studieoversikt

Status

Fullført

Forhold

Intervensjon / Behandling

Detaljert beskrivelse

To verify the SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) with the Nellcor OxiMax pulse oximetry technology during non-motion conditions over the range of 70-100% SaO2, when compared with the reference SaO2 obtained from CO-Oximetry analysis of arterial blood samples.

Studietype

Observasjonsmessig

Registrering (Faktiske)

43

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

    • Colorado
      • Louisville, Colorado, Forente stater, 80027
        • Element Materials Technology

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

  • Voksen

Tar imot friske frivillige

Ja

Prøvetakingsmetode

Ikke-sannsynlighetsprøve

Studiepopulasjon

The study population will include a minimum of 36 to 72 healthy adult volunteers.

Beskrivelse

Inclusion Criteria:

  • Participant must have the ability to understand and provide written informed consent
  • Participant is an adult between 18-50 years of age
  • Participant must be willing and able to comply with study procedures and duration
  • Participant is a non-smoker or who has not smoked within 2 days prior to the study

Exclusion Criteria:

  • Participant is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation (i.e., Raynaud's Syndrome), injury, or physical malformation of fingers, hands, nose, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos (e.g. permanent, Henna) or artificial dyes (e.g. spray tan, artificial tanning lotion) in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
  • Females who are pregnant
  • Females who are trying to get pregnant with confirmation of positive urine pregnancy test
  • Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Participants with known respiratory conditions such as: (self-reported)

    • uncontrolled / severe asthma
    • flu
    • pneumonia / bronchitis
    • shortness of breath / respiratory distress
    • unresolved respiratory or lung surgery
    • emphysema, COPD, lung disease
    • recent COVID with hospitalization
  • Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)

    • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
    • have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
    • chest pain (angina)
    • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
    • previous heart attack
    • blocked artery
    • unexplained shortness of breath
    • congestive heart failure (CHF)
    • history of stroke
    • transient ischemic attack
    • carotid artery disease
    • myocardial ischemia
    • myocardial infarction
    • cardiomyopathy
    • implantable active medical device such as pacemaker or automatic defibrillator
  • Participants with self-reported health conditions

    • diabetes
    • uncontrolled thyroid disease
    • kidney disease / chronic renal impairment,
    • history of seizures (except childhood febrile seizures)
    • epilepsy
    • history of unexplained syncope
    • recent history of frequent migraine headaches
    • recent symptomatic head injury (within the last 2 months) with continued symptoms
    • cancer requiring chemotherapy, radiation, or current treatment - Participants with known clotting disorders (self reported)
    • history of bleeding disorders or personal history of prolonged bleeding from injury
    • history of blood clots
    • hemophilia
    • current use of blood thinner: prescription or daily use of aspirin
    • Sickle Cell Trait or Disease - Participants with self-reported dermatological conditions at sensor application sites
    • severe dermatitis
    • hyperkeratosis
    • nail fungus - Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self reported)
  • Participants with severe allergies to iodine (only applicable if iodine is used)
  • Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine)
  • Participants with allergies to ultrasound gel
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
  • Unwillingness or inability to remove nail polish, nail jewelry, or artificial nails from test digits. Bruised nail bed of test digits.
  • Unwillingness or inability to remove makeup, creams, lotions, sunscreen, or powders from application sites on the nose or ear.
  • Unwillingness or inability to cut/trim fingernail(s) of test digits if determined length will interfere with correct application of the sensor
  • Piercing at application site of the ear or nose sensors that may interfere with correct placement of sensor
  • Participants received colored intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
  • Surgical hardware in pathway of Device Under Test
  • Other known health conditions should be considered upon disclosure

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

Kohorter og intervensjoner

Gruppe / Kohort
Intervensjon / Behandling
Healthy Volunteers
Participants are healthy, non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years.

The experimental Philips AlarX5 SpO2 sensor will be tested with Nellcor OxiMax pulse oximetry technology. The AlarX5 SpO2 sensor will be connected via an adapter cable to an IntelliVue X3 Patient Monitor integrated with a Nellcor OxiMax SpO2 Board. Within the study, biomedical/health related outcomes as it relates to the healthy volunteers are not evaluated. The data collected will be used to evaluate how accurate the pulse oximeter (DUT) is to the reference.

Devices under test are listed below:

  • Nellcor OxiMax SpO2 Board (453564624031) integrated within Philips IntelliVue X3 Patient Monitor (867030)
  • Philips AlarX5, Experimental SpO2 sensor based on the Nasal Alar design (nasal ala and earlobe placement)
  • Philips M1943NL SpO2 Adapter Cable (3m)

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Tidsramme: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.
Up to 6 weeks

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Tidsramme: Up to 6 weeks
Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.
Up to 6 weeks

Andre resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Evaluation of the pigmentation bias of the AlarX5 SpO2 sensor (nasal ala and earlobe application).
Tidsramme: Up to 6 weeks
Non-disparate performance for the AlarX5 SpO2 sensor.
Up to 6 weeks

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Faktiske)

28. mai 2026

Primær fullføring (Faktiske)

29. juni 2026

Studiet fullført (Faktiske)

29. juni 2026

Datoer for studieregistrering

Først innsendt

16. juni 2026

Først innsendt som oppfylte QC-kriteriene

25. juni 2026

Først lagt ut (Faktiske)

2. juli 2026

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

16. juli 2026

Siste oppdatering sendt inn som oppfylte QC-kriteriene

14. juli 2026

Sist bekreftet

1. juli 2026

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • SEF-2025-301123

Plan for individuelle deltakerdata (IPD)

Planlegger du å dele individuelle deltakerdata (IPD)?

UBESLUTTE

IPD-planbeskrivelse

This is currently under consideration.

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Ja

produkt produsert i og eksportert fra USA

Nei

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .

Abonnere