Safety and Clinical Performance of the Next-Generation SAFESYNC Transcatheter Heart Valve in Patients With Severe Aortic Stenosis (SAFESYNC NG)
Evaluation of the Safety and Clinical Performance of the Next-Generation SAFESYNC Balloon-Expandable Transcatheter Heart Valve in Patients With Severe Aortic Stenosis: A Prospective, Multicenter, Single-Arm Pivotal Study
This pivotal, prospective, multicenter (Brazil, Chile and Argentina), single-arm study will evaluate the safety and clinical performance of the Inovare® Safesync Next-Generation bioprosthetic heart valve, implanted by transfemoral transcatheter aortic valve implantation (TAVI) with the B-Delivery System, in patients aged 70 years or older with severe aortic stenosis and an indication for TAVI, regardless of surgical risk. The study will enroll 210 participants.
Panoramica dello studio
Stato
Intervento / Trattamento
Tipo di studio
Iscrizione (Stimato)
Fase
- Non applicabile
Contatti e Sedi
Contatto studio
- Nome: Clinical Research Department - Braile Biomédica
- Numero di telefono: +5517 21367000
- Email: pesquisa@braile.com.br
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
Severe aortic valve stenosis diagnosed according to criteria established in current guidelines; Age ≥70 years, regardless of surgical risk; Indication for transcatheter aortic valve implantation (TAVI); Willing to participate and to sign the written informed consent form.
Exclusion Criteria:
Any condition considered a contraindication to implantation of a bioprosthetic heart valve (e.g., patient with an indication for a mechanical heart valve);
Known hypersensitivity or contraindication to any of the following that cannot be adequately managed with premedication:
- Aspirin, clopidogrel, or heparin (heparin-induced thrombocytopenia [HIT] or heparin-induced thrombocytopenia with thrombosis [HITTS]);
- Nitinol (titanium or nickel);
- Contrast media; Blood dyscrasias: leukopenia (white blood cell count <1,000/mm³), thrombocytopenia (platelet count <100,000 cells/mm³), history of coagulopathy, or hypercoagulable states; Active infection, including endocarditis; Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or need for mechanical hemodynamic support; Recent stroke or transient ischemic attack (within 6 months); History of gastrointestinal bleeding or other bleeding that precludes anticoagulant or antiplatelet therapy; Estimated life expectancy of less than 5 years; Other medical, social, or psychological conditions that, in the investigator's opinion, prevent the patient from providing consent or adequately complying with the protocol; Pre-existing prosthetic heart valve in the aortic position (biological or mechanical); Concomitant severe mitral or tricuspid valve disease; Hypertrophic obstructive cardiomyopathy; Percutaneous coronary intervention performed within 72 hours prior to the study procedure; Femoral artery diameter <5.5 mm (for Safesync 20 to 24 mm), <6.0 mm (for Safesync 26 and 28 mm), or <6.5 mm (for Safesync 30 mm); High risk of annular rupture or significant residual paravalvular leak due to severe calcification of the aortic valve or left ventricular outflow tract; High risk of coronary obstruction (coronary ostial height <10 mm with narrow sinuses of Valsalva [SOV], defined as a mean SOV diameter <3 mm larger than the bioprosthesis size), as determined by computed tomography angiography; Left ventricular dysfunction with left ventricular ejection fraction (LVEF) ≤30%; Severe chronic kidney disease, defined as glomerular filtration rate <30 mL/min/1.73 m².
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: N / A
- Modello interventistico: Assegnazione di gruppo singolo
- Mascheramento: Nessuno (etichetta aperta)
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: INOVARE® SAFESYNC Next-Generation TAVI
|
Participants will undergo TAVI via transfemoral access using the INOVARE® SAFESYNC Next-Generation Bioprosthetic Heart Valve and the B-Delivery System. |
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Device Success According to VARC-3 Criteria
Lasso di tempo: 30 days after the procedure
|
Technical success; Absence of mortality; Absence of surgery or intervention related to the device or to major vascular, access-related, or cardiac structural complications; Intended valve performance, defined as mean transvalvular gradient <20 mmHg, peak velocity <3 m/s, Doppler velocity index ≥0.25, and less than moderate aortic regurgitation. |
30 days after the procedure
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Percentage of Participants Achieving Technical Success According to VARC-3 Criteria
Lasso di tempo: During the procedure
|
Defined as freedom from mortality; successful vascular access, delivery of the device, and retrieval of the delivery system; correct positioning of a single prosthetic heart valve in the intended anatomical location; and freedom from surgery or intervention related to the device or to a major vascular, access-related, or cardiac structural complication. Results will be reported as the number and percentage of participants achieving technical success. |
During the procedure
|
|
Incidence and Severity of Paravalvular Leak
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Paravalvular leak will be assessed by transthoracic echocardiography and classified according to severity. The number and percentage of participants with paravalvular leak in each severity category will be reported. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Mean Transvalvular Aortic Pressure Gradient Assessed
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Mean transvalvular aortic pressure gradient will be assessed by transthoracic echocardiography and reported in mmHg. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Distribution of Participants by New York Heart Association Functional Class
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Functional status will be assessed using the New York Heart Association (NYHA) Functional Classification. The classification ranges from Class I to Class IV, where Class I represents no limitation of physical activity and Class IV represents inability to perform physical activity without symptoms and/or symptoms at rest. Higher NYHA classes indicate worse functional status. Results will be reported as the number and percentage of participants in each NYHA functional class. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of New Permanent Pacemaker Implantation
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
New permanent pacemaker implantation following the index transcatheter aortic valve implantation procedure will be assessed. Results will be reported as the number and percentage of participants who undergo new permanent pacemaker implantation. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of All-Cause Mortality
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
All-cause mortality will be assessed throughout follow-up. Results will be reported as the number and percentage of participants who die from any cause. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of Stroke
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Stroke events will be assessed based on clinical neurological evaluation and classified according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants experiencing stroke. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of Bioprosthetic Valve-Related Reintervention
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Bioprosthetic valve-related reintervention will be defined as any surgical or transcatheter intervention performed to treat dysfunction or failure of the implanted study valve. Results will be reported as the number and percentage of participants undergoing valve-related reintervention. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of Bioprosthetic Valve Dysfunction According to VARC-3 Definitions
Lasso di tempo: 30 days, 6 months, 12 months, and annually through 5 years
|
Bioprosthetic valve dysfunction will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Bioprosthetic valve dysfunction includes structural valve deterioration, non-structural valve dysfunction, valve thrombosis, and endocarditis. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve dysfunction and according to the corresponding category. |
30 days, 6 months, 12 months, and annually through 5 years
|
|
Incidence of Bioprosthetic Valve Failure According to VARC-3 Definitions
Lasso di tempo: 30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure
|
Bioprosthetic valve failure will be assessed according to Valve Academic Research Consortium 3 (VARC-3) definitions. Results will be reported as the number and percentage of participants meeting the criteria for bioprosthetic valve failure. |
30 days, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the index procedure
|
Collaboratori e investigatori
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- SAFESYNC NG
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .