Automated Oxygen Titration With the EMILY.AI Device During the Six-Minute Walk Test
Functional, Clinical, and Predictive Validation Study of the EMILY AI Device in Patients With Acute and Chronic Respiratory Failure: Exercise Substudy (Six-Minute Walk Test) in Ambulatory Patients
This study evaluates EMILY.AI, an investigational (non-CE-marked) smart oxygen valve that automatically adjusts the oxygen flow delivered to a patient, second by second, according to oxygen saturation (SpO2), heart rate and respiratory rate measured by a pulse oximeter.
Adults with chronic respiratory disease who need oxygen during exertion perform two six-minute walk tests on the same day, 30 minutes apart: one with oxygen adjusted automatically by EMILY.AI and one with a fixed oxygen flow delivered through the same device. The order is randomized and participants do not know which test uses automatic titration.
The study compares the time spent within the target saturation range, desaturation, oxygen use, walking distance, breathlessness and participant preference between the two tests. This substudy is part of clinical investigation FJD-EMILY-23-02, whose other substudies are registered in separate records.
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
Background Patients with chronic respiratory disease often desaturate during exertion, and fixed-flow oxygen cannot follow the rapid changes in oxygen demand during walking and recovery. EMILY.AI combines a medical-grade oxygen valve with an edge-computing fuzzy-logic algorithm that sets the oxygen flow every second, with continuous logging of physiological signals and delivered flow.
Design Pilot, pragmatic, single-blind, randomized crossover study at Hospital Universitario Fundación Jiménez Díaz (Madrid, Spain). Each participant performs two supervised six-minute walk tests (6MWT) according to the American Thoracic Society guideline, separated by 30 minutes of rest: one with automatic titration by EMILY.AI and one with a fixed flow set by the physician, delivered through the same device so that the participant cannot tell them apart. The order of the two tests is randomized. Each test includes 6 minutes of walking and 2 minutes of standardized recovery.
Safety The medical team knows the device configuration and supervises the valve throughout each test.
Measurements SpO2, heart rate and respiratory rate are recorded with medical-grade pulse oximeters (Nonin, Masimo or Nihon Kohden models), together with distance walked, Borg dyspnoea score, oxygen delivered and participant-reported comfort, breathlessness and preference.
The protocol FJD-EMILY-23-02 also includes a substudy in the Respiratory Intermediate Care Unit and a substudy in a conventional hospital ward, registered separately.
Tipo di studio
Iscrizione (Stimato)
Fase
- Non applicabile
Contatti e Sedi
Contatto studio
- Nome: Alba Naya Prieto, MD
- Numero di telefono: +34 915 50 48 00
- Email: alba.naya@fjd.es
Backup dei contatti dello studio
- Nome: Sarah Beatrice Heili Frades, MD, PhD
- Numero di telefono: +34 915 50 48 00
- Email: SHeili@fjd.es
Luoghi di studio
-
-
-
Madrid, Spagna, 28040
- Reclutamento
- Hospital Universitario Fundación Jiménez Díaz
-
Contatto:
- Sarah Beatrice Heili Frades Zimmermann MD, PhD
- Numero di telefono: +34 915 50 48 00
- Email: SHeili@fjd.es
-
Contatto:
- Alba Naya Prieto, MD
- Numero di telefono: +34 915 50 48 00
- Email: alba.naya@fjd.es
-
Investigatore principale:
- Sarah Beatrice Heili Frades Zimmermann, MD, PhD
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
- Aged 40 to 90 years with a chronic respiratory disease.
- Indication for exertional oxygen therapy, defined as SpO2 below 88% during a six-minute walk test and/or resting respiratory failure (PaO2 below 60 mmHg).
- Able to understand the study and sign informed consent.
Exclusion Criteria:
- Decompensated acute or chronic disease with severe organ involvement, such as cardiovascular disease.
- Musculoskeletal or neurological disease preventing stable and safe walking.
- Comorbidities or treatments that, by intensity or dose, limit performance of the tests or interpretation of the results (such as alcohol use, beta-blockers, or other neurological or psychiatric drugs).
- Walking limitation preventing the six-minute walk test, or cardiovascular or connective tissue disease limiting accurate transcutaneous pulse oximetry.
- Pregnancy or breastfeeding (women of childbearing potential require a negative pregnancy test at screening).
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione incrociata
- Mascheramento: Separare
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: EMILY.AI first, then fixed flow
First 6MWT with automatic titration by EMILY.AI; after 30 minutes of rest, second 6MWT with fixed-flow oxygen through the same device. |
Investigational smart oxygen valve with an embedded fuzzy-logic algorithm that adjusts oxygen flow every second from pulse-oximetry SpO2, heart rate and respiratory rate during walking and recovery. Other names: EMILY.AI Smart Valve Standard care: constant oxygen flow set by the physician, delivered through the EMILY.AI device in fixed mode so that the participant cannot distinguish it from automatic titration. |
|
Sperimentale: Fixed flow first, then EMILY.AI
First 6MWT with fixed-flow oxygen through the device; after 30 minutes of rest, second 6MWT with automatic titration by EMILY.AI. |
Investigational smart oxygen valve with an embedded fuzzy-logic algorithm that adjusts oxygen flow every second from pulse-oximetry SpO2, heart rate and respiratory rate during walking and recovery. Other names: EMILY.AI Smart Valve Standard care: constant oxygen flow set by the physician, delivered through the EMILY.AI device in fixed mode so that the participant cannot distinguish it from automatic titration. |
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Percentage of time with SpO2 within the target range
Lasso di tempo: Day 1
|
Percentage of test time (6 minutes of walking plus 2 minutes of recovery) with SpO2, measured by pulse oximetry, within the predefined target range (88-92%), without delivering oxygen while SpO2 is above the range. Compared between the test with automatic titration by EMILY.AI and the test with fixed flow. |
Day 1
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Percentage of test time with SpO2 below 88%
Lasso di tempo: Day 1
|
Percentage of test time (walking and recovery) with SpO2 below 88%, compared between conditions. |
Day 1
|
|
Number of desaturation episodes with SpO2 below 88%
Lasso di tempo: Day 1
|
Number of episodes per test in which SpO2 falls below 88%, compared between conditions. |
Day 1
|
|
Time to start of oxygen flow correction
Lasso di tempo: Day 1
|
Time in seconds from SpO2 leaving the target range (88-96%) to the start of an oxygen flow change, compared between conditions. |
Day 1
|
|
Time to return of SpO2 to the target range
Lasso di tempo: Day 1
|
Time in seconds from SpO2 leaving the target range (88-96%) to its return to the range, compared between conditions. |
Day 1
|
|
Percentage of test time with oxygen delivered while SpO2 is above 92%
Lasso di tempo: Day 1
|
Percentage of test time (walking and recovery) during which oxygen is delivered while SpO2 is above 92%, compared between conditions. |
Day 1
|
|
Number of episodes with oxygen delivered while SpO2 is above 92%
Lasso di tempo: Day 1
|
Number of episodes per test in which oxygen is delivered while SpO2 is above 92%, compared between conditions. |
Day 1
|
|
Total oxygen delivered
Lasso di tempo: Day 1
|
Total volume of oxygen delivered (litres) during each test including recovery, compared between conditions. |
Day 1
|
|
Mean oxygen flow
Lasso di tempo: Day 1
|
Mean oxygen flow (litres per minute) delivered during each test including recovery, compared between conditions. |
Day 1
|
|
Six-minute walk distance
Lasso di tempo: Day 1
|
Distance walked (metres), compared between conditions. |
Day 1
|
|
Change in dyspnoea measured with the modified Borg Dyspnoea Scale
Lasso di tempo: Day 1: before and immediately after each six-minute walk test
|
Change from the start to the end of each test in dyspnoea measured with the modified Borg Dyspnoea Scale, which ranges from 0 (no breathlessness at all) to 10 (maximal breathlessness). Higher scores mean a worse outcome. Compared between conditions. |
Day 1: before and immediately after each six-minute walk test
|
|
Maximum heart rate
Lasso di tempo: Day 1
|
Maximum heart rate (beats per minute) during each test, compared between conditions. |
Day 1
|
|
Maximum respiratory rate
Lasso di tempo: Day 1
|
Maximum respiratory rate (breaths per minute) during each test, compared between conditions. |
Day 1
|
|
Perceived difference in exercise capacity between the two tests
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Did you notice differences in your exercise capacity in either of the two tests?", and, if so, in which test. Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
|
Perceived difference in breathlessness between the two tests
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Did you notice differences in shortness of breath in either of the two tests?", and, if so, in which test. Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
|
Perceived difference in comfort of oxygen delivery between the two tests
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Did you notice differences in the comfort of oxygen delivery in either of the two tests?", and, if so, in which test. Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
|
Preference between the two oxygen delivery systems
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Do you prefer either of the two oxygen systems used in the tests?", and, if so, which one. Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
|
Preference for continuous automatic flow regulation at home
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Would you prefer your home oxygen system to regulate the oxygen flow continuously?". Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
|
Preference for telemonitored home oxygen
Lasso di tempo: Day 1: within 1 hour after the second six-minute walk test
|
Number of participants answering no, partially or yes to the question "Would you prefer your home oxygen system to be telemonitored by the medical team?". Categorical 3-option response, not a scored scale. |
Day 1: within 1 hour after the second six-minute walk test
|
Collaboratori e investigatori
Sponsor
Collaboratori
Investigatori
- Investigatore principale: Sarah Beatrice Heili Frades, MD, PhD, MD, PhD, Hospital Universitario Fundación Jiménez Díaz
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
- Malattie vascolari
- Malattia cardiovascolare
- Processi patologici
- Malattia cronica
- Attributi della malattia
- Malattie delle vie respiratorie
- Malattie polmonari
- Malattie polmonari, ostruttive
- Ipertensione
- Condizioni patologiche, segni e sintomi
- Malattia polmonare, cronica ostruttiva
- Malattie polmonari, interstiziale
- Ipertensione, polmonare
Altri numeri di identificazione dello studio
- FJD-EMILY-23-02-TM6M
- FJD-EMILY-23-02 (Altro identificatore: Aether Tech SL)
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .