NBM-200 Clinical Study in Blood Donation Setting
The NBM-200 is a portable Hemoglobin and oximetry monitor, based on occlusion spectroscopy technology, for non-invasive spot checking of hemoglobin (Hb), estimated Hematocrit (Hct), SpO2 and pulse rate.
The NBM-200 includes a reusable ring-shaped sensor probe that fits on the patient's finger, and a portable desktop monitor that calculates and displays the measurement result.
The sensor probe consists of a multi-wavelength optical measuring system and inflatable cuff employing pneumatic tissue manipulation. Blood flow in the finger can be briefly occluded and the resulting changes in its optical behavior are analyzed to provide accurate measurements of Hb.
調査の概要
詳細な説明
The proposed study plan will support a 510(k) submission for the non-invasive NBM-200 system with a blood bank indication. In accordance with the agency's suggestion, OrSense plans to use as a predicate an FDA-cleared blood analyzer with a hemoglobin test method that is traceable to the cyanmethemoglobin method or that was standardized and calibrated using the International Council for Standardization in Hematology (ICSH) hemoglobin standard, e.g., the HemoCue-301 or EKF's Hemo Control Hemoglobin Measurement System. OrSense further plans to use the cleared NBM-200 (K142209) as a reference device. Clinical data will be collected in a blood donor population using 3 methods to check hemoglobin (Hb): (1) the non-invasive NBM-200 system; (2) a point-of-care device based on finger sticks that is FDA-cleared for use in blood banks, such as the HemoCue or EKF device; and (3) a laboratory analyzer based on venous blood samples.
Participants will be recruited via email and during their visit to the designated donor rooms.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Florida
-
Palm Harbor、Florida、アメリカ、34684
- Palm Harbor Donor Center
-
Saint Petersburg、Florida、アメリカ、33716
- OneBlood St. Pete Main Donor Center
-
Tampa、Florida、アメリカ、33624
- Tampa- Dale Mabry OneBlood Donor Center
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Regular donors and volunteers
Exclusion Criteria:
- Pregnancy
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:非ランダム化
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
他の:Comparison
Comparison study: The participants (at least 300) will be tested once with each of the 3 devices. First, a measurement with the NBM-200 will be performed according to the user manual. The best finger to use, in order of preference: 1) right thumb 2) left thumb 3) right index 4) left index. The POC capillary test will follow the NBM-200 test, and the venous test will be performed last. A venous blood sample will be obtained from all participants in the comparison study, even if a subject is designated to donate and is deferred from donation because of low hemoglobin. |
Finger Prick and Venipuncture
他の名前:
|
|
他の:Precision A & B
Precision study:
A laboratory test with venous blood is not required for participants in the precision study. |
Finger Prick and Venipuncture
他の名前:
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Bias
時間枠:24 hours
|
The mean bias between the NBM-200 and HemoCue readings and the laboratory reference
|
24 hours
|
協力者と研究者
スポンサー
協力者
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- OrSense-001
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
IPD プランの説明
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。